MISSION 1410MSKRecalls & reported problems
Bard Peripheral Vascular, Inc. · Mission Disposable Core Biopsy Instrument Kit - 14Gx10cm · Class 2
All models of MISSION (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801741142543), received between Apr 14, 2022 and Aug 6, 2025 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2023Open, Classified
The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.
Z-0643-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Remove11
- Material Twisted/Bent11
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Device Misassembled During Manufacturing /Shipping1
- Failure to Fire1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2025Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED BREAST BIOPSY THROUGH NORMAL DENSITY TISSUE WITH THE MISSION KIT. DURING THE PROCEDURE, THE NEEDLE WAS BENT. IT WAS FURTHER REPORTED THAT IT WAS SLIGHTLY DIFFICULT TO REMOVE. COAXIAL WAS USED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22724286 ↗MalfunctionJul 2, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE AFTER SAMPLING PROCESS WHILE REMOVING THE NEEDLE, A RECTANGLE-SHAPED PLASTIC ALLEGEDLY WAS FOUND IN THE NOTCH OF THE NEEDLE INTO THE SAMPLING CONTAINER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19653271 ↗MalfunctionDec 13, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE ALLEGEDLY HAD DIFFICULTY IN REMOVING FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18316839 ↗MalfunctionDec 6, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE OF THE DEVICE WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED WITH FORCE FROM THE PATIENT'S BREAST. A COAXIAL WAS USED AND PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18273373 ↗MalfunctionNov 6, 2023Difficult to Remove; Device Misassembled During Manufacturing /Shipping; Material Twisted/Bent; Difficult to Remove; Device Misassembled During Manufacturing /Shipping; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE, THE NOTCH OF NEEDLE WAS ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED OUT FROM THE BREAST BY FORCE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18071075 ↗MalfunctionNov 3, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NOTCH OF NEEDLE WAS ALLEGEDLY BENT WHILE PENETRATION AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED OUT FROM THE BREAST BY FORCE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18065309 ↗MalfunctionOct 25, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE, THE NOTCH OF NEEDLE ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY REMOVED FROM THE PATIENT'S BREAST BY EXCESSIVE FORCE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18002158 ↗MalfunctionOct 12, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE INNER NEEDLE ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED OUT FROM THE BREAST BY FORCE. A COAXIAL NEEDLE WAS USED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17925861 ↗MalfunctionOct 5, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE OF THE DEVICE WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED WITH FORCE FROM THE PATIENT'S BREAST. A COAXIAL WAS USED AND PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17875601 ↗MalfunctionOct 5, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE OF THE DEVICE WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED WITH FORCE FROM THE PATIENT'S BREAST. A COAXIAL WAS USED AND PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17875217 ↗MalfunctionOct 5, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE OF THE DEVICE WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED WITH FORCE FROM THE PATIENT'S BREAST. A COAXIAL WAS USED AND PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17875138 ↗MalfunctionMar 25, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE INNER NEEDLE ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY REMOVED BY EXCESSIVE FORCE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16616734 ↗MalfunctionApr 14, 2022Failure to Fire; Detachment of Device or Device Component; Failure to Fire; Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO FIRE. A COAXIAL NEEDLE WAS USED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14110439 ↗
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page