MAGEC® Spinal Bracing and Distraction System MC2-5590SRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 5.5 mm standard rod with 90 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This NuVasive MAGEC spinal rod is a growing rod implanted to help the spine grow and straighten. The 26 reports mostly describe rods failing to lengthen, end caps working loose, or parts shifting, sometimes needing removal or revision surgery. There are no deaths and 5 injuries. One open recall, started November 2022, covers a problem where the sterile packaging barrier could fail, leaving the device non-sterile.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00812258022662), received between Sep 25, 2019 and Oct 23, 2024 — classified as 5 injuries, 21 malfunctions.

The most frequently reported specific device problem is “Migration” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2022Class IIOpen, Classified

    During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

    Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration7
  • Failure to Advance4
  • Mechanical Problem4
  • Migration or Expulsion of Device4
  • Corroded2
  • Degraded2
  • Detachment of Device or Device Component2
  • Fracture2

Patient problems

  • Failure of Implant20
  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 23, 2024Failure to Advance; Failure to Advance

    INFORMATION WAS RECEIVED THAT ONE OF THE IMPLANTED RODS FAILED TO LENGTHEN SO BOTH RODS WERE REMOVED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 20517467 ↗
  • InjuryOct 23, 2024Failure to Advance; Failure to Advance

    INFORMATION WAS RECEIVED THAT ONE OF THE IMPLANTED RODS FAILED TO LENGTHEN SO BOTH RODS WERE REMOVED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 20516591 ↗
  • InjuryAug 14, 2024Break; Failure to Advance; Break; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 19982049 ↗
  • MalfunctionJun 16, 2023Mechanical Jam; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT FORCE TESTING WAS UNABLE TO BE COMPLETED DUE TO THE EXTERNAL REMOTE CONTROLLER (ERC) NOT BEING FUNCTIONAL. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 17148921 ↗
  • MalfunctionJul 19, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE OCCURRED DUE TO THE END CAP POPPING OFF AND THE ROD NO LONGER WORKING. NO PATIENT INJURY HAS BEEN REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 15055723 ↗
  • MalfunctionMar 17, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD COULD NOT BE DISTRACTED ANYMORE DUE TO THE END CAP LOOSENING OF ACTUATOR. THE ROD WAS EXPLANTED AND A FUSION TOOK PLACE FROM T2-L4. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13791897 ↗
  • MalfunctionMar 17, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD COULD NOT BE DISTRACTED ANYMORE DUE TO THE END CAP LOOSENING OF THE ACTUATOR. THE ROD WAS EXPLANTED AND A FUSION TOOK PLACE FROM T2-L4. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13791432 ↗
  • MalfunctionSep 24, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT THE END CAP CAME OFF AND THE ROD WAS UNABLE TO DISTRACT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12528369 ↗
  • MalfunctionSep 9, 2021Corroded; Fracture; Corroded; Fracture

    NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12446879 ↗
  • MalfunctionSep 2, 2021Degraded; Degraded

    INFORMATION WAS RECEIVED THAT THE ROD PRESENTS WITH FRETTING. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12417156 ↗
  • MalfunctionSep 2, 2021Degraded; Degraded

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12414674 ↗
  • MalfunctionJul 13, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT WHILE USING ULTRA-SOUND TO MEASURE DISTRACTION LENGTH, A LARGER MEASUREMENT WAS RECORDED THAN WAS PROGRAMMED. THE SURGEON REQUESTED ANOTHER X-RAY AND UPON REVIEW, IT WAS NOTED THAT THE ENDCAP HAD SEPARATED FROM THE ACTUATOR. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12159411 ↗
  • MalfunctionJul 13, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT WHILE USING ULTRA-SOUND TO MEASURE DISTRACTION LENGTH, A LARGER MEASUREMENT WAS RECORDED THAN WHAT WAS PROGRAMMED. THE SURGEON REQUESTED ANOTHER X-RAY AND UPON REVIEW, THE ENDCAP WAS NOTED TO BE SEPARATED FROM THE ACTUATOR. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12159355 ↗
  • MalfunctionJul 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD EVIDENCE OF GALLING.

    Patient problems: Insufficient Information

    MAUDE report 12144726 ↗
  • MalfunctionApr 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    INFORMATION WAS RECEIVED THAT THE ROD ENDCAP WAS FOUND TO BE SEPARATED; HOWEVER, THE ROD ACHIEVED ITS INTENDED LENGTH AND THERE WAS NO REPORTED PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641243 ↗
  • MalfunctionApr 9, 2021Fracture; Fracture

    INFORMATION WAS RECEIVED THAT DURING AN EXPLANT PROCEDURE FOR FINAL FUSION AFTER THE ROD HAD ACHIEVED ITS INTENDED PURPOSE, IT WAS NOTED THERE WAS A FRACTURE THROUGH THE ROD END CAP. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641238 ↗
  • MalfunctionNov 3, 2020Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD WAS NOT LENGTHENING. THERE WAS NO PATIENT INJURY REPORTED. IT IS UNKNOWN IF THE PATIENT WILL HAVE A REVISION PROCEDURE.

    Patient problems: Failure of Implant

    MAUDE report 10782789 ↗
  • MalfunctionNov 3, 2020Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD WAS NOT LENGTHENING. THERE WAS NO PATIENT INJURY REPORTED. IT IS UNKNOWN IF THE PATIENT WILL HAVE A REVISION PROCEDURE.

    Patient problems: Failure of Implant

    MAUDE report 10782786 ↗
  • MalfunctionOct 30, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT ULTRASOUND IMAGES REVEALED THAT THE END CAP HAD MIGRATED. IT IS UNKNOWN IF A REVISION PROCEDURE IS PLANNED AT THIS TIME. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10766472 ↗
  • MalfunctionJun 2, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT X-RAY IMAGES REVEALED THAT THE END CAP HAD DISENGAGED. AT THIS TIME IT IS UNKNOWN IF A REVISION PROCEDURE IS PLANNED. NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10112897 ↗
  • MalfunctionApr 27, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT X-RAY IMAGING REVEALED THAT THE END CAP HAD DISENGAGED. AS PER THE SURGEON, THE RODS WILL REMAIN IMPLANTED UNTIL DEFINITIVE FUSION.

    Patient problems: Failure of Implant

    MAUDE report 10003644 ↗
  • MalfunctionMar 16, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT THE END CAP WAS DISENGAGED FROM THE ROD. THE SURGEON HAS NEVER ATTEMPTED ANY DISTRACTIONS.

    Patient problems: Failure of Implant

    MAUDE report 9837641 ↗
  • MalfunctionMar 3, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT THE END CAP HAD DISLODGED FROM THE ROD. AT THIS TIME IS UNKNOWN IF PATIENT WILL UNDERGO A REVISION PROCEDURE.

    Patient problems: Failure of Implant

    MAUDE report 9783264 ↗
  • InjuryFeb 27, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER THE PATIENT WAS BEING TREATED FOR INFLAMMATION AND X-RAYS REVEALED THAT BOTH END CAPS HAD DISLODGED FROM THE RODS.

    Patient problems: Failure of Implant

    MAUDE report 9766211 ↗
  • MalfunctionFeb 27, 2020Corroded; Structural Problem; Failure to Advance; Detachment of Device or Device Component; Migration; Physical Resistance/Sticking; Corroded; Structural Problem; Failure to Advance; Detachment of Device or Device Component; Migration; Physical Resistance/Sticking

    INFORMATION WAS RECEIVED THAT X-RAY IMAGES REVEALED THAT THE END CAP HAD DISENGAGED FROM THE ROD. IT IS UNKNOWN IF A REVISION PROCEDURE IS PLANNED. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 9765878 ↗

Other Growing Rod System- Magnetic Actuation devices

All Growing Rod System- Magnetic Actuation devices →