ALTRUA™ 2 EL DR S722Recalls & reported problems
Boston Scientific Corporation · Pacemaker · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00802526559365), received between Apr 29, 2019 and Apr 6, 2022 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- Adverse Event Without Identified Device or Use Problem1
- Interrogation Problem1
- Premature Discharge of Battery1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- Unspecified Infection1
- Unspecified Mental, Emotional or Behavioral Problem1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 6, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PATIENT IMPLANTED WITH THIS DEVICE PRESENTED TO THE EMERGENCY ROOM FEELING UNEASY. UPON INTERROGATION IT WAS IDENTIFIED THE DEVICE WAS OPERATING IN SAFETY MODE. NEW INFORMATION RECEIVED INDICATES THAT THE DEVICE WAS EXPLANTED AND WILL BE RETURNED FOR EVALUATION.
Patient problems: Unspecified Mental, Emotional or Behavioral Problem
MAUDE report 14032489 ↗InjuryJan 22, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THIS PACEMAKER WAS SUSPECTED TO BE EXHIBITING PREMATURE BATTERY DEPLETION. A SAVE TO DISK WAS SENT IN FOR ENGINEERING ANALYSIS AND IT WAS CONFIRMED THAT THERE WAS AN INCREASE IN POWER CONSUMPTION. BOSTON SCIENTIFIC TECHNICAL SERVICES RECOMMENDED DEVICE REPLACEMENT BECAUSE OF THIS ANOMALY. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IN SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13329640 ↗InjuryJan 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT IMPLANTED WITH THIS PACEMAKER EXPERIENCED AN INFECTION. THIS PACEMAKER WAS EXPLANTED AND REPLACED WITH A NEW ONE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE EXPLANTED DEVICE IS EXPECTED FOR RETURN.
Patient problems: Unspecified Infection
MAUDE report 13139212 ↗InjuryApr 14, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING A POST IMPLANT INTERROGATION, THREE DEVICE RESETS WERE EXHIBITED AND THE DEVICE CONSEQUENTLY IN SAFETY MODE. TECHNICAL SERVICES (TS) WAS CALLED TO PROVIDE GUIDANCE REGARDING THE OPERATION OF THIS UNEXPECTED PROGRAMING OPERATION. TS STATED THAT ONCE IN SAFETY MODE, THE DEVICE WOULD NEED TO BE EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE EVENTS WERE REPORTED AND REPLACEMENT LATER THAT SAME DAY WAS CONFIRMED. THE EXPLANTED DEVICE WAS RETURNED FOR ANALYSIS. ANOTHER DEVICE OF SAME MODEL TYPE WAS IMPLANTED WITHOUT FURTHER COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11668826 ↗MalfunctionApr 29, 2019Interrogation Problem; Interrogation Problem
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THIS PACEMAKER WAS CONNECTED TO THE LEADS HOWEVER THE FIELD WAS UNABLE TO ESTABLISH TELEMETRY AND INTERROGATE IT. DESPITE TROUBLESHOOTING, THE PHYSICIAN COULD NOT RESOLVE THE ISSUE SO THE PACEMAKER WAS REMOVED AND REPLACED PRIOR TO POCKET CLOSURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8562894 ↗
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page