MAGEC® Spinal Bracing and Distraction System MC2-5570RRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 5.5 mm offset rod with 70 mm actuator · Class 2
All models of MAGEC Spinal Bracing and Distraction System (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00812258022655), received between Jul 9, 2021 and May 31, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Corroded” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 1, 2022Class IIOpen, Classified
During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.
Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Corroded1
- Mechanical Jam1
- Mechanical Problem1
- No Apparent Adverse Event1
Patient problems
- Failure of Implant2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION AND GALLING. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17034318 ↗MalfunctionSep 15, 2021Mechanical Problem; Mechanical Problem
MalfunctionJul 9, 2021No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR A FINAL FUSION. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD EVIDENCE OF GALLING.
Patient problems: Insufficient Information
MAUDE report 12144539 ↗
Other Growing Rod System- Magnetic Actuation devices
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation112 reports
- MAGEC MC2-5090SNuvasive, Inc. · Growing rod system- magnetic actuation1 recall33 reports
- MAGEC® Spinal Bracing and Distraction System RA002-4545SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation82 reports
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLRNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation81 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall26 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall25 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page