Bard® Nasogastric Sump Tube and PreVent® Filter and Lopez Valve® with ENFit™ EN0056160Recalls & reported problems
C. R. Bard, Inc. · Bard® Nasogastric Sump Tube and PreVent® Anti-Reflux Filter and Lopez Valve® with ENFit™ · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741113864), received between Mar 29, 2019 and Feb 14, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Complete Blockage” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 20, 2023Class IIOpen, Classified
BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced suction, decompression, and drainage occurred.
Z-0804-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Complete Blockage1
- Defective Component1
- Inadequacy of Device Shape and/or Size1
- Incorrect Measurement1
- Partial Blockage1
- Suction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 14, 2024Defective Component; Defective Component
IT WAS REPORTED THAT THE NASOGASTRIC TUBES HAD UNKNOWN ISSUE. (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18701844 ↗MalfunctionDec 28, 2023Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE CURRENTLY STOCKED BARD SUMP NASOGASTRIC TUBES WERE NOT DRAINING APPROPRIATELY WHEN CONNECTED TO SUCTION DRAINAGE. CUSTOMER WAS TOLD OF UP TO 4 INSTANCES WHERE THE TUBE HAD TO BE REPLACED SHORTLY AFTER BEING PLACED DUE TO CLOGGING OR NOT DRAINING. THIS RESULTED IN DELAYS IN CARE FOR THEIR PATIENT (I.E. DRAINAGE OF STOMACH CONTENTS), AND DISSATISFACTION WITH MULTIPLE INVASIVE PROCEDURES TO REPLACE TUBES, NOT TO MENTION THE ADDITIONAL X-RAYS EXPOSURE REQUIRED TO VERIFY TUBE PLACEMENT. COMPARED TO THE COVIDIEN NG SALEM SUMP TUBES CUSTOMER HAD BEEN USING, THE SIZE OF THE HOLE TO ALLOW DRAINAGE FROM SUCTION WAS SIGNIFICANTLY SMALLER IN THE BARD. THEY WERE CONCERNED ABOUT THE REPEATED INVASIVE PROCEDURE THAT CAN CAUSE TRAUMA, DISCOMFORT, PATIENT DISSATISFACTION AND REPEATED EXPOSURE TO RADIATION. THERE WAS ANOTHER SET OF STAFF WITH THE SAME CONCERNS. CUSTOMER HAD MULTIPLE AREAS EXPRESSED CONCERNS WITH THE BARD SUB FOR THE NASOGASTRIC TUBES. CUSTOMER THOUGHT THE BARD TUBE SHOULD BE PULLED FROM USE IMMEDIATELY. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 14DEC2023, IT WAS REPORTED THAT THE NASOGASTRIC TUBE HAVING ISSUES WITH CLOGGING AS THEY USED SUCTION. STATE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18410268 ↗MalfunctionMar 29, 2019Complete Blockage; Incorrect Measurement; Inadequacy of Device Shape and/or Size; Suction Problem; Incorrect Measurement
IT WAS REPORTED THAT THE SUMP TUBES WITH ENFIT CONNECTION HAD A SMALLER INTERNAL DIAMETER THAN THE TUBES THEY HAD PREVIOUSLY RECEIVED. THE NURSES HAVE STATED THAT THEY WERE HAVING A HARD TIME GETTING ADEQUATE SUCTION DUE TO THE SMALLER SIZE AND AS A RESULT THE TUBES WERE GETTING CLOGGED WHEN THERE WERE LARGE FOOD PARTICLES AND/OR PILL PARTICLES IN THE GASTRIC CONTENTS. THIS WAS A COMMON OCCURRENCE IN THE EMERGENCY DEPARTMENT AND FOR PATIENT'S GOING EMERGENTLY TO THE OPERATING ROOM.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8467207 ↗
Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices
- Salem Sump 8888266114Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION2 recalls0 reports
- CARDINAL HEALTH 8888266122ECardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall9 reports
- CARDINAL HEALTH 8888266130ECardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall7 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046160C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall7 reports
- Salem Sump 8888266148Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall5 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046180C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall5 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter 0046160C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION54 reports
- Bard® Nasogastric Sump Tube and PreVent® Filter and Lopez Valve® with ENFit™ EN0056140C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page