VITALIO™ K272Recalls & reported problems
Boston Scientific Corporation · Pacemaker · Class 3 · Implant
All models of VITALIO (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00802526516955), received between Apr 22, 2016 and Jun 25, 2026 — classified as 9 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Capture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Failure to Capture2
- High impedance2
- Incorrect, Inadequate or Imprecise Result or Readings2
- Pacing Problem2
- Ambient Noise Problem1
- Device Operates Differently Than Expected1
- Device Sensing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient2
- Unspecified Infection2
- Cardiac Arrest1
- Discomfort1
- Dizziness1
- Erosion1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryJun 25, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT WITH AN IMPLANTABLE PACEMAKER STATED THE DEVICE IS DEAD AND IS CONSIDERING DEVICE REMOVAL. AS OF THIS TIME, THE DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25628241 ↗InjuryDec 2, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS PACEMAKER HAD REVERTED TO SAFETY MODE. TECHNICAL SERVICES (TS) WAS CONSULTED, DISCUSSED SAFETY MODE PARAMETERS, AND RECOMMENDED DEVICE REPLACEMENT. THIS PACEMAKER REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION RECEIVED DETAILED THIS DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694168 ↗InjuryMay 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SYSTEM WAS EXPLANTED DUE TO THE PATIENT FELT DISCOMFORT. ACCORDING TO THE PATIENT HE FELT THE DEVICE AROUND HIS ABDOMEN MORE THAN BEFORE. A NEW LEADLESS SYSTEM WAS SUCCESSFULLY IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Discomfort
MAUDE report 19270280 ↗MalfunctionNov 29, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PATIENT WITH THIS PACEMAKER DEVICE WAS FOUND IN SAFETY MODE. TECHNICAL SERVICES (TS) RECOMMENDED TO HAVE THE DEVICE REPLACED. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18230077 ↗MalfunctionSep 20, 2023Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THIS PATIENT WITH THIS PACEMAKER DEVICE EXHIBITED T WAVE OVERSENSING SINCE LAST FEW MONTHS NOW. THE PATIENT HAD HISTORY OF RAPID VENTRICULAR RESPONSE (RVR) AND NON-SUSTAINED EPISODES WERE NOTED ON THE PRESENTING. UPON FURTHER REVIEW OF THE EPISODES, IT LOOKED LIKE INTERMITTENT T WAVE OVERSENSING. TECHNICAL SERVICES (TS) DISCUSSED OPTIONS FOR OPTIMIZING SENSITIVITY, RETRACTOR AS MUCH AS POSSIBLE TO MINIMIZE TWOS AND OPTION OF ADJUSTING TACHY DETECTION AS APPROPRIATE. THIS DEVICE REMAINS IN SERVICE, NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17780863 ↗InjuryJul 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS PACEMAKER DEVICE UNDERWENT A REVISION. THE PATIENT WAS FEELING THE EDGE OF THE DEVICE. MOREOVER, THE DEVICE WAS REPOSITIONED FROM SUBCUTANEOUS TO SUBMUSCULAR. THIS DEVICE REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12243467 ↗MalfunctionJan 22, 2019High impedance; Pacing Problem; High impedance; Pacing Problem
IT WAS REPORTED THAT THIS PACEMAKER EXHIBITED HIGH, OUT-OF-RANGE PACING IMPEDANCE ON THE RIGHT VENTRICULAR (RV) CHANNEL WHICH CAUSED A LEAD SAFETY SWITCH. PACING INHIBITION OF GREATER THAN TWO SECONDS WAS ALSO NOTED. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IN SERVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8268539 ↗InjuryOct 22, 2018Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER WAS PART OF A SYSTEM REVISION DUE TO INFECTION WITH SEPSIS. THE PACEMAKER WAS EXPLANTED AND THEN WAS USED AS AN EXTERNAL TEMPORARY PACEMAKER. SUBSEQUENTLY, THE PACEMAKER WAS REMOVED FROM SERVICE WHEN A NEW SYSTEM WAS IMPLANTED. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THE PACEMAKER WAS EXPLANTED.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 7989902 ↗MalfunctionOct 9, 2017High impedance; Device Sensing Problem; High impedance; Device Sensing Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER EXHIBITED VARYING IMPEDANCES BETWEEN 600-700 OHMS AND 1,500 OHMS. THE THRESHOLDS ARE STABLE; HOWEVER, THE P-WAVE AMPLITUDES ARE NOT AS EXPECTED. IT WAS STATED THAT THE PATIENT WILL CONTINUE TO BE FOLLOWED CLOSELY. THE PATIENT'S RIGHT ATRIAL (RA) LEAD IS A NON-BOSTON SCIENTIFIC PRODUCT. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS PRODUCT REMAINS IN-SERVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6932138 ↗InjuryApr 19, 2017Premature Discharge of Battery; Low impedance; Premature Discharge of Battery; Low impedance
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE AND ASSOCIATED LEAD DISPLAYED LOW PACE IMPEDANCE MEASUREMENTS. THE DEVICE WAS NOTED TO HAVE EXPERIENCED PREMATURE BATTERY DEPLETION. THE PATIENT WAS SEEN FOR A REVISION PROCEDURE WHERE THE DEVICE WAS EXPLANTED AND THE LEAD WAS ABANDONED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6503982 ↗InjuryJul 16, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS PART OF A SYSTEM REVISION DUE TO INFECTION AND EROSION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THE PRODUCT WAS EXPLANTED.
Patient problems: Erosion; Unspecified Infection
MAUDE report 5800152 ↗InjuryJul 14, 2016Failure to Capture; Failure to Capture
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE. AN INVASIVE PROCEDURE WAS PERFORMED. THIS LEAD WAS EXPLANTED AND REPLACED. THE DEVICE REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5795685 ↗MalfunctionJul 12, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT'S REMOTE HOME MONITORING SYSTEM DISPLAYED A RED BLINKING LIGHT, INDICATIVE OF A POTENTIAL CHANGE IN THEIR DEVICE. ADDITIONAL INFORMATION HAS BEEN REQUESTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE REMAINS IN SERVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5787444 ↗InjuryApr 22, 2016Failure to Capture; Pacing Problem; Ambient Noise Problem; Failure to Capture; Pacing Problem; Ambient Noise Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER EXHIBITED NOISE AND PACING INHIBITION. THERE WAS ASYSTOLE GREATER THAN 2 SECONDS. IT WAS SUSPECTED THAT THERE WAS A LEAD FRACTURE ON THE RIGHT ATRIAL (RA) COMPETITOR LEAD. A PROCEDURE WAS PERFORMED TO EXPLANT THE LEAD AT WHICH TIME A LEAD FRACTURE WAS CONFIRMED ON THE LEAD NEAR THE POCKET. DURING THE PROCEDURE, THERE WAS PACING INHIBITION WHILE THE DEVICE WAS IN CAUTERY MODE. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THAT THE PACING INHIBITION WAS LIKELY DUE TO THE EXPOSED FRACTURE BEING NEAR THE CAUTERY. THE FRACTURED LEAD WAS SURGICALLY ABANDONED. THE DEVICE WAS ALSO EXPLANTED AND REPLACED WITH AN UPGRADED DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiac Arrest; Dizziness
MAUDE report 5601741 ↗
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page