MAGEC® Spinal Bracing and Distraction System MC2-4590SRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 4.5 mm standard rod with 90 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This MAGEC spinal growth-rod system is intended to support and gradually lengthen the spine. Reports mostly describe rods that fail to lengthen or have other mechanical problems, sometimes leading to revision surgery; there are 25 reports, 5 injuries, and no deaths. One open recall began in November 2022 over packaging that might fail to keep the device sterile.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00812258022624), received between Jan 13, 2020 and Apr 21, 2025 — classified as 5 injuries, 20 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2022Class IIOpen, Classified

    During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

    Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Advance7
  • Mechanical Problem4
  • Migration or Expulsion of Device4
  • Break2
  • Detachment of Device or Device Component2
  • Fracture2
  • Mechanical Jam2
  • Naturally Worn2

Patient problems

  • Failure of Implant20
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Metal Related Pathology2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2025Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD FAILING TO DISTRACT. THE REVISION SURGERY WAS COMPLETED WITH NO ISSUE OR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 21875661 ↗
  • MalfunctionAug 14, 2024Break; Failure to Advance; Break; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 19983194 ↗
  • InjuryAug 13, 2024Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 19976981 ↗
  • MalfunctionMar 23, 2024Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 18965572 ↗
  • InjuryJan 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    UPON FURTHER REVIEW, THIS REPORT HAS BEEN DETERMINED TO BE A DUPLICATE OF MFR # 3006179046-2024-00100. THE RECORD ASSOCIATED WITH THIS REPORT WILL BE VOIDED AND ALL RELEVANT INFORMATION WILL BE DOCUMENTED AND REPORTED UNDER MFR # 3006179046-2024-00100.

    Patient problems: Metal Related Pathology

    MAUDE report 18597324 ↗
  • InjuryJan 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Metal Related Pathology

    MAUDE report 18520247 ↗
  • MalfunctionSep 12, 2023Failure to Advance; Failure to Advance

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD NOT LENGTHENING AFTER MULTIPLE ATTEMPTS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 17732666 ↗
  • MalfunctionJul 26, 2023Fracture; Fracture

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17398277 ↗
  • MalfunctionJul 26, 2023Structural Problem; Structural Problem

    INFORMATION WAS RECEIVED THAT THE LEFT SIDE ROD'S O-RING WAS DISPLACED IN 2019. ADDITIONALLY, IT WAS REPORTED THAT THERE WAS BLACKENING OF THE TISSUE AROUND THE ACTUATOR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17398263 ↗
  • MalfunctionJan 18, 2023Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 16196461 ↗
  • MalfunctionJan 18, 2023Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 16196336 ↗
  • MalfunctionJul 28, 2022Mechanical Jam; Naturally Worn; Mechanical Jam; Naturally Worn

    INFORMATION WAS RECEIVED THAT THE ROD HAS FAILED TO LENGTHEN AFTER (B)(6) 2020. AS PER THE REPORTER, THE ROD IS POSSIBLY JAMMED AND THERE WAS NOTICEABLE WEAR DEBRIS IN-SITU. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 15125166 ↗
  • MalfunctionJul 28, 2022Mechanical Jam; Naturally Worn; Mechanical Jam; Naturally Worn

    INFORMATION WAS RECEIVED THAT THE ROD HAS FAILED TO LENGTHEN AFTER (B)(6) 2020. AS PER THE REPORTER, THE ROD IS POSSIBLY JAMMED AND THERE WAS NOTICEABLE WEAR DEBRIS IN-SITU. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 15125130 ↗
  • MalfunctionJan 5, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD FAILED TO REACH TERMINAL DISTRACTION LENGTHEN. THE RODS WERE REMOVED FOR FINAL FUSION WITH A COMPETITOR PRODUCT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13162943 ↗
  • MalfunctionJan 5, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD FAILED TO REACH TERMINAL DISTRACTION LENGTHEN. THE RODS WERE REMOVED FOR FINAL FUSION WITH A COMPETITOR PRODUCT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13162870 ↗
  • MalfunctionJun 4, 2021Mechanical Problem; Mechanical Problem

    N/A.

    Patient problems: Failure of Implant

    MAUDE report 11938885 ↗
  • MalfunctionJun 4, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 11938857 ↗
  • MalfunctionMar 8, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT X-RAYS SHOWED THE ACTUATOR END CAP IS MOVING WITH THE ROD. THERE IS NO REVISION PROCEDURE PLANNED AT THIS TIME. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 11439902 ↗
  • MalfunctionMar 1, 2021Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE DUE TO A BROKEN ROD. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD WAS BROKEN.

    Patient problems: Failure of Implant

    MAUDE report 11397107 ↗
  • MalfunctionDec 30, 2020Unstable; Unstable

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON AN (B)(6) 2020. AS PER THE REPORTER THE ROD PULLED OUT. THERE WAS NO PATIENT INJURY REPORTED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) THE ROD FUNCTIONED AS INTENDED.

    Patient problems: Failure of Implant

    MAUDE report 11098839 ↗
  • MalfunctionDec 29, 2020Fracture; Fracture

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED (B)(6) 2019. AS PER THE REPORTER THE ROD FRACTURED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) IT WAS IDENTIFIED THAT THE ROD WAS WORKING AND POSSIBLE TO BE DETRACTED, BUT NOT CONFORM TO THE FORCE TEST.

    Patient problems: Failure of Implant

    MAUDE report 11091224 ↗
  • MalfunctionAug 18, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. AS PER THE REPORTER, X-RAY IMAGES REVEALED THAT THE END CAP HAD MIGRATED. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10423745 ↗
  • MalfunctionJul 31, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER, X-RAY IMAGES REVEALED THE END CAP DISCONNECTED FROM THE ACTUATOR. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10348638 ↗
  • MalfunctionJul 17, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED AFTER THE THIRD LENGTHENING OF THE RODS, X-RAY IMAGES REVEALED THAT THE END CAP HAD LOOSENED. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10292813 ↗
  • InjuryJan 13, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT X-RAYS REVEALED THAT THE END CAP HAD DISENGAGED. AT THIS TIME IT IS UNKNOWN IF THE PATIENT WILL UNDERGO A REVISION PROCEDURE.

    Patient problems: Failure of Implant

    MAUDE report 9583884 ↗

Other Growing Rod System- Magnetic Actuation devices

All Growing Rod System- Magnetic Actuation devices →