Skip to content

MAGEC® Spinal Bracing and Distraction System MS1-5590RRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 5.5 mm offset rod with 90 mm actuator · Class 2

All models of MAGEC Spinal Bracing and Distraction System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00812258026325), received between Nov 5, 2019 and Aug 3, 2023 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Defective Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Defective Device3
  • Break1
  • Mechanical Problem1
  • No Apparent Adverse Event1

Patient problems

  • Failure of Implant4
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 3, 2023Break; Break

    INFORMATION WAS RECEIVED THAT ROD FRACTURED AND LOOKS LIKE IT CAME "UNHINGED." THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 17451581 ↗
  • MalfunctionApr 19, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD FAILED TO DISTRACT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14145388 ↗
  • MalfunctionJul 1, 2021No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD EXPERIENCED A PIN BREAK, SURFACE WEAR, GALLING, SCRATCHING, DISCOLORATION, AND MECHANICAL DAMAGE. ADDITIONALLY, FORCE TESTING WAS NOT PERFORMED AS IT WAS IMPOSSIBLE TO RETRACT THE ROD. NO PATIENT HARM WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 12101102 ↗
  • InjuryApr 10, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019 FOR AN UNKNOWN REASON. AS PER THE REPORTER DURING SERVICING IT WAS IDENTIFIED THAT THE O-RING WAS DAMAGED.

    Patient problems: Failure of Implant

    MAUDE report 9950973 ↗
  • InjuryNov 25, 2019Defective Device; Defective Device

    THIS REPORT HAS BEEN UPDATED TO INCLUDE INVESTIGATION FINDINGS.

    Patient problems: Failure of Implant

    MAUDE report 9370964 ↗
  • InjuryNov 5, 2019Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A PATIENT UNDERWENT A POSTERIOR SPINAL INSTRUMENTATION IMPLANT APPROXIMATELY THREE YEARS AGO. THE IMPLANTS WERE STRYKER SCREWS BUT MAGEC RODS ALONG WITH BONE GRAFTING POSTERIOR ARTHRODESIS T2, T3, T4 AND ALSO L1 AND L2. AFTER TWO ADJUSTMENTS OF THE RODS, THE PATIENT HAD THE RODS REMOVED DUE TO THE FAILURE OF THE RODS LENGTHENING TRANSCUTANEOUSLY. THIS PATIENT UNDERWENT REMOVAL OF THE RODS LAST MONTH AND HAD REPLACEMENT WITH SOLERA SCREWS, CONNECTORS, AND RODS. NO COMPLICATIONS AND PATIENT WAS DISCHARGED FROM THE HOSPITAL WITHOUT CONCERNS.

    Patient problems: Failure of Implant

    MAUDE report 9279366 ↗

Other Growing Rod System- Magnetic Actuation devices

All Growing Rod System- Magnetic Actuation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page