Ultraverse® 035 PTA Dilatation Catheter U3513084Recalls & reported problems
Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 8.0mm x 40mm balloon, 130cm shaft · Class 2
All models of Ultraverse 035 PTA Dilatation Catheter (92)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741094170), received between May 1, 2017 and Jun 18, 2025 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture10
- Detachment of Device or Device Component4
- Retraction Problem4
- Burst Container or Vessel2
- Device-Device Incompatibility2
- Difficult to Insert2
- Break1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE 035 PTA BALLOON CATHETER. DURING THE PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 7ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22282216 ↗MalfunctionNov 13, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN RIGHT COMMON ILIAC ARTERY, THE PTA BALLOON WAS ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20672693 ↗MalfunctionAug 30, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE RUPTURED AT THE RATED BURST PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17648988 ↗MalfunctionOct 26, 2022Inflation Problem; Material Rupture; Inflation Problem; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY HAD INFLATION ISSUE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15672749 ↗MalfunctionOct 10, 2022Break; Burst Container or Vessel; Difficult to Insert; Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Device-Device Incompatibility; Break; Burst Container or Vessel; Difficult to Insert; Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT COMMON FEMORAL, THE PTA BALLOON WAS ALLEGEDLY UNABLE TO BE INSERTED THROUGH THE SHEATH. IT WAS FURTHER REPORTED THAT THE BALLOON RUPTURED AND HAD DIFFICULTY IN RETRACTING THROUGH THE INTRODUCER SHEATH AND EVENTUALLY THE CATHETER BROKE OFF WITHIN THE SHEATH. IT WAS FURTHER REPORTED THAT THE DEVICE WAS REMOVED AS SINGLE UNIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15574057 ↗MalfunctionAug 31, 2021Burst Container or Vessel; Difficult to Insert; Material Rupture; Device-Device Incompatibility; Burst Container or Vessel; Difficult to Insert; Material Rupture; Device-Device Incompatibility
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY HAD FAULT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12390631 ↗MalfunctionNov 30, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10918056 ↗MalfunctionOct 27, 2020Retraction Problem; Retraction Problem
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL U3513084 PTA BALLOON DILATATION CATHETER ALLEGEDLY RETRACTION ISSUE. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO PATIENT CONSEQUENCES. AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10744368 ↗MalfunctionApr 15, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. THE INFORMATION REVIEWED INDICATED THAT MODEL U3513084 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCES. PATIENT INFORMATION WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9965005 ↗InjuryApr 2, 2018Retraction Problem; Material Rupture; Detachment of Device or Device Component; Retraction Problem; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DURING AN ANGIOPLASTY PROCEDURE IN A A/V FISTULA, THE PTA BALLOON CATHETER ALLEGEDLY RUPTURED PAST RATED BURST PRESSURE. REPORTEDLY, THE PTA BALLOON ALLEGEDLY BECAME STUCK INSIDE THE 6FR SHEATH DURING RETRACTION. IT WAS FURTHER REPORTED THAT WHEN THE BALLOON AND SHEATH WERE REMOVED FROM THE PATIENT, A SEGMENT OF THE BALLOON MATERIAL ALLEGEDLY DETACHED. THE PATIENT WAS SENT TO SURGERY FOR REMOVAL OF THE DETACHED BALLOON SEGMENT. THE PATIENT WAS REPORTED TO BE DOING WELL POST SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7390848 ↗MalfunctionFeb 15, 2018Retraction Problem; Detachment of Device or Device Component; Retraction Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT AFTER A SUCCESSFUL ANGIOPLASTY PROCEDURE THE PTA BALLOON CATHETER ALLEGEDLY GOT STUCK INSIDE THE 5FR SHEATH AND SUBSEQUENTLY THE BIFURCATION DETACHED FROM THE CATHETER SHAFT. REPORTEDLY, THE REMAINING CATHETER WAS PULLED OUT FROM THE SHEATH AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7273774 ↗MalfunctionMay 1, 2017Retraction Problem; Material Rupture; Detachment of Device or Device Component; Retraction Problem; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT RENAL ARTERY THE PTA BALLOON ALLEGEDLY RUPTURED CAUSING THE TIP OF THE BALLOON TO DETACH INSIDE THE SHEATH UPON REMOVAL FROM THE PATIENT. REPORTEDLY, THE HEALTH CARE PROVIDER (HCP) FOUND THAT THE TIP OF THE BALLOON AND RADIOPAQUE MARKER WAS MISSING. UPON FURTHER EXAMINATION THE HCP WAS ABLE TO REMOVE THE RADIOPAQUE MARKER. HOWEVER, IT WAS DECIDED BY THE HCP NOT REMOVE THE BALLOON TIP DESPITE MULTIPLE ATTEMPTS. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6535849 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page