MAGEC® Spinal Bracing and Distraction System MC2-4570RRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 4.5 mm offset rod with 70 mm actuator · Class 2
All models of MAGEC Spinal Bracing and Distraction System (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00812258022617), received between Jul 30, 2020 and Jun 8, 2022 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Migration” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 1, 2022Class IIOpen, Classified
During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.
Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Migration3
- Migration or Expulsion of Device2
- Break1
- Fracture1
- Material Fragmentation1
- Naturally Worn1
- No Apparent Adverse Event1
- Scratched Material1
Patient problems
- Failure of Implant7
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 8, 2022Fracture; Material Fragmentation; Naturally Worn; Scratched Material; Fracture; Material Fragmentation; Naturally Worn; Scratched Material
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO UNEXPECTED PAIN. DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND THE BEARING MECHANISM TO BE FRACTURED AND OUT OF POSITION, AND METALLIC FRAGMENTS AROUND THE LEAD SCREW. ADDITIONALLY, WEAR, SCRATCHES, DISCOLORATION, O-RING DAMAGE AND OTHER SURFACE DAMAGE WAS NOTED. THERE WAS NO FURTHER REPORTED PATIENT IMPACT AND NO PATIENT OR USER INJURY WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 1 OF 2.
Patient problems: Pain
MAUDE report 14629729 ↗MalfunctionMar 1, 2021Break; Break
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE DUE TO A BROKEN ROD. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THERE WAS NO PROBLEM FOUND.
Patient problems: Failure of Implant
MAUDE report 11397104 ↗MalfunctionJan 11, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
INFORMATION WAS RECEIVED THAT X-RAY IMAGED REVEALED SEPARATION OF ACTUATOR AND END CAP. THERE WAS NO PATIENT INJURY REPORTED. NO REVISION SURGERY IS PLANNED AS THE SURGEON WANTS TO OBSERVE THE FUNCTIONALITY OF THE ROD.
Patient problems: Failure of Implant
MAUDE report 11151203 ↗MalfunctionDec 30, 2020No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 FOR A FINAL FUSION. THERE WAS NO PATIENT INJURY REPORTED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) THE ROD FAILED TO DISTRACT.
Patient problems: Failure of Implant
MAUDE report 11098820 ↗MalfunctionDec 29, 2020Migration; Migration
INFORMATION WAS RECEIVED THAT X-RAY IMAGES REVEALED THAT THE END CAP HAD DISENGAGED. IT IS UNKNOWN IF A REVISION HAS BEEN PLANNED AT THIS TIME. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Failure of Implant
MAUDE report 11092296 ↗MalfunctionOct 22, 2020Migration; Migration
INFORMATION WAS RECEIVED THAT X-RAY IMAGING REVEALED THAT THE ACTUATOR END CAP OFF THE OFFSET ROD CAME LOOSE. THERE WAS NO PATIENT INJURY REPORTED. NO REVISION PROCEDURE IS PLANNED AT THIS TIME.
Patient problems: Failure of Implant
MAUDE report 10722951 ↗MalfunctionOct 9, 2020Migration; Migration
SEE UPDATED INFORMATION IN H2, H3, H6 AND H10.
Patient problems: Failure of Implant
MAUDE report 10659675 ↗MalfunctionJul 30, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
INFORMATION WAS RECEIVED THAT X-RAY IMAGES REVEALED THAT THE END CAP HAD DISENGAGED. NO REVISION PROCEDURE IS PLANNED AS THE ROD IS STILL IN-SITU. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Failure of Implant
MAUDE report 10348385 ↗
Other Growing Rod System- Magnetic Actuation devices
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation112 reports
- MAGEC MC2-5090SNuvasive, Inc. · Growing rod system- magnetic actuation1 recall33 reports
- MAGEC® Spinal Bracing and Distraction System RA002-4545SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation82 reports
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLRNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation81 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall26 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall25 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page