Site-Rite 9770177RRecalls & reported problems

Bard Access Systems, Inc. · REFURB, SR VISION II TOWER · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741125362), received between Jun 4, 2018 and Apr 8, 2019 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Difficult or Delayed Activation1
  • Loss of Power1
  • Operating System Becomes Nonfunctional1
  • Reset Problem1
  • Unexpected Shutdown1

Patient problems

  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2019Unexpected Shutdown

    SALES REP REPORTED: THE SYSTEM SHUTS DOWN MULTIPLE TIMES PER PICC PLACEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8494117 ↗
  • MalfunctionSep 28, 2018Reset Problem

    BIOMED REPORTED THAT VISION II TOWER IS FREEZING UP IN SHERLOCK MODE DURING PROCEDURE CAUSING THE USERS TO HAVE TO REBOOT IN THE MIDDLE OF PROCEDURE. THE SYSTEM'S TOUCHSCREEN FUNCTIONS ALSO INTERMITTENTLY SCROLL THROUGH OPTIONS WITHOUT COMMAND. THE TWO REPORTED ISSUES MAY BE RELATED TO THE SAME ROOT CAUSE, ALTHOUGH IT CANNOT YET BE DETERMINED UNTIL EVALUATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7920457 ↗
  • MalfunctionSep 7, 2018Loss of Power

    IT WAS REPORTED BY THE PICC TEAM THAT VISION II ULTRASOUND SYSTEM IS SHUTTING DOWN WITHOUT COMMAND EVEN WHILE BEING PLUGGED INTO AC POWER. THE SYSTEM DID SHUT DOWN DURING A PROCEDURE, HOWEVER, THE PICC TEAM WAS ABLE TO REBOOT SYSTEM AND CONTINUE WITH PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7858217 ↗
  • MalfunctionJun 4, 2018Difficult or Delayed Activation; Operating System Becomes Nonfunctional

    SR VISION II ULTRASOUND SYSTEM WAS REPORTED TO HAVE AN INTERMITTENT DELAY IN SYSTEM OPERATION AND WOULD OCCASIONALLY FREEZE MID-PROCEDURE. THE BIOMEDICAL ENGINEER WAS UNABLE TO DUPLICATE THE REPORTED ISSUE. FIELD ASSURANCE TECH SUPPORT HAD THE ENGINEER RUN THE SYSTEM DISK OPTIMIZATION AND ADVISED THE ENGINEER TO CONTACT FIELD ASSURANCE IF THE ISSUE PERSISTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7567528 ↗

Other System, imaging, pulsed doppler, ultrasonic devices

All System, imaging, pulsed doppler, ultrasonic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.