Skip to content

Sleek OTW PTA Dilatation Catheter 426-5010XRecalls & reported problems

Clearstream Technologies Limited · Sleek OTW PTA Dilatation Catheter 5.0mm x 100mm balloon, 150cm shaft · Class 2

All models of Sleek OTW PTA Dilatation Catheter (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741133992), received between Jan 6, 2021 and Dec 14, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove3
  • Detachment of Device or Device Component2
  • Break1
  • Deflation Problem1
  • Device-Device Incompatibility1
  • Material Deformation1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2023Break; Deflation Problem; Difficult to Remove; Device-Device Incompatibility; Material Deformation; Break; Deflation Problem; Difficult to Remove; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY FAILED TO DEFLATE. IT WAS FURTHER REPORTED THAT THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO BE REMOVED FROM THE GUIDEWIRE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18325320 ↗
  • InjuryJan 6, 2021Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    AS THE BALLOON OF A 5MM X 10CM X 150CM SLEEK OVER-THE-WIRE (OTW) PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER (BC) WAS BEING WITHDRAWN FROM THE PATIENT, IT GOT CAUGHT UP ON THE TIP OF THE 6F NON-CORDIS SHEATH AND WASN¿T ABLE TO BE REMOVED. AS FURTHER ATTEMPTS WERE MADE TO WITHDRAW THE SLEEK BALLOON, THE ENTIRE BALLOON SEPARATED FROM THE SHAFT AND REMAINED IN THE PATIENT¿S TIBIAL ARTERY. THE REMAINING SHAFT WAS REMOVED OVER THE WIRE AND THE PATIENT WAS SENT TO SURGERY AFTER FAILED ATTEMPT AT SNARING THE BALLOON. THE BALLOON WAS REMOVED BY A VASCULAR SURGEON IN THE OPERATION ROOM (OR) BY CUTDOWN AFTER BEING TRANSPORTED TO THE HOSPITAL. THE PATIENT IS DOING WELL WITH NO ISSUES. THE DEVICE WAS STORED AND HANDLED AS PER THE INSTRUCTION FOR USE (IFU). THERE WAS NO DAMAGE NOTED TO THE PACKAGING OF THE DEVICE. THERE WERE NO ANOMALIES NOTED WHEN REMOVED FROM THE PACKAGE. THE DEVICE WAS OPENED IN STERILE FIELD. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. THE INTENDED PROCED

    Patient problems: Insufficient Information

    MAUDE report 11125058 ↗
  • MalfunctionJan 6, 2021Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PTA PROCEDURE WITH THE INTENDED TREATMENT SITE TO BE THE SFA FROM PEDAL ACCESS, THE LESION WAS SEVERELY CALCIFIED, THE VESSEL WAS MODERATELY TORTOUS IN NATURE AND THE PATIENT HAD 99% STENOSIS. THE DEVICE WAS ALLEGEDLY DIFFICULT TO BE REMOVED. IT WAS FURTHER REPORTED THAT THE DEVICE SEPARATED FROM THE SHAFT AND REMAINED INSIDE THE TIBIAL ARTERY PATIENT. THE DETACHED PIECE WAS REMOVED AFTER FURTHER INTERVENTION. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11122266 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →