Skip to content

M6-C 6MRecalls & reported problems

Spinal Kinetics, Inc. · Artificial Cervical Disc, 6M (6mm H x 15mm W x 12.5mm D) · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00812388030018), received between Feb 23, 2023 and Oct 6, 2025 — classified as 12 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Collapse” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Collapse4
  • Adverse Event Without Identified Device or Use Problem3
  • Insufficient Device Problem Information3
  • Off-Label Use2
  • Patient Device Interaction Problem2
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • Neck Pain5
  • Pain3
  • Implant Pain2
  • Insufficient Information2
  • Osteolysis2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Dizziness1
  • Erosion1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THAT DR. (B)(6) EXPERIENCED AN M6-C DEVICE FAILURE WHICH WAS SUBSEQUENTLY EXPLANTED FROM ONE OF HIS PATIENTS. DR. (B)(6) REQUESTED THAT THE EXPLANTED DEVICE BE RETURNED FOR INTERNAL INSPECTION. ACCORDING TO INFORMATION OBTAINED FROM THE PATIENT'S HUSBAND, THE PATIENT PREVIOUSLY HAD A PRODISC IMPLANTED (IMPLANT DATE UNKNOWN) AND LATER UNDERWENT A REVISION PROCEDURE IN 2020 PERFORMED BY DR. (B)(6). THE PATIENT WAS INITIALLY INFORMED THAT A CAGE WOULD BE IMPLANTED, HOWEVER INTRAOPERATIVELY DR. (B)(6) ELECTED TO IMPLANT A TWO-LEVEL M6-C. THE PATIENT LATER PRESENTED TO DR. (B)(6), WHO EXPLANTED BOTH OF THE M6-C DEVICES ON (B)(6) 2025. THE PATIENT'S HUSBAND REPORTED THAT THE PATIENT HAS EXPERIENCED CHRONIC PAIN FOR SEVERAL YEARS AND THAT BONE RESORPTION HAD BEGUN. NO IMAGES WERE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 23227760 ↗
  • InjuryOct 6, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THAT DR. (B)(6) EXPERIENCED AN M6-C DEVICE FAILURE WHICH WAS SUBSEQUENTLY EXPLANTED FROM ONE OF HIS PATIENTS. DR. (B)(6) REQUESTED THAT THE EXPLANTED DEVICE BE RETURNED FOR INTERNAL INSPECTION. ACCORDING TO INFORMATION OBTAINED FROM THE PATIENT'S HUSBAND, THE PATIENT PREVIOUSLY HAD A PRODISC IMPLANTED (IMPLANT DATE UNKNOWN) AND LATER UNDERWENT A REVISION PROCEDURE IN 2020 PERFORMED BY DR. (B)(6). THE PATIENT WAS INITIALLY INFORMED THAT A CAGE WOULD BE IMPLANTED; HOWEVER INTRAOPERATIVELY DR. (B)(6) ELECTED TO IMPLANT A TWO-LEVEL M6-C. THE PATIENT LATER PRESENTED TO DR. (B)(6), WHO EXPLANTED BOTH OF THE M6-C DEVICES ON (B)(6) 2025. THE PATEINT'S HUSBAND REPORTED THAT THE PATIENT HAS EXPERIENCED CHRONIC PAIN FOR SEVERAL YEARS AND THAT BONE RESORPTION HAD BEGUN. NO IMAGES WERE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 23223401 ↗
  • InjuryApr 11, 2025Collapse; Collapse

    IT WAS REPORTED THAT THERE WAS DISC LUCENCY AND COLLAPSE OF M6-C DISC AFTER PATIENT HAD A FALL AND HAD NECK PAIN, INCIDENTAL FINDING.

    Patient problems: Neck Pain

    MAUDE report 21827540 ↗
  • InjurySep 25, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS -M6C REVISION CASE-- BLUMENTHAL ET AL (ESJ 2024). DISCUSSES THE OUTCOMES AND MANAGEMENT STRATEGIES FOR PATIENTS WHO UNDERWENT REVISION OR REMOVAL OF CERVICAL TOTAL DISC REPLACEMENTS (TDR). THE MOST COMMON REASON FOR REMOVAL/REVISION ENCOUNTERED IN THIS COHORT WAS SEVERE OSTEOLYSIS, 7 WERE M6-C.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Osteolysis

    MAUDE report 20306495 ↗
  • InjurySep 25, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS A REMOVAL OF C3/4 AND C4/5 M6-C IMPLANTED IN 2021 - SOUTH AFRICAN PATIENT - PROBABLY FIRST IMPLANTED THERE.

    Patient problems: Implant Pain

    MAUDE report 20304354 ↗
  • InjuryJan 23, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C IMPLANTED AT LEVEL C5/6 IN (B)(6) 2019. PATIENT BEGAN EXPERIENCING PAIN 4 YEARS POST IMPLANTING. THE TOTAL DISC REPLACEMENT WAS EXPLANTED AