MISSION recallZ-0643-2024 · Bard Peripheral Vascular Inc · initiated Nov 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

FDA root cause category: Process control

What the company is doing

On 11/8/23, recall notices were distributed to customers and distributors who were asked to do the following: 1) Discontinue use of affected product, and immediately dispose of all affected product remaining in your possession in accordance with your local facility's process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this recall. If you purchased this product from a distributor, contact your distributor for further instructions and credit resolution. - Distributors were asked to Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this expanded field action notification. 3) Complete and return the attached Customer Response Form via email to [email protected] Customers requiring additional assistance can contact the firm's North American Regional Complaint Center by phone: 1-844-823-5433, say "recall" when prompted M-F 8am - 5pm CT, or via email: [email protected]

Recall details

Recall number
Z-0643-2024
Status
Open, Classified
Initiated by the company
Nov 8, 2023
Posted by the FDA
Jan 5, 2024
Recalling firm
Bard Peripheral Vascular Inc
Quantity
1,243,937
Product code
KNW
FDA submissions
K171953
Distribution
Worldwide - US Nationwide distribution including in the states of PR, IL, MA, GA, KY, CA, TX, PA, IA, MN, NH, MD, FL, ID, IN, AL, VA, CO, WI, NY, NV, OK, OH, WA, MI, MS, AZ, HI, SD, NJ, NE, MT, NC, DC, CT, WY, UT, TN, LA, NM, SC, VT, MO, ME, AR, OR, KS, RI, AK, WV, DE, ND, PR, GU and the countries of CA, BE, AU, KR, SG, CN, TH, HK, VI, VN, AR, JP, CL, IN, TW, CO, MX, DO, BR, BD, MM, PK, NP, PA.

Product description

Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G x 25CM; 2010MSK/ Mission Kit 20G x 10CM; 2016MSK/ Mission Kit 20G x 16CM; 2020MSK/ Mission Kit 20G x 20CM

Code information (models, lots, serial numbers)

Show the full code information (16k characters)

REF/UDI-DI/Lot (Expiration Date): 1410MSK/ 00801741142543/ 0001418630(05/28/24), 0001429851(07/28/24), 0001433778(08/28/24), 0001441794(10/28/24), 0001443076(10/28/24, 0001446454(11/28/24, 0001451131(01/28/25, 0001452840(01/28/25, 0001463142(03/28/25, 0001468867(04/28/25, 0001475757(06/28/25, 0001481508(07/28/25, 0001485193(08/28/25, 0001487094(08/28/25, 0001488512(08/28/25, 0001491080(09/28/25, 0001491583(09/28/25, 0001498859(11/28/25, 0001502349(12/28/25); 1416MSK/ 00801741142550/0001402196(01/28/24), 0001423061(06/28/24), 0001430525(07/28/24), 0001434059(08/28/24), 0001435619(09/28/24), 0001452376(01/28/25), 0001458437(02/28/25), 0001464693(03/28/25), 0001464694(03/28/25); 1610MSK/ 00801741142567/0001390419(10/28/23), 0001392740(10/28/23), 0001402197(01/28/24), 0001415815(04/28/24), 0001418632(05/28/24), 0001426735(06/28/24), 0001426839(07/28/24), 0001430408(07/28/24), 0001430528(07/28/24), 0001433779(08/28/24), 0001468480(04/28/25), 0001468479(04/28/25), 0001474221(05/28/25), 0001478981(06/28/25), 0001481036(07/28/25), 