MAGEC® Spinal Bracing and Distraction System MC2-6090SRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 6.0 mm standard rod with 90 mm actuator · Class 2
All models of MAGEC Spinal Bracing and Distraction System (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00812258022709), received between Oct 2, 2019 and Jun 16, 2023 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 1, 2022Class IIOpen, Classified
During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.
Z-0559-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Jam4
- Break2
- Defective Device2
- Migration2
- Corroded1
- Failure to Advance1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- Failure of Implant7
- Insufficient Information3
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2023Break; Corroded; Mechanical Jam; Break; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED DUE TO A BROKEN ROD; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17148877 ↗MalfunctionMay 31, 2023Break; Mechanical Jam; Break; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. DURING DISASSEMBLY THE RADIAL BEARING WAS FOUND TO BE BROKEN. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER THE ROD HAD FULLY EXTENDED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17032014 ↗MalfunctionMay 26, 2023Mechanical Jam; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED AS THE ELECTRONIC REMOTE CONTROLLER (ERC) WAS NOT FUNCTIONAL. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17014946 ↗MalfunctionOct 29, 2022Failure to Advance; Mechanical Jam; Failure to Advance; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, DURING SURFACE GRADING GALLING WAS NOTED. THE REASON FOR REMOVAL IS UNKNOWN; THERE WAS INSUFFICIENT INFORMATION TO KNOW IF THERE WAS AN ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 15696357 ↗MalfunctionJul 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD EVIDENCE OF GALLING.
Patient problems: Insufficient Information
MAUDE report 12144387 ↗MalfunctionJun 3, 2021Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD NOT LENGTHENING. NO PATIENT ADVERSE EVENT WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 11935661 ↗InjuryDec 17, 2020Migration; Migration
A REVISION PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE.
Patient problems: Failure of Implant
MAUDE report 11033833 ↗MalfunctionApr 12, 2020Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2018 FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) IT WAS IDENTIFIED THAT THE ROD HAD A PIN FRACTURE AND A LARGE AMOUNT OF DEBRIS IN HOUSING TUBE. NO OTHER INFORMATION IN RELATION TO PATIENT HAS BEEN REPORTED.
Patient problems: Failure of Implant
MAUDE report 9951843 ↗MalfunctionFeb 28, 2020Migration; Migration
INFORMATION WAS RECEIVED THAT THE END CAP ON THE ROD WAS DISLODGED. AS PER REPORTER, A REVISION PROCEDURE HAS NOT BEEN PERFORMED BUT IS PLANNED.
Patient problems: Failure of Implant
MAUDE report 9771983 ↗MalfunctionOct 2, 2019Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT THE END CAP HAD BECOME DISENGAGED FROM THE ROD. AS PER REPORTER, NO REVISION PROCEDURE WILL BE CONDUCTED AT THIS TIME.
Patient problems: Failure of Implant
MAUDE report 9145665 ↗
Other Growing Rod System- Magnetic Actuation devices
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation112 reports
- MAGEC MC2-5090SNuvasive, Inc. · Growing rod system- magnetic actuation1 recall33 reports
- MAGEC® Spinal Bracing and Distraction System RA002-4545SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation82 reports
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLRNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation81 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall26 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall25 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page