MISSION 1610MSKRecalls & reported problems
Bard Peripheral Vascular, Inc. · Mission Disposable Core Biopsy Instrument Kit - 16Gx10cm · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741142567), received between Jun 12, 2019 and Feb 5, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2023Open, Classified
The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.
Z-0643-2024 · Root cause: Process control · FDA record ↗
- Feb 28, 2023Open, Classified
The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.
Z-1260-2023 · Root cause: Mixed-up of materials/components · FDA record ↗
Most reported problems
Device problems
- Break1
- Device Markings/Labelling Problem1
- Difficult to Remove1
- Material Twisted/Bent1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Injury1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2024Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY THROUGH NORMAL DENSITY TISSUE, THE END OF INNER NEEDLE; NEEDLE NOTCH WAS ALLEGEDLY BENT WHILE PENETRATING AND THE CUTTING DIDN'T GO ALL THE WAY. FURTHER IT WAS REPORTED THAT THE NEEDLE WAS ALLEGEDLY PULLED OUT OF THE PATIENT'S BREAST WITH FORCE. A COAXIAL WAS USED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18641080 ↗MalfunctionJan 9, 2024Unsealed Device Packaging; Unsealed Device Packaging
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE PACKAGING WAS ALLEGEDLY OPEN WITHOUT SEALING. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18473178 ↗MalfunctionFeb 14, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT PRIOR TO A PROCEDURE, THE DEVICE PACKAGE WAS ALLEGEDLY MISLABELED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13520740 ↗InjuryJun 12, 2019Break; Break
IT WAS REPORTED THAT DURING A TIBIA BONE BIOPSY, THE DEVICE ALLEGEDLY BROKE INSIDE THE PATIENT. THE NEEDLE WAS REMOVED FROM THE PATIENT 2 WEEKS POST PROCEDURE.
Patient problems: Injury
MAUDE report 8691245 ↗
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.