MISSIONRecalls & reported problems across all models

Bard Peripheral Vascular, Inc. · 20 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one MISSION model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
216
Deaths reported
1
Injuries reported
37
Recalls
3

We found 20 models of MISSION by Bard Peripheral Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 216 adverse event reports in the FDA's MAUDE database, received between Mar 14, 2017 and Sep 29, 2026 — classified as 1 death, 37 injuries, 178 malfunctions.

The most frequently reported specific device problem is “Break” (53 reports).

3 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 5 reportsDec 2023: 9 reportsJan 2024: 10 reportsFeb 2024: 5 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 27 reportsAug 2025: 9 reportsSep 2025: 4 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 2 reportsFeb 2026: 14 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202627
Last 36 months. The most recent month may be incomplete.

MISSION recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Nov 8, 2023Open, Classified

    The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

    Z-0643-2024 · Root cause: Process control · names 11 of these models · Recall details · FDA record ↗

  2. Apr 11, 2023Open, Classified

    Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.

    Z-1564-2023 · Root cause: Mixed-up of materials/components · names 2 of these models · Recall details · FDA record ↗

  3. Feb 28, 2023Open, Classified

    The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.

    Z-1260-2023 · Root cause: Mixed-up of materials/components · names 2 of these models · Recall details · FDA record ↗

Most reported MISSION problems

Device problems

  • Break53
  • Device Markings/Labelling Problem44
  • Material Twisted/Bent44
  • Difficult to Remove40
  • Detachment of Device or Device Component30
  • Device-Device Incompatibility24
  • Device Contamination with Chemical or Other Material11
  • Adverse Event Without Identified Device or Use Problem10

Patient problems

  • No Clinical Signs, Symptoms or Conditions161
  • No Consequences Or Impact To Patient12
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available11
  • Foreign Body In Patient9
  • Hemorrhage/Blood Loss/Bleeding5
  • Injury5
  • Pneumothorax5
  • Device Embedded In Tissue or Plaque4

MISSION models

20 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MISSION adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Bard Peripheral Vascular, Inc. products

All Bard Peripheral Vascular, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page