Savvy Long PTA Dilatation Catheter 436-2022LRecalls & reported problems
Clearstream Technologies Limited · Savvy Long PTA Dilatation Catheter 2.0mm x 220mm balloon, 120cm shaft · Class 2
All models of Savvy Long PTA Dilatation Catheter (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741132742), received between Apr 16, 2019 and Jul 11, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Break1
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 11, 2019Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 436-2022L PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED A BREAK AND DETACHMENT. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8783147 ↗InjuryApr 16, 2019Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
POST DILATION, IT WAS DIFFICULT TO REMOVE THE SAVVY LONG BALLOON CATHETER VIA THE SHEATH AS IT BECAME ELONGATED AND DETACHED FROM THE CATHETER. THERE WAS NO PATIENT INJURY. THE DEVICE WAS REMOVED FROM THE PATIENT WITH USE OF A GOOSE NECK SNARE. THE TARGET LESION IS THE ANTERIOR TIBIAL ARTERY WHICH WAS MODERATELY CALCIFIED WITH A COMBINATION OF STENOSES AND OCCLUSIONS, APPROX 20 CM IN LENGTH. THE DEVICE WAS PREP ACCORDING TO THE INSTRUCTION FOR USE WITH NO DEFECTS NOTED DURING PREP. A CONSTANT AND DEDICATED SALINE SOURCE WAS UTILIZED THROUGH THE BALLOON DURING PREP. THERE WERE NO DIFFICULTIES REMOVING THE PRODUCT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. A COMPETITOR'S CONTRAST MEDIA AND INFLATION DEVICE WAS USED WITH A RATION OF SIXTY/FORTY. THE INFLATION DEVICE WAS SUCCESSFULLY USED WITH OTHER DEVICES. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. THE CATHETER HAS NEVER BEEN IN AN ACUTE BEND. A 4FR BRITE TIP SHEATH WAS USED AND THEN UPSIZED TO 6FR BRITE TIP SHEATH. THE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8520326 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page