Peripheral Catheter Insertion Kit ST020101DRecalls & reported problems
Bard Access Systems, Inc. · Sldgr IDC, 20gx10cm, Max · Class 2
All models of Peripheral Catheter Insertion Kit (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741139376), received between Jan 30, 2020 and Mar 4, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Fluid/Blood Leak1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient2
- Exposure to Body Fluids1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT THE POWERGLIDE LEAKED AT EXTENSION. THE CLINICIAN TOOK OFF EXTENSION AND PUT ON A DIFFERENT EXTENSION, LEFT FOR AN HOUR AND CAME BACK TO THE BED AND IT WAS LEAKING. TOOK OFF EXTENSION AND NO CRACK IN EXTENSION OR HUB THAT WAS NOTICED. THE CLINICIAN SAID THAT IT LOOKS LIKE THERE IS A DENT IN THE HUB OF CATHETER. CHANGED POWERGLIDE OVER WIRE. CHANGED EXTENSION AND NO ISSUES OR LEAKING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9789084 ↗MalfunctionFeb 7, 2020Break; Break
IT WAS REPORTED THAT MIDLINE PLACED AND PATIENT DISCHARGED TO HOMECARE. PATIENT RETURNED TO THE ED WITH MIDLINE BROKEN OFF AND MISSING HUB. APPROXIMATELY 2CM OF CATHETER OUT OF THE SKIN. MIDLINE REMOVED WITH TIP INTACT, MEASURED 8CM. THIS WAS A 10CM CATHETER AT INSERTION. NURSE SPOKE WITH HOMECARE RN WHO STATED "IT LOOKED KINKED AT A 90 DEGREE ANGLE AND WHEN SHE TOOK THE DRESSING OFF THE HUB WAS NOT ATTACHED, "THE LINE WAS SEVERED". PATIENT WAS ALERT AND ORIENTED WHEN BEING ADMITTED TO THE ED. CATHETER WAS REMOVED WITH NO ISSUES. NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9681646 ↗MalfunctionJan 30, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE POWERGLIDE WAS INSERTED ON (B)(6) 2019 IN AC OF COMBATIVE PATIENT. RN FLUSHED CATHETER ON (B)(6) 2019 AND NOTICED PINHOLE IN EXPOSED PART OF CATHETER. BODILY FLUID SPRAYED IN FACE OF RN UPON FLUSHING. RN UNDERWENT TESTING TO ELIMINATE RISK OF TRANSMISSION. THE CATHETER WAS REMOVED.
Patient problems: Exposure to Body Fluids
MAUDE report 9650088 ↗
Other Midline catheter devices
- Midline Catheter Kit P4154108DBard Access Systems, Inc. · Midline catheter1 recall1 death126 reports
- ARROW IPN921959Teleflex Incorporated · Midline catheter2 recalls0 reports
- Midline Catheter Kit S4153108BDPBard Access Systems, Inc. · Midline catheter1 recall36 reports
- Midline Catheter Kit S4153108DBard Access Systems, Inc. · Midline catheter1 recall32 reports
- Midline Catheter Kit S4153108DGPBard Access Systems, Inc. · Midline catheter1 recall15 reports
- Midline Catheter Kit S4254108BDPBard Access Systems, Inc. · Midline catheter1 recall11 reports
- Midline Catheter Kit P4154108Bard Access Systems, Inc. · Midline catheter1 recall10 reports
- ARROW IPN037698Teleflex Incorporated · Midline catheter1 recall9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