OverStitch NXT Endoscopic Suturing System ESS-ESG-NXTRecalls & reported problems

Apollo Endosurgery, Inc. · Needle Driver and Anchor Exchange · Class 2

All models of OverStitch NXT Endoscopic Suturing System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00811955020834), received between Jul 11, 2025 and Oct 27, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Perforation1
  • Perforation of Esophagus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 27, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN OVERSTITCH NXT ENDOSCOPIC SUTURING SYSTEM WAS USED DURING AN EDG PROCEDURE ON OCTOBER 1, 2025. DURING THE PROCEDURE, THE PHYSICIAN PREPARED TO PLACE A SECOND SUTURE. TOO MUCH TISSUE WAS CAPTURED BY THE DEVICE AND THE ANCHOR EXCHANGE FAILED TO UNLOAD FROM THE NEEDLE BODY. WITH THE ANCHOR STILL IN THE PATIENT'S TISSUE, THE PHYSICIAN USED SCISSORS TO CUT THE SUTURE AND LEFT THE ANCHOR IN THE ESOPHAGEAL WALL. THE PROCEDURE WAS COMPLETED WITH MANTIS CLIPS TO FIXATE THE STENT. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23389928 ↗
  • MalfunctionSep 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN OVERSTITCH NXT ENDOSCOPIC SUTURING SYSTEM WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE AND AFTER THE FIRST SUTURE WAS COMPLETED, THE OVERSTITCH NXT TOWER SLID FORWARD AND WAS NO LONGER TOUCHING THE DISTAL END OF THE SCOPE. A NEW OVERSTITCH NXT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22954211 ↗
  • InjuryAug 22, 2025Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Difficult to Advance; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN OVERSTITCH NXT ENDOSCOPIC SUTURING SYSTEM WAS USED DURING A STENT SUTURING PROCEDURE ON (B)(6) 2025. DURING INSERTION, THE OVERSTITCH DEVICE COULD NOT BE ADVANCED INTO THE PATIENT'S ESOPHAGUS. THE PHYSICIAN ATTEMPTED A JAW THRUST AND APPLIED EXTRA LUBE TO THE DEVICE BUT WAS STILL UNSUCCESSFUL. THEN, THE PHYSICIAN ATTEMPTED TO USE AN OVERTUBE DEVICE AND SAW A PERFORATION IN THE PATIENT'S ESOPHAGUS BELIEVED TO BE CAUSED BY THE OVERSTITCH NXT AND ABORTED THE PROCEDURE. THE PATIENT WAS ADMITTED TO THE ICU. AFTER 3 NIGHTS IN THE HOSPITAL, THE PATIENT WAS DISCHARGED.

    Patient problems: Perforation

    MAUDE report 22872867 ↗
  • InjuryJul 11, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN OVERSTITCH NXT WAS USED DURING AN ENDOSCOPIC SLEEVE GASTROPLASTY PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE OVERSTITCH NXT DEVICE COULD NOT BE ADVANCED INTO THE ESOPHAGUS. A DIAGNOSTIC SCOPE WAS THEN USED, WHICH REVEALED A PERFORATION JUST BELOW THE UPPER ESOPHAGEAL SPHINCTER (UES). THE PERFORATION WAS PARTIALLY CLOSED, AND THE ORIGINAL PROCEDURE WAS CANCELLED. THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR OBSERVATION, AND TWO CT SCANS WERE PERFORMED. THE PATIENT WAS DISCHARGED ON (B)(6) 2025. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: Perforation of Esophagus

    MAUDE report 22480861 ↗