MAGEC® Spinal Bracing and Distraction System MC2-5590RRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 5.5 mm offset rod with 90 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00812258022679), received between Jul 9, 2020 and Aug 14, 2024 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Migration” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2022Class IIOpen, Classified

    During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

    Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration3
  • Migration or Expulsion of Device3
  • Mechanical Problem2
  • Break1
  • Corroded1
  • Failure to Advance1
  • Fracture1
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant8
  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2024Break; Failure to Advance; Break; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 19981711 ↗
  • MalfunctionJun 16, 2023Mechanical Jam; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 17148919 ↗
  • MalfunctionAug 9, 2022Migration; Migration

    INFORMATION WAS RECEIVED THAT THE RIGHT SIDE ROD HAS ENDCAP SEPARATION. THE PATIENT IS ASYMPTOMATIC AND THE ROD IS STILL LENGTHENING. NO PATIENT INJURY HAS BEEN REPORTED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15192417 ↗
  • MalfunctionApr 18, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT IMMEDIATELY UPON CLOSING, LENGTHENING WAS ATTEMPTED AND THE LEFT ROD WAS NOT LENGTHENING. SURGEON PLANS TO MONITOR AND CONTINUE TO TRY TO LENGTHEN. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 14137600 ↗
  • MalfunctionSep 24, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT THE END CAP CAME OFF AND THE ROD WAS UNABLE TO DISTRACT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12528408 ↗
  • MalfunctionSep 9, 2021Corroded; Fracture; Corroded; Fracture

    INFORMATION WAS RECEIVED THAT THE MAGNET APPEARED TO BE FRACTURED INTERNALLY AND DISASSOCIATED. ADDITIONALLY, IT WAS REPORTED THAT FRETTING/CORROSION IS VISIBLE. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12447032 ↗
  • MalfunctionJul 13, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT END CAP LOOSENED. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12162165 ↗
  • MalfunctionJul 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD EVIDENCE OF GALLING.

    Patient problems: Insufficient Information

    MAUDE report 12144749 ↗
  • MalfunctionJun 3, 2021Migration; Migration

    INFORMATION WAS RECEIVED THAT AFTER THE FOURTH APPOINTMENT FOR LENGTHENING, THE MAGEC ROD END CAP WAS LOOSE. A REVISION PROCEDURE IS PLANNED BUT HAS YET TO OCCUR.

    Patient problems: Insufficient Information

    MAUDE report 11930954 ↗
  • InjuryJul 17, 2020Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration

    INFORMATION WAS RECEIVED THAT PRIOR TO THE FIRST DISTRACTION, X -RAY IMAGES REVEALED LOOSENING OF THE END CAP. A REVISION PROCEDURE IS NOT PLANNED AT THIS TIME. THERE WAS NO PATIENT INJURY REPORTED. AS PER THE SURGEON, THE INITIAL DISTRACTION SESSION WENT WELL.

    Patient problems: Failure of Implant

    MAUDE report 10290767 ↗
  • MalfunctionJul 9, 2020Mechanical Problem; Mechanical Problem

    AS PER REPORTER THE FIRST X-RAY (PRIOR TO 1ST DISTRACTION) X-RAY INDICATED A LOOSENING OF THE ENDCAP. FIRST DISTRACTION WENT WELL, NO REVISION PROCEDURE IS PLANNED AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 10255596 ↗