ALTRUA™ 20 S205Recalls & reported problems

Boston Scientific Corporation · Pacemaker · Class 3 · Implant

All models of ALTRUA 20 (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00802526499210), received between Oct 12, 2016 and Oct 13, 2024 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing4
  • Signal Artifact/Noise3
  • Pacing Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Ambient Noise Problem1
  • Battery Problem1
  • High Capture Threshold1
  • Telemetry Discrepancy1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Overstimulation of Tissue1
  • No Consequences Or Impact To Patient1
  • Pocket Erosion1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 13, 2024Telemetry Discrepancy; Telemetry Discrepancy

    IT WAS REPORTED THAT THIS PACEMAKER COULD NOT BE INTERROGATED BY DIFFERENT PROGRAMMERS. DIFFERENT TROUBLESHOOTING OPTIONS WERE DISCUSSED WITH THE DIFFERENT PROGRAMMERS AND STILL THE DEVICE COULD NOT BE INTERROGATED. AN X RAY WAS COMPLETED TO CONFIRM THE DEVICE WAS STILL IMPLANTED. IT WAS RECOMMENDED TO CONSIDER THE PATIENT UNPROTECTED AND REPLACE DEVICE. THE PACEMAKER REMAINS IN SERVICE AT THIS TIME. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20436512 ↗
  • MalfunctionJul 11, 2023Signal Artifact/Noise; Over-Sensing; High Capture Threshold; Signal Artifact/Noise; Over-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THIS PACEMAKER EXHIBITED NOISE AND OVERSENSING ON THE VENTRICULAR CHANNEL. HIGH CAPTURE THRESHOLDS WERE NOTED AS WELL. THE INVOLVED RIGHT VENTRICULAR (RV) LEAD IS A NON BOSTON SCIENTIFIC PRODUCT. THE PATIENT REPORTED EXPERIENCING POCKET STIMULATION. A FOLLOW UP WITH CARDIOLOGY WAS SCHEDULED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. AT THIS TIME, THIS DEVICE SYSTEM REMAINS IN SERVICE.

    Patient problems: Device Overstimulation of Tissue

    MAUDE report 17302155 ↗
  • InjuryJul 13, 2022Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing

    IT WAS REPORTED THAT THIS PACEMAKER SHOWED AN EVENT WERE NOISY SIGNALS WERE OVERSENSED. THERE WAS NO MALFUNCTION RELATED TO THE SYSTEM LEADS. THESE LEADS WERE UNIPOLAR. A NEW SYSTEM WAS IMPLANTED ON THE OPPOSITE SIDE WITHOUT ANY PROBLEMS. THE EXISTING SYSTEM WAS NOT REMOVED AND REMAINS IMPLANTED. THE SENSE AND PACE RATE WAS REPROGRAMMED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14999184 ↗
  • MalfunctionApr 22, 2022Signal Artifact/Noise; Over-Sensing; Pacing Problem; Battery Problem; Signal Artifact/Noise; Over-Sensing; Pacing Problem; Battery Problem

    IT WAS REPORTED THAT THIS PACEMAKER AND A NON-BOSTON SCIENTIFIC RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING OF NOISE RESULTING IN AN UNKNOWN DURATION OF PACING INHIBITION, AND ATRIAL TACHY RESPONSE (ATR) MODE SWITCHES. THE PATIENT WAS NOTED TO BE RA AND RV PACEMAKER DEPENDENT. PROGRAMMING CHANGES WERE MADE. SUBSEQUENTLY, THE BATTERY STATUS WAS NOTED TO HAVE INCREASED FROM LESS THAN 0.5 YEARS REMAINING TO 1 YEAR REMAINING. TECHNICAL SERVICES (TS) DISCUSSED THE INCREASE IN LONGEVITY WAS LIKELY A RESULT OF THE PROGRAMMING CHANGES THAT WERE MADE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14182138 ↗
  • InjuryNov 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS PACEMAKER WAS PART OF A SYSTEM REVISION DUE TO SKIN INFECTION WHICH LED TO POCKET EROSION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THE PACEMAKER WAS EXPLANTED.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 12749680 ↗
  • InjuryOct 12, 2016Over-Sensing; Pacing Problem; Ambient Noise Problem; Over-Sensing; Pacing Problem; Ambient Noise Problem

    BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS PACEMAKER AND RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS FROM ANOTHER MANUFACTURER EXHIBITED NOISE AND OVERSENSING LEADING TO PACING INHIBITION. THE NOISE WAS ABLE TO BE REPRODUCED WITH ISOMETRICS. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED DECREASING SENSITIVITY IN THE ATRIUM AND VENTRICLE TO INCREASE PACING. THE SYSTEM WAS ABANDONED AND A NEW SYSTEM WAS IMPLANTED IN THE RIGHT SIDE. THE DEVICE WAS LATER EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6021733 ↗