Quick summary

This NuVasive MAGEC growing rod is a magnetic, remotely adjustable spinal rod used to lengthen the spine as a child grows. The 30 reports mostly describe the rod failing to lengthen, breaking, or jamming, with several patients needing revision surgery and metal debris found on removal; 14 reports involved injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00812258026295), received between Apr 9, 2019 and May 1, 2024 — classified as 14 injuries, 16 malfunctions.

The most frequently reported specific device problem is “Defective Device” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device10
  • Fracture6
  • Mechanical Jam5
  • Mechanical Problem5
  • Break3
  • Failure to Advance3
  • Appropriate Device Problem Term/Code Not Available2
  • Material Erosion2

Patient problems

  • Failure of Implant29
  • No Known Impact Or Consequence To Patient3
  • Metal Related Pathology2
  • Foreign Body Reaction1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2024Material Erosion; Failure to Advance; Material Erosion; Failure to Advance

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD NOT DISTRACTING. DURING THE REMOVAL PROCEDURE THERE WAS METAL DEBRIS NOTED AT THE SUPERIOR ASPECT OF THE LENGTHENING SEGMENT.

    Patient problems: Failure of Implant; Metal Related Pathology

    MAUDE report 19222196 ↗
  • InjuryMay 1, 2024Material Erosion; Failure to Advance; Material Erosion; Failure to Advance

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD NOT DISTRACTING. DURING THE REMOVAL PROCEDURE THERE WAS METAL DEBRIS NOTED AT THE SUPERIOR ASPECT OF THE LENGTHENING SEGMENT.

    Patient problems: Failure of Implant; Metal Related Pathology

    MAUDE report 19222174 ↗
  • InjurySep 20, 2023Mechanical Problem; Failure to Advance; Material Deformation; Mechanical Problem; Failure to Advance; Material Deformation

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Foreign Body Reaction

    MAUDE report 17782957 ↗
  • MalfunctionAug 9, 2023Mechanical Jam; Mechanical Jam

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17492158 ↗
  • MalfunctionJan 3, 2023Break; Break

    INFORMATION WAS RECEIVED THAT DURING A CLINIC A PATIENT PRESENTED WITH A BROKEN ROD. THE PATIENT'S PARENTS NOR THE PATIENT RECALL ANY INCIDENT THAT WOULD HAVE CAUSED THE BREAKAGE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 16088335 ↗
  • MalfunctionNov 22, 2022Fracture; Fracture

    INFORMATION WAS RECEIVED THAT AFTER TAKING X-RAYS POST-DISTRACTION, THE ROD WAS FOUND TO BE BROKEN ABOVE THE DISTAL SCREWS. AS A RESULT, A REVISION PROCEDURE WAS PERFORMED. IT WAS NOTED THAT THE FRACTURE TOOK PLACE NEAR THE END OF THE INTENDED DISTRACTION PERIOD FOR THE PATIENT. NO FURTHER PATIENT IMPACT AND NO PATIENT HARM WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 15846469 ↗
  • MalfunctionNov 22, 2022Fracture; Fracture

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO A ROD FRACTURE. IT WAS REPORTED THE ROD HAD ACHIEVED INTENDED DISTRACTION GOALS AND HAD FRACTURED ONLY A SHORT TIME PRIOR TO THE REVISION SURGERY. THERE WAS NO REPORT OF PATIENT FALL, ACCIDENT, OR OTHER TRAUMATIC EVENT. NO FURTHER ADVERSE PATIENT IMPACT AND NO PATIENT HARM WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 15846272 ↗
  • MalfunctionJun 17, 2022Fracture; Mechanical Jam; Fracture; Mechanical Jam

    DURING EVALUATION OF A RETURNED ROD, IT WAS FOUND THE ROD EXHIBITED A BROKEN RADIAL BEARING AND WOULD NOT DISTRACT. INITIALLY, THE ROD WAS RETURNED WITHOUT A REPORTED PROBLEM AFTER IT HAD BEEN EXPLANTED AS PART OF A PLANNED REMOVAL FOR FINAL FUSION. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT ASSOCIATED WITH THIS DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14734781 ↗
  • MalfunctionJan 12, 2022Fracture; Fracture

    INFORMATION WAS RECEIVED THAT THE ROD HAS A BROKEN PIN AND A BROKEN RADIAL BEARING. IT WAS REPORTED THE ROD IS STILL ABLE LENGTHEN. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13240621 ↗
  • MalfunctionNov 30, 2021Mechanical Problem; Mechanical Problem

    N/A.

    Patient problems: Failure of Implant

    MAUDE report 12896743 ↗
  • MalfunctionNov 30, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12896129 ↗
  • InjuryOct 6, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12584695 ↗
  • MalfunctionAug 3, 2021Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED AS THE ROD WAS BROKEN. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12268339 ↗
  • InjuryJun 30, 2021Fracture; Fracture

    INFORMATION WAS RECEIVED THAT THE ROD WAS FOUND TO BE BROKEN DURING A CLINIC VISIT. THE ROD WAS ALREADY FULLY DISTRACTED. THE SURGEON OPTED TO LEAVE THE ROD IMPLANTED AND STABILIZED IT BY PLACING A DOMINO CONNECTOR TO THE FIXATION.

    Patient problems: Failure of Implant

    MAUDE report 12096739 ↗
  • MalfunctionJun 29, 2021Fracture; Fracture

    INFORMATION WAS RECEIVED THAT THE ROD WAS FOUND TO BE FRACTURED. THE ROD WAS EXPLANTED; NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 12089657 ↗
  • MalfunctionJun 14, 2021Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE SURGEON BELIEVING THAT THE RODS WERE NOT DEPLOYING DUE TO EXTENDED TIME BETWEEN DISTRACTIONS. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 11996244 ↗
  • MalfunctionJun 3, 2021Break; Break

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 11935704 ↗
  • InjuryMay 6, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2017. AS PER THE REPORTER THE ROD WAS NOT FUNCTIONING PROPERLY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 10031418 ↗
  • MalfunctionApr 27, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED IN (B)(6) 2019. BASED ON ENGINEERING  REVIEW PERFORMED ON (B)(6) 2020, IT WAS IDENTIFIED THAT THE ROD FAILED DISTRACT. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10001717 ↗
  • MalfunctionApr 12, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2018 FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) IT WAS IDENTIFIED THAT THE ROD FAILED TO DISTRACT. NO OTHER INFORMATION IN RELATION TO PATIENT HAS BEEN REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 9951930 ↗
  • InjuryFeb 26, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER REPORTER, THE PATIENT DID NOT LENGTHEN DURING MULTIPLE ATTEMPTS OVER THE LAST YEAR. REPORTEDLY, THE RODS MAY BE JAMMED AND DID NOT REACH THEIR MAXIMUM DISTRACTION LENGTH.

    Patient problems: Failure of Implant

    MAUDE report 9757163 ↗
  • MalfunctionFeb 20, 2020Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS COMPLETED ON (B)(6) 2019. AS PER THE REPORTER, THE RODS WERE JAMMED.

    Patient problems: Failure of Implant

    MAUDE report 9736118 ↗
  • InjuryFeb 20, 2020Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS COMPLETED ON (B)(6) 2019. AS PER REPORTER, THE ROD WAS JAMMED.

    Patient problems: Failure of Implant

    MAUDE report 9736062 ↗
  • MalfunctionFeb 19, 2020Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS COMPLETED ON (B)(6) 2020. AS PER REPORTER, THE ROD WAS JAMMED.

    Patient problems: Failure of Implant

    MAUDE report 9729120 ↗
  • InjuryDec 20, 2019Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019. AS PER THE REPORTER THE PATIENT HAD A BROKEN ROD.

    Patient problems: Failure of Implant

    MAUDE report 9506953 ↗