Skip to content

Ultraverse® 035 PTA Dilatation Catheter U351301010Recalls & reported problems

Bard Peripheral Vascular, Inc. · Ultraverse 035 PTA Dilatation Catheter 10.0mm x 100mm balloon, 130cm shaft · Class 2

All models of Ultraverse 035 PTA Dilatation Catheter (92)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741094736), received between Nov 7, 2023 and Nov 7, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Device-Device Incompatibility1
  • Difficult to Insert1
  • Difficult to Remove1
  • Failure to Advance1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2023Difficult to Insert; Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Device Dislodged or Dislocated; Material Deformation; Difficult to Insert; Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Device Dislodged or Dislocated; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE ILIAC ARTERY VIA THE FEMORAL ARTERY APPROACH, THE JOINT OF THE BALLOON AND THE DELIVERY SYSTEM WAS ALLEGEDLY MISLOCATED, AND THE JOINT WAS ALLEGEDLY DEFORMED. IT WAS FURTHER REPORTED THAT DUE TO MISLOCATION, THERE WAS OBSTRUCTION OF THE GUIDE WIRE PASSING AND THE GUIDEWIRE ALLEGEDLY GOT STUCK IN THE SHEATH, AND WAS UNABLE TO BE ADVANCED. FURTHERMORE, THE CATHETER WAS ALLEGEDLY DIFFICULT TO BE REMOVED, AND EVENTUALLY THE CATHETER AND SHEATH WERE WITHDRAWN TOGETHER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18083359 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates