MAGEC® Spinal Bracing and Distraction System MS1-4590RRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 4.5 mm offset rod with 90 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00812258026301), received between Aug 29, 2019 and Apr 18, 2024 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem4
  • Defective Device3
  • Degraded2
  • Break1
  • Fracture1
  • Material Discolored1

Patient problems

  • Failure of Implant8
  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024Break; Break

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19140424 ↗
  • MalfunctionJul 7, 2021Mechanical Problem; Mechanical Problem

    CORRECTED DEVICE INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12132597 ↗
  • MalfunctionJul 1, 2021Degraded; Fracture; Degraded; Fracture

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD HAVING A PIN FRACTURE. DURING A REVIEW OF INVESTIGATION FINDINGS FROM THE (B)(6) IT WAS IDENTIFIED THAT THE ROD HAD A PIN FRACTURE WITH DEBRIS IN THE HOUSING TUBE, AND THE PRESENCE OF DISCOLORATION/CORROSION WAS NOTED.

    Patient problems: Failure of Implant

    MAUDE report 12105068 ↗
  • MalfunctionJul 1, 2021Material Discolored; Material Discolored

    IT WAS REPORTED THAT AN EXPLANT PROCEDURE WAS PERFORMED SINCE THE ROD WAS FULLY EXTENDED; HOWEVER, DURING A REVIEW OF INVESTIGATION FINDINGS FROM THE (B)(6), IT WAS IDENTIFIED THAT THE ROD HAD MINOR DISCOLORATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12105050 ↗
  • MalfunctionJul 1, 2021Degraded; Mechanical Problem; Degraded; Mechanical Problem

    IT WAS REPORTED THAT A ROD WAS EXPLANTED ON AN UNKNOWN DATE FOR AN UNKNOWN REASON; HOWEVER, DURING A REVIEW OF INVESTIGATION FINDINGS FROM THE (B)(6), IT WAS IDENTIFIED THAT THE ROD WAS UNABLE TO RETRACT AND MINOR DISCOLORATION WAS NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12104971 ↗
  • MalfunctionNov 4, 2020Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD WAS NOT LENGTHENING WELL. THERE WAS NO PATIENT INJURY REPORTED AND IT IS UNKNOWN IF A REVISION PROCEDURE IS PLANNED.

    Patient problems: Failure of Implant

    MAUDE report 10787325 ↗
  • MalfunctionSep 19, 2020Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT X-RAY IMAGES TAKEN (B)(6) 2019, IDENTIFIED A LOCKING PIN FAILURE. THERE WAS NO PATIENT INJURY REPORTED. IT IS UNKNOWN IF A REVISION PROCEDURE IS PLANNED.

    Patient problems: Failure of Implant

    MAUDE report 10557815 ↗
  • InjuryMay 6, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION SURGERY WAS PERFORMED ON (B)(6) 2017. AS PER THE REPORTER THE ROD WAS NOT FUNCTIONING PROPERLY .

    Patient problems: Failure of Implant; Pain

    MAUDE report 10031422 ↗
  • MalfunctionApr 8, 2020Defective Device; Defective Device

    SEE UPDATED INFORMATION IN H2, H3, H6 AND H10.

    Patient problems: Failure of Implant

    MAUDE report 9938544 ↗
  • MalfunctionAug 29, 2019Defective Device; Defective Device

    INFORMATION RECEIVED ALLEGED A PIN BREAK WAS NOTED. ON (B)(6) 2018 A REVISION PROCEDURE WAS PERFORMED TO EXCHANGE THE DAMAGED ROD FOR BIGGER SIZES TO TAKE THE PATIENT THROUGH TO DEFINITIVE FUSION.

    Patient problems: Failure of Implant

    MAUDE report 8945708 ↗

Other Growing Rod System- Magnetic Actuation devices