0001481035(07/28/25), 0001480804(07/28/25), 0001480805(07/28/25), 0001485185(08/28/25), 0001487093(08/28/25), 0001486540(08/28/25), 0001491078(09/28/25), 0001502344(12/28/25); 1616MSK/ 00801741142574/0001410073(02/28/24), 0001418633(05/28/24), 0001429719(07/28/24), 0001432764(08/28/24), 0001441990(10/28/24), 0001445629(11/28/24), 0001451613(01/28/25), 0001468728(04/28/25), 0001474220(05/28/25), 0001474506(05/28/25), 0001476311(06/28/25), 0001480541(07/28/25), 0001480542(07/28/25), 0001498110(11/28/25); 1810MSK/ 00801741097058/0001390415(10/28/23), 0001390414(10/28/23), 0001392714(10/28/23), 0001393862(10/28/23), 0001394226(10/28/23), 0001394535(10/28/23), 0001398597(12/28/23), 0001398596(12/28/23), 0001399830(12/28/23), 0001398961(12/28/23), 0001400410(12/28/23), 0001400409(12/28/23), 0001402163(01/28/24), 0001402164(01/28/24), 0001402165(01/28/24), 0001411232(03/28/24), 0001412043(03/28/24), 0001412042(03/28/24), 0001412044(03/28/24), 0001411871(03/28/24), 0001411233(03/28/24), 0001415831(04/28/24), 0001415832(04/28/24), 0001415833(04/28/24), 0001418634(05/28/24), 0001418637(05/28/24), 0001418638(05/28/24), 0001418639(05/28/24), 0001418635(05/28/24), 0001423157(06/28/24), 0001423153(06/28/24), 0001423889(06/28/24), 0001423818(06/28/24), 0001425382(06/28/24), 0001425996(06/28/24), 0001427153(07/28/24), 0001427154(07/28/24), 0001427151(07/28/24), 0001427971(07/28/24), 0001428062(07/28/24), 0001428063(07/28/24), 0001428595(07/28/24), 0001428594(07/28/24), 0001428593(07/28/24), 0001432766(08/28/24), 0001432765(08/28/24), 0001435553(09/28/24), 0001435552(09/28/24), 0001436593(09/28/24), 0001436594(09/28/24), 0001436596(09/28/24), 0001437223(09/28/24), 0001439097(10/28/24), 0001439098(10/28/24), 0001440545(10/28/24), 0001440550(10/28/24), 0001440546(10/28/24), 0001440547(10/28/24), 0001440548(10/28/24), 0001441274(10/28/24), 0001441333(10/28/24), 0001441334(10/28/24), 0001442964(10/28/24), 0001442322(10/28/24), 0001444346(11/28/24), 0001444344(11/28/24), 0001444348(11/28/24), 0001444345(11/28/24), 0001445638(11/28/24), 0001444347(11/28/24), 0001448511(12/28/24), 0001448510(12/28/24), 0001448509(12/28/24), 0001448508(12/28/24), 0001448746(12/28/24), 0001448745(12/28/24), 0001450208(12/28/24), 0001450000(12/28/24), 0001452266(01/28/25), 0001452380(01/28/25), 0001452267(01/28/25), 0001457033(02/28/25), 0001457034(02/28/25), 0001458731(02/28/25), 0001458730(02/28/25), 0001459349(02/28/25), 0001458728(02/28/25), 0001458729(02/28/25), 0001460530(03/28/25), 0001465424(04/28/25), 0001465423(04/28/25), 0001465703(04/28/25), 0001465704(04/28/25), 0001466004(04/28/25), 0001466003(04/28/25), 0001466075(04/28/25), 0001465493(04/28/25), 0001465883(04/28/25), 0001465491(04/28/25), 0001465492(04/28/25), 0001467088(04/28/25), 0001465882(04/28/25), 0001467181(04/28/25), 0001467180(04/28/25), 0001468726(04/28/25), 0001470276(05/28/25), 0001470007(05/28/25), 0001469995(05/28/25), 0001473075(05/28/25), 0001473604(05/28/25), 0001470254(05/28/25), 0001473043(05/28/25), 0001473602(05/28/25), 0001473603(05/28/25), 0001474075(05/28/25), 0001475710(06/28/25), 0001477992(06/28/25), 0001477987(06/28/25), 0001477993(06/28/25), 0001477990(06/28/25), 0001475711(06/28/25), 0001477985(06/28/25), 0001478765(06/28/25), 0001478764(06/28/25), 0001478960(06/28/25), 0001478959(06/28/25), 0001479238(06/28/25), 0001478581(06/28/25), 0001481263(07/28/25), 0001481261(07/28/25), 0001482202(07/28/25), 0001481262(07/28/25), 0001482322(07/28/25), 0001483661(07/28/25), 0001483662(07/28/25), 0001486541(08/28/25), 0001487721(08/28/25), 0001487089(08/28/25), 0001487722(08/28/25), 0001487720(08/28/25), 0001488976(08/28/25), 0001488974(08/28/25), 0001488975(08/28/25), 0001486542(08/28/25), 0001490113(09/28/25), 0001490114(09/28/25), 0001490116(09/28/25), 0001490112(09/28/25), 0001490117(09/28/25), 0001490115(09/28/25), 0001490739(09/28/25), 0001490738(09/28/25), 0001491038(09/28/25), 0001490742(09/28/25), 0001491039(09/28/25), 0001491021(09/28/25), 0001490741(09/28/25), 0001491022(09/28/25), 0001492549(09/28/25), 0001494131(10/28/25), 0001494132(10/28/25), 0001494195(10/28/25), 0001494843(10/28/25), 0001494841(10/28/25), 0001494844(10/28/25), 0001494842(10/28/25), 0001497923(10/28/25), 0001497741(10/28/25), 0001497740(10/28/25), 0001497925(10/28/25), 0001497739(10/28/25), 0001497924(11/28/25), 0001499062(11/28/25), 0001501269(12/28/25), 0001501273(12/28/25), 0001501270(12/28/25), 0001501272(12/28/25), 0001501263(12/28/25), 0001501267(12/28/25), 0001501271(12/28/25), 0001501264(12/28/25), 0001501268(12/28/25), 0001501265(12/28/25), 0001501266(12/28/25), 0001503385(12/28/25), 0001503382(12/28/25), 0001503384(12/28/25); 1816MSK/00801741097065/ 0001390396(10/28/23), 0001390397(10/28/23), 0001390398(10/28/23), 0001392710(10/28/23), 0001393863(10/28/23), 0001394534(10/28/23), 0001396998(11/28/23), 0001396999(11/28/23), 0001398599(12/28/23), 0001398919(12/28/23), 0001398598(12/28/23), 0001400412(12/28/23), 0001400411(12/28/23), 0001402168(01/28/24), 0001402166(01/28/24), 0001402167(01/28/24), 0001410074(02/28/24), 0001411878(03/28/24), 0001411882(03/28/24), 0001411879(03/28/24), 0001412690(03/28/24), 0001412691(03/28/24), 0001412693(03/28/24), 0001413032(03/28/24), 0001413033(03/28/24), 0001415838(04/28/24), 0001415839(04/28/24), 0001415835(04/28/24), 0001415841(04/28/24), 0001415837(04/28/24), 0001415840(04/28/24), 0001418554(05/28/24), 0001418649(05/28/24), 0001418644(05/28/24), 0001418646(05/28/24), 0001418645(05/28/24), 0001418648(05/28/24), 0001418647(05/28/24), 0001418643(05/28/24), 0001422221(05/28/24), 0001422328(05/28/24), 0001423158(06/28/24), 0001423160(06/28/24), 0001423159(06/28/24), 0001425987(06/28/24), 0001425990(06/28/24), 0001425989(06/28/24), 0001427603(07/28/24), 0001427607(07/28/24), 0001427610(07/28/24), 0001427617(07/28/24), 0001427609(07/28/24), 0001428585(07/28/24), 0001428587(07/28/24), 0001428586(07/28/24), 0001429849(07/28/24), 0001430932(07/28/24), 0001429850(07/28/24), 0001431200(08/28/24), 0001431199(08/28/24), 0001436070(09/28/24), 0001437562(09/28/24), 0001437563(09/28/24), 0001439251(10/28/24), 0001440554(10/28/24), 0001440555(10/28/24), 0001440557(10/28/24), 0001439252(10/28/24), 0001440552(10/28/24), 0001441545(10/28/24), 0001441546(10/28/24), 0001441548(10/28/24), 0001441547(10/28/24), 0001442311(10/28/24), 0001442308(10/28/24), 0001442310(10/28/24), 0001442307(10/28/24), 0001443670(11/28/24), 0001443662(11/28/24), 0001443660(11/28/24), 0001443661(11/28/24), 0001443659(11/28/24), 0001443658(11/28/24), 0001460536(03/28/25), 0001460525(03/28/25), 0001462195(03/28/25), 0001460526(03/28/25), 0001462867(03/28/25), 0001460521(03/28/25), 0001462866(03/28/25), 0001463988(03/28/25), 0001463586(03/28/25), 0001463987(03/28/25), 0001465104(03/28/25), 0001467002(04/28/25), 0001467001(04/28/25), 0001468229(04/28/25), 0001468230(04/28/25), 0001468228(04/28/25), 0001468231(04/28/25), 0001469728(04/28/25), 0001469729(04/28/25), 0001469758(04/28/25), 0001469759(04/28/25), 0001474077(05/28/25), 0001474076(05/28/25), 0001474218(05/28/25), 0001474573(05/28/25), 0001474572(05/28/25), 0001474768(05/28/25), 0001474767(05/28/25), 0001474980(05/28/25), 0001476218(06/28/25), 0001477154(06/28/25), 0001477155(06/28/25), 0001477156(06/28/25), 0001481268(07/28/25), 0001481270(07/28/25), 0001481523(07/28/25), 0001481520(07/28/25), 0001481269(07/28/25), 0001481521(07/28/25), 0001481522(07/28/25), 0001485403(08/28/25), 0001485404(08/28/25), 0001485405(08/28/25), 0001485406(08/28/25), 0001486548(08/28/25), 0001486547(08/28/25), 0001486910(08/28/25), 0001487344(08/28/25), 0001488715(08/28/25), 0001489811(09/28/25), 0001489812(09/28/25), 0001490109(09/28/25), 0001490110(09/28/25), 0001490108(09/28/25), 0001490107(09/28/25), 0001491313(09/28/25), 0001491314(09/28/25), 0001491315(09/28/25), 0001491774(09/28/25), 0001491316(09/28/25), 0001491773(09/28/25), 0001493542(09/28/25), 0001493543(09/28/25), 0001493686(09/28/25), 0001493685(09/28/25), 0001493818(10/28/25), 0001494200(10/28/25), 0001494203(10/28/25), 0001494204(10/28/25), 0001494201(10/28/25), 0001494836(10/28/25), 0001494202(10/28/25), 0001494838(10/28/25), 0001494198(10/28/25), 0001494196(10/28/25), 0001494837(10/28/25), 0001494197(10/28/25), 0001494199(10/28/25), 0001495250(10/28/25), 0001494835(10/28/25), 0001495244(10/28/25), 0001495251(10/28/25), 0001495247(10/28/25), 0001495248(10/28/25), 0001495253(10/28/25), 0001495245(10/28/25), 0001495971(10/28/25), 0001496079(10/28/25), 0001496078(10/28/25), 0001495252(10/28/25), 0001495973(10/28/25), 0001495970(10/28/25), 0001496387(10/28/25), 0001495972(10/28/25), 0001495969(10/28/25), 0001496384(10/28/25), 0001496389(10/28/25), 0001496386(10/28/25), 0001496565(10/28/25), 0001497084(10/28/25), 0001496954(10/28/25), 0001496956(10/28/25), 0001496385(10/28/25), 0001496955(10/28/25), 0001496566(10/28/25), 0001497352(11/28/25), 0001497353(11/28/25), 0001498289(11/28/25), 0001498290(11/28/25), 0001498291(11/28/25), 0001499112(11/28/25), 0001499666(11/28/25), 0001499665(11/28/25), 0001499111(11/28/25), 0001499109(11/28/25), 0001499108(11/28/25), 0001499745(11/28/25), 0001499746(11/28/25), 0001499748(11/28/25), 0001499742(11/28/25), 0001500140(11/28/25), 0001500386(11/28/25), 