Skip to content

Ultraverse® 035 PTA Dilatation Catheter U357594Recalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 9.0mm x 40mm balloon, 75cm shaft · Class 2

All models of Ultraverse 035 PTA Dilatation Catheter (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741094347), received between Jan 6, 2021 and Jun 9, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove5
  • Material Rupture4
  • Detachment of Device or Device Component3
  • Device-Device Incompatibility2
  • Difficult to Insert2
  • Break1
  • Entrapment of Device1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2025Material Rupture; Material Rupture

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRAVERSE 035 PTA BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY RUPTURED AT 8 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22165367 ↗
  • MalfunctionFeb 13, 2025Difficult to Insert; Difficult to Remove; Stretched; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation; Difficult to Insert; Difficult to Remove; Stretched; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation

    ON (B)(6) 2024, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING ULTRAVERSE 035 DILATATION CATHETER. DURING THE PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO REMOVE THROUGH THE SHEATH. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE WAS LEFT AND SWITCHED THE SHEATH TO KEEP ACCESS TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21371422 ↗
  • MalfunctionOct 21, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 8 ATM WHEN THE LESION EXPANDS. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20490925 ↗
  • InjuryJun 12, 2023Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Difficult to Remove; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THREE CENTIMETER OF THE PTA BALLOON WAS ALLEGEDLY REMAINED IN THE VESSEL AND WAS UNABLE TO REMOVE THE PIECE. REPORTEDLY, THE PTA BALLOON PIECE WAS ATTACHED TO THE VESSEL WALL WITH A STENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 17105360 ↗
  • MalfunctionJun 28, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 7ATM AT THE FIRST INFLATION ATTEMPT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14840133 ↗
  • MalfunctionApr 7, 2022Difficult to Insert; Difficult to Remove; Device-Device Incompatibility; Difficult to Insert; Difficult to Remove; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE BELOW THE KNEE, THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO INSERT. IT WAS FURTHER REPORTED THAT THE DEVICE WAS HARD TO REMOVE AFTER INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14039674 ↗
  • MalfunctionDec 13, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY STUCK UPON REMOVING FROM THE SHEATH. IT WAS FURTHER REPORTED THAT DEVICE WAS REMOVED AS SINGLE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12982337 ↗
  • MalfunctionJan 6, 2021Break; Entrapment of Device; Difficult to Remove; Retraction Problem; Detachment of Device or Device Component; Break; Entrapment of Device; Difficult to Remove; Retraction Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY STUCK AND UNABLE TO REMOVE. IT WAS FURTHER REPORTED THAT THE SHAFT OF THE CATHETER ALLEGEDLY BROKE. THE BALLOON AND THE SHEATH WERE REMOVED TOGETHER AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11123016 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →