TRUE DILATATION 0284514Recalls & reported problems
Bard Peripheral Vascular, Inc. · TRUE™ DILATATION Balloon Valvuloplasty Catheter, 28 mm x 4.5 cm,110 cm catheter · Class 2
All models of TRUE DILATATION (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741097478), received between Jan 10, 2020 and Aug 5, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove3
- Material Rupture3
- Break2
- Deflation Problem2
- Detachment of Device or Device Component2
- Leak/Splash1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50137947 ↗MalfunctionJul 15, 2026
TWO DEVICES: 1. DURING A TAVR [TRANSCATHETER AORTIC VALVE REPLACEMENT] VALVE PROCEDURE, A BARD TRUE DILATION BALLOON (26MM) BURST DURING NORMAL USE. THE MANUFACTURER WAS NOTIFIED BY THE CATH LAB PERSONNEL¿NO HARM TO THE PATIENT. 2. THE SURGEON INFLATED THE TRUE DILATION BALLOON, AND IT RUPTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25826152 ↗MalfunctionJan 2, 2025Difficult to Remove; Detachment of Device or Device Component; Material Deformation; Difficult to Remove; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING A BALLOON AORTIC VALVULOPLASTY PROCEDURE IN THE PULMONARY VALVE, THE BALLOON ALLEGEDLY FAILED TO PASS THROUGH THE SHEATH. IT WAS FURTHER REPORTED THAT THE BALLOON MIGRATED TOWARDS THE DISTAL END OF THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21052635 ↗MalfunctionMay 16, 2024Deflation Problem; Leak/Splash; Deflation Problem; Leak/Splash
T WAS REPORTED THAT DURING A VALVULOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY HAD A LEAK ON THE SHAFT. IT WAS FURTHER REPORTED THAT THE AIR ALLEGEDLY KEPT COMING IN WHEN TRYING TO DEFLATE THE BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19328691 ↗MalfunctionApr 17, 2024Break; Deflation Problem; Difficult to Remove; Break; Deflation Problem; Difficult to Remove
IT WAS REPORTED THAT DURING A VALVULOPLASTY PROCEDURE, THE PTV BALLOON ALLEGEDLY PARTIALLY SEPARATED AT THE SHAFT AND THE BALLOON ALLEGEDLY WASN'T COMPLETELY DEFLATED. IT WAS FURTHER REPORTED THAT THE ACTUAL BALLOON WAS STILL CONNECTED VIA THE LUMEN TO FILL THE BALLOON. FURTHERMORE, IT WAS ALLEGEDLY HARD TO REMOVE THROUGH THE SHEATH. REPORTEDLY, IT WAS REMOVED ALL TOGETHER AND THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19123674 ↗InjuryJan 23, 2024Break; Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Break; Difficult to Remove; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A VALVULOPLASTY PROCEDURE, THE PTV BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE BALLOON ALLEGEDLY DETACHED FROM THE CATHETER SHAFT BUT STAYED ATTACHED AT THE TIP WHICH CAUSED IT TO FOLD OVER ITSELF AND PREVENTED REMOVAL FROM THE VEIN. FURTHERMORE, THE PHYSICIAN WAS ALLEGEDLY UNABLE TO PULL THE BALLOON OUT OF THE SHEATH. REPORTEDLY, A RIGHT COMMON FEMORAL ARTERY EXPLORATION WAS MADE FOR FOREIGN BODY REMOVAL THROUGH THE COMMON FEMORAL VEIN LONGITUDINAL VENOTOMY, AND PRIMARY REPAIR OF RIGHT COMMON FEMORAL VEIN VENOTOMY WAS PERFORMED TO REMOVE THE DEVICE AND REPAIR VASCULATURE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18564074 ↗MalfunctionJan 10, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0284514 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE 2 MATERIAL RUPTURES, 2 INVOLVED PATIENTS WITH NO PATIENT CONSEQUENCES. THERE WAS NO PROVIDED PATIENT INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9576566 ↗
Other Balloon aortic valvuloplasty devices
- TRUE DILATATION 0224512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty23 reports
- TRUE DILATATION 0204511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty21 reports
- TRUE DILATATION 0244512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty2 deaths19 reports
- TRUE DILATATION 0184511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty12 reports
- TRUE DILATATION 0264513Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty11 reports
- TRUE FLOW TF0203511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE DILATATION 0214512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE DILATATION 0234512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty1 death10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.