0001500385(11/28/25), 0001499744(11/28/25), 0001500888(11/28/25), 0001500890(11/28/25), 0001500891(11/28/25), 0001501017(11/28/25),0001501018(11/28/25), 0001501247(12/28/25), 0001447510(12/28/24), 0001447511(12/28/24), 0001447513(12/28/24), 0001447512(12/28/24), 0001448119(12/28/24), 0001448118(12/28/24), 0001448121(12/28/24), 0001448120(12/28/24), 0001448743(12/28/24), 0001450414(12/28/24), 0001450413(12/28/24), 0001451076(01/28/25), 0001451075(01/28/25), 0001451625(01/28/25), 0001451626(01/28/25), 0001451621(01/28/25), 0001451622(01/28/25), 0001453517(01/28/25), 0001453516(01/28/25), 0001454295(01/28/25), 0001457030(02/28/25), 0001456850(02/28/25), 0001456856(02/28/25); 1820MSK/ 00801741097072/0001390395(10/28/23), 0001393302(10/28/23), 0001395412(11/28/23), 0001399924(12/28/23), 0001398970(12/28/23), 0001402194(01/28/24), 0001412041(03/28/24), 0001412040(03/28/24), 0001414697(04/28/24), 0001414698(04/28/24), 0001415131(04/28/24), 0001415130(04/28/24), 0001418624(05/28/24), 0001422573(05/28/24), 0001422798(06/28/24), 0001424731(06/28/24), 0001435536(09/28/24), 0001435620(09/28/24), 0001439254(10/28/24), 0001439255(10/28/24), 0001439256(10/28/24), 0001443406(11/28/24), 0001443409(11/28/24), 0001449979(12/28/24), 0001458050(02/28/25), 0001458051(02/28/25), 0001461273(03/28/25), 0001464301(03/28/25), 0001464468(03/28/25), 0001465731(04/28/25), 0001476033(06/28/25), 0001476034(06/28/25), 0001476035(06/28/25), 0001477256(06/28/25), 0001480266(07/28/25), 0001480368(07/28/25), 0001480369(07/28/25), 0001487716(08/28/25), 0001488502(08/28/25), 0001487718(08/28/25), 0001487719(08/28/25), 0001489904(09/28/25), 0001489890(09/28/25), 0001489891(09/28/25), 0001493060(09/28/25), 0001490111(09/28/25), 0001496111(10/28/25), 0001496110(10/28/25), 0001498327(11/28/25), 0001498326(11/28/25), 0001500878(11/28/25), 0001502903(12/28/25), 0001504427(12/28/25); 1825MSK/ 00801741142581/0001440070(10/28/24), 0001443305(11/28/24), 0001455151(01/28/25), 0001480267(07/28/25), 0001498155(11/28/25), 0001502905(12/28/25); 2010MSK/00801741097089/ 0001390416(10/28/23), 0001392742(10/28/23), 0001395411(11/28/23), 0001398976(12/28/23), 0001402093(01/28/24), 0001414353(03/28/24), 0001414355(03/28/24), 0001412036(03/28/24), 0001414683(04/28/24), 0001415823(04/28/24), 0001415822(04/28/24), 0001418626(05/28/24), 0001420606(05/28/24), 0001423813(06/28/24), 0001427149(07/28/24), 0001430092(07/28/24), 0001430336(07/28/24), 0001430524(07/28/24), 0001432741(08/28/24), 0001437858(09/28/24), 0001442202(10/28/24), 0001445947(11/28/24), 0001445948(11/28/24), 0001445945(11/28/24), 0001448749(12/28/24), 0001448750(12/28/24), 0001448747(12/28/24), 0001448748(12/28/24)0001449432(12/28/24), 0001449433(12/28/24), 0001456597(02/28/25), 0001456153(02/28/25), 0001456573(02/28/25), 0001456936(02/28/25), 0001456596(02/28/25),0001459554(02/28/25),0001461729(03/28/25), 0001461728(03/28/25), 0001462125(03/28/25), 0001475009(05/28/25), 0001475008(05/28/25), 0001478968(06/28/25), 0001478963(06/28/25), 0001479643(06/28/25), 0001479644(06/28/25), 0001482928(07/28/25), 