Skip to content

Peripheral Catheter Insertion Kit ST118101Recalls & reported problems

Bard Access Systems, Inc. · Sldgr IDC, 18gx10cm, Full · Class 2

All models of Peripheral Catheter Insertion Kit (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741139291), received between Jan 23, 2020 and Nov 11, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED BY THE CUSTOMER, "WE HAVE A NURSE THAT PLACED A MIDLINE ON A PATIENT. SHE USED THE ST KIT, WITH THE INTRODUCER/WIRE. WHEN PLACING MIDLINE, SHE ACCESSED THE VEIN AND ADVANCED THE WIRE AS NORMAL. SHE PLACED THE MIDLINE/INTRODUCER COMBO ON THE END OF THE WIRE AND ADVANCED INTO PLACE APPROPRIATELY. SHE THEN WITHDREW THE WIRE, WHILE LEAVING THE INTRODUCER IN PLACE. THE EXTENSION TUBING WAS HOOKED UP TO THE END OF THE INTRODUCER AND THE MIDLINE WAS USED THIS WAY, WITH THE INTRODUCER IN PLACE. A FEW DAYS INTO THE DWELL TIME, A FLOOR NURSE NOTICED IT WAS FLUSHING SLUGGISHLY AND WAS HAVING DIFFICULTY WITH BLOOD DRAW. WHEN THE IV TEAM NURSE WENT TO ASSESS THE MIDLINE, SHE NOTED THE WHITE INTRODUCER PIECE WAS IN PLACE WITH THE EXTENSION HOOKED UP TO THE INTRODUCER. WHEN REMOVING MIDLINE, SHE ALSO NOTED THE INTRODUCER IN PLACE AT THE TIP." ADDITIONAL INFORMATION RECEIVED ON OCTOBER 21, 2024: IT WAS REPORTED DATE OF EVENT WAS 10/13/2024. MIDLINE WAS FLUSHING WITH RESISTANCE, SO LINE WAS PULLED. THE INTRODUCER WAS FOUND TO BE IN PLACE STILL. DUE TO PATIENT BEING DIFFICULT INTRAVENOUS ACCESS AND HAVING THE MIDLINE PULLED, PATIENT HAD A JUGULAR CENTRAL LINE PLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20654692 ↗
  • MalfunctionJan 23, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE MIDLINE WAS PLACED(B)(6)2019 , WAS FOUND TO BE LEAKING ON 10/24/19. THE LINE WAS REPLACED ON 10/25/19. PATIENT WAS NOT HARMED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9624859 ↗

Other Midline catheter devices

All Midline catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page