ESSENTIO™ MRI EL DR L131Recalls & reported problems

Boston Scientific Corporation · Pacemaker · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific Essentio MRI pacemaker is an implanted device that paces the heart. Its 730 reports often describe the battery running down early and sensing or pacing problems, and others mention infections. The reports mention 1 death and 400 injuries, and reports rose from 114 in the prior 12 months to 252 in the latest 12 months. There are 2 open recalls, both about the device entering safety mode because of high battery impedance.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
774
Deaths reported
1
Injuries reported
399
Recalls naming it
2

Summary

The FDA's MAUDE database contains 774 adverse event reports that list this exact device (DI 00802526559006), received between Jan 26, 2017 and Sep 30, 2026 — classified as 1 death, 399 injuries, 374 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (293 reports).

2 FDA recall records name this device's identifier, 2 still open.

26 other recalls cover devices cleared under the same FDA submission (P150012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 7 reportsJan 2024: 11 reportsFeb 2024: 5 reportsMar 2024: 6 reportsApr 2024: 12 reportsMay 2024: 7 reportsJun 2024: 9 reportsJul 2024: 4 reportsAug 2024: 8 reportsSep 2024: 6 reportsOct 2024: 11 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 8 reportsFeb 2025: 9 reportsMar 2025: 10 reportsApr 2025: 8 reportsMay 2025: 8 reportsJun 2025: 16 reportsJul 2025: 14 reportsAug 2025: 4 reportsSep 2025: 11 reportsOct 2025: 20 reportsNov 2025: 18 reportsDec 2025: 26 reportsJan 2026: 39 reportsFeb 2026: 35 reportsMar 2026: 27 reportsApr 2026: 31 reportsMay 2026: 22 reportsJun 2026: 34 reportsJul 2026: 28 reportsAug 2026: 13 reportsSep 2026: 23 reportsOct 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2025Class IOpen, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0084-2026 · Root cause: Other · Recall details · FDA record ↗

  2. Dec 12, 2024Class IOpen, Classified

    A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

    Z-0824-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Discharge of Battery293
  • Adverse Event Without Identified Device or Use Problem153
  • Over-Sensing114
  • Pacing Problem72
  • High impedance64
  • Signal Artifact/Noise63
  • Under-Sensing49
  • Defective Device40

Patient problems

  • No Clinical Signs, Symptoms or Conditions510
  • Unspecified Infection128
  • Sepsis32
  • No Known Impact Or Consequence To Patient31
  • No Code Available21
  • Asystole16
  • Syncope14
  • Bradycardia13

Recent adverse event reports

The 25 most recent of 774 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950540 ↗
  • MalfunctionSep 30, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948933 ↗
  • InjurySep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that this pacemaker was found to be in Safety Mode. Subsequently, this device was explanted and replaced with another device. The device changeout was brought on by a surgical repair of the hip after experiencing a fall, which was determined to be mechanical related for the patient as she was not dependent. This pacemaker has not been received for investigation. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946997 ↗
  • MalfunctionSep 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937442 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923044 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918986 ↗
  • MalfunctionSep 25, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892576 ↗
  • MalfunctionSep 24, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890397 ↗
  • MalfunctionSep 24, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886924 ↗
  • MalfunctionSep 24, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. No adverse patient effects were reported. This device remains in service at this time.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885905 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this pacemaker was part of the system revision due to infection. No adverse patient effects were reported. The pacemaker was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50866136 ↗
  • MalfunctionSep 17, 2026Premature Discharge of Battery

    It was reported that this pacemaker exhibited a Code 1003 voltage alert, indicative that the voltage was too low for the projected remaining battery capacity. Technical Services (TS) reviewed the available device data and noted that the device had recently been updated to newest software update. The voltage alert was determined to have likely triggered shortly after the software update due to activation of the monitoring algorithm. Engineering analysis confirmed that remote monitoring functionality remained available and that approximately six months of RF telemetry availability remained. Continued monitoring was recommended, until radio frequency (RF) telemetry is disabled, at which time replacement should be performed. No adverse patient effects were reported. This pacemaker remains in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806043 ↗
  • MalfunctionSep 10, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50724482 ↗
  • MalfunctionSep 8, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50679870 ↗
  • InjurySep 7, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50658539 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50658382 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50658039 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50649895 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Device replacement was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50612079 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602467 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602362 ↗
  • MalfunctionSep 4, 2026Over-Sensing

    It was reported that the health care professional (HCP) called Technical Services (TS) for further review of this pacemaker stored Atrial Tachy Response (ATR) and Signal Artifact Monitor (SAM) events. Upon review, the ATR episodes showed oversensing noise on both right atrial (RA) and right ventricular (RV) leads. The oversensed noise appears to be electromagnetic interference (EMI) in nature. Further review of the SAM events shows the EMI noise led to minute ventilation (MV) to be disabled by the SAM feature. The HCP noted patient was in a procedure at the time the events were recorded. It is believed that the EMI oversensing from the electrocautery led to the inappropriate ATR and SAM events. TS discussed programming optimization recommendations. No adverse patient effects were reported. This pacemaker remains in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50601334 ↗
  • MalfunctionSep 4, 2026Over-Sensing; Pacing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50600903 ↗
  • MalfunctionAug 24, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50421670 ↗
  • MalfunctionAug 24, 2026Failure to Read Input Signal; Under-Sensing; Low impedance

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50419132 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150012. One submission often covers many models, so check the product and lot details in each record.

  • Z-1770-2026 · Mar 19, 2026 · Open, Classified
    Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL
  • Z-0089-2026 · Aug 20, 2025 · Open, Classified
    Model Number L211 PROPONENT DR SL MRI Pacemaker
  • Z-0082-2026 · Aug 20, 2025 · Open, Classified
    ACCOLADE DR SL MRI (Model L311)
  • Z-0097-2026 · Aug 20, 2025 · Open, Classified
    Model Number L131, ESSENTIO DR EL MRI Pacemaker
  • Z-0088-2026 · Aug 20, 2025 · Open, Classified
    Model Number L210 PROPONENT SR SL Pacemaker
  • Z-0095-2026 · Aug 20, 2025 · Open, Classified
    Model Number L111, ESSENTIO DR SL MRI Pacemaker
  • Z-0091-2026 · Aug 20, 2025 · Open, Classified
    Model Number L231, PROPONENT DR EL MRI Pacemaker
  • Z-0081-2026 · Aug 20, 2025 · Open, Classified
    ACCOLADE SR SL MRI (Model L310)
  • Z-0094-2026 · Aug 20, 2025 · Open, Classified
    Model Number L110, ESSENTIO SR SL MRI Pacemaker
  • Z-0823-2025 · Dec 12, 2024 · Open, Classified
    PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
  • Z-0822-2025 · Dec 12, 2024 · Open, Classified
    ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
  • Z-1999-2021 · Jun 3, 2021 · Open, Classified
    Proponent pacemaker, Models: a) L200 b) L201 c) L209 d) L210 e) L211 f) L221 g) L231
  • Z-2000-2021 · Jun 3, 2021 · Open, Classified
    Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331
  • Z-2017-2021 · Jun 3, 2021 · Open, Classified
    VITALIO DR EL (pacemaker), Model Numbers: a) J274 b) J277 c) K274 (added 7/15/2021) d) K277 (added 7/15/2021) e) K284 (added 7/15/2021) f) K287 (added 7/15/2021)
  • Z-1998-2021 · Jun 3, 2021 · Open, Classified
    Essentio pacemaker, Models: a) L100 b) L101 c) L110 d) L111 e) L121 f) L131