Skip to content

MAGEC® Spinal Bracing and Distraction System MC2-6070SRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 6.0 mm standard rod with 70 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812258022686), received between Jul 1, 2021 and Jul 1, 2021 — classified as 1 malfunction.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 1, 2022Class IIOpen, Classified

    During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

    Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2021No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON DUE TO ROD HAVING FULLY EXTENDED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD SURFACE WEAR, MINOR GALLING, SCRATCHING, DISCOLORATION AND MINOR MECHANICAL DAMAGE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12100785 ↗

Other Growing Rod System- Magnetic Actuation devices

All Growing Rod System- Magnetic Actuation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page