0001482931(07/28/25), 0001482930(07/28/25), 0001482932(07/28/25), 0001482929(07/28/25), 0001482934(07/28/25), 0001488082(08/28/25), 0001487710(08/28/25), 0001487711(08/28/25), 0001487602(08/28/25), 0001488510(08/28/25), 0001488081(08/28/25), 0001490705(09/28/25), 0001490708(09/28/25), 0001490706(09/28/25), 0001498139(11/28/25), 0001498140(11/28/25), 0001503043(12/28/25), 0001503044(12/28/25), 0001504002(12/28/25), 0001503958(12/28/25), 0001445946(11/28/24); 2016MSK/ 00801741097096/0001390417(10/28/23), 0001390418(10/28/23), 0001392738(10/28/23), 0001394806(11/28/23), 0001395410(11/28/23), 0001398980(12/28/23), 0001398979(12/28/23), 0001401523(12/28/23), 0001401520(12/28/23), 0001402087(01/28/24), 0001402086(01/28/24), 0001415824(04/28/24), 0001415826(04/28/24), 0001418627(05/28/24), 0001418628(05/28/24), 0001423816(06/28/24), 0001423815(06/28/24), 0001428803(07/28/24), 0001428592(07/28/24), 0001429954(07/28/24), 0001430089(07/28/24), 0001430339(07/28/24), 0001432740(08/28/24), 0001432739(08/28/24), 0001434756(08/28/24), 0001437864(09/28/24), 0001441795(10/28/24), 0001441796(10/28/24), 0001442314(10/28/24), 0001442315(10/28/24), 0001442313(10/28/24), 0001442312(10/28/24), 0001445803(11/28/24), 0001445944(11/28/24), 0001445802(11/28/24), 0001445801(11/28/24), 0001448753(12/28/24), 0001448752(12/28/24),0001450642(12/28/24), 0001450646(12/28/24), 0001450644(12/28/24), 0001453528(01/28/25), 0001453530(01/28/25), 0001453529(01/28/25), 0001454098(01/28/25), 0001456155(02/28/25), 0001456154(02/28/25), 0001457603(02/28/25), 0001457602(02/28/25), 0001458843(02/28/25), 0001469201(04/28/25), 0001469202(04/28/25), 0001469200(04/28/25), 0001469577(04/28/25), 0001475585(05/28/25), 0001475411(05/28/25), 0001475410(05/28/25), 0001475412(05/28/25), 0001476363(06/28/25), 0001476364(06/28/25), 0001478962(06/28/25), 0001478961(06/28/25), 0001480801(07/28/25), 0001483581(07/28/25), 0001483582(07/28/25), 0001486538(08/28/25), 0001487713(08/28/25), 0001486539(08/28/25), 0001489271(08/28/25), 0001489612(08/28/25), 0001487714(08/28/25), 0001492480(09/28/25), 0001492475(09/28/25), 0001492477(09/28/25), 0001492478(09/28/25), 0001492479(09/28/25), 0001492476(09/28/25), 0001495451(10/28/25), 0001497095(10/28/25), 0001497094(10/28/25), 0001497674(10/28/25), 0001497354(10/28/25), 0001498772(11/28/25), 0001499753(11/28/25), 0001499751(11/28/25), 0001499752(11/28/25), 0001500387(11/28/25), 0001503045(12/28/25), 0001503830(12/28/25), 0001503828(12/28/25), 0001503046(12/28/25), 0001503829(12/28/25), 0001458844(02/28/25); 2020MSK/ 00801741097102/0001412035(03/28/24), 0001414700(04/28/24), 0001415128(04/28/24), 0001422837(06/28/24), 0001425299(06/28/24), 0001425389(06/28/24), 0001435087(09/28/24), 0001439257(10/28/24), 0001443304(11/28/24), 0001451614(01/28/25), 0001463280(03/28/25), 0001464594(03/28/25), 0001469727(04/28/25), 0001469431(04/28/25), 0001469767(04/28/25), 0001469768(04/28/25), 0001471378(05/28/25), 0001487723(08/28/25), 0001491546(09/28/25), 0001492487(09/28/25), 0001494059(09/28/25), 0001499269(11/28/25), 0001504259(12/28/25)