Skip to content

Sleek OTW PTA Dilatation Catheter 426-2522XRecalls & reported problems

Clearstream Technologies Limited · Sleek OTW PTA Dilatation Catheter 2.5mm x 220mm balloon, 150cm shaft · Class 2

All models of Sleek OTW PTA Dilatation Catheter (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741133701), received between Jul 16, 2019 and May 23, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Break2
  • Device Dislodged or Dislocated1
  • Difficult to Open or Remove Packaging Material1
  • Material Deformation1
  • Material Rupture1
  • Packaging Problem1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2025Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    ON (B)(6) 2024, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING A SLEEK OTW PTA CATHETER. PRIOR TO THE PROCEDURE, THE PTA BALLOON CATHETER FELL APART AND DISASSEMBLED WHEN REMOVING THE PROTECTIVE BALLOON COVER. AN UNKNOWN DEVICE WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22076218 ↗
  • MalfunctionFeb 24, 2025Break; Stretched; Detachment of Device or Device Component; Break; Stretched; Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, AFTER REMOVING THE BALLOON SLEEVE THE BALLOON WAS ALLEGEDLY DETACHED FROM THE CATHETER SHAFT. IT WAS FURTHER REPORTED THAT A BREAK WAS CONFIRMED AT THE BALLOON TAPER POINT AND THE CATHETER OUTER SHAFT. IT WAS FURTHER REPORTED THAT THERE WAS EVIDENCE OF INNER STRETCHED WITHIN THE BREAK AREA. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21444597 ↗
  • MalfunctionMar 14, 2024Break; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material; Break; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE IN THE TIBIAL FEMORAL ARTERY VIA THE COMMON FEMORAL ARTERY, THE DILATATION CATHETER WAS ALLEGEDLY STRIPPED RIGHT OFF THE SHAFT WHEN THE SCRUB TECHNICIAN REMOVED THE BALLOON FROM THE PLASTIC BALLOON HOUSING. IT WAS FURTHER REPORTED THAT THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER OR ANY OF THE STERILE PACKAGING COMPONENTS. REPORTEDLY, THE DEVICE DID NOT SEPARATE INSIDE THE PATIENT AND THE BALLOON WAS NOT CAUGHT IN THE LESION. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18902795 ↗
  • MalfunctionJul 15, 2020Detachment of Device or Device Component; Material Deformation; Packaging Problem; Detachment of Device or Device Component; Material Deformation; Packaging Problem

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL 426-2522X PTA CATHETER ALLEGEDLY EXPERIENCED PACKAGING PROBLEM, MATERIAL DEFORMATION AND DETACHMENT OF DEVICE OR DEVICE COMPONENT. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS EVENT DID NOT INVOLVE A PATIENT AS THERE WAS NO PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 10279918 ↗
  • MalfunctionJul 16, 2019Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 436-2522X PTA CATHETER EXPERIENCED AN ALLEGED MATERIAL RUPTURE. THIS REPORT WAS RECEIVED FROM A SINGLE SOURCE. THE ALLEGED MATERIAL RUPTURE INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8795521 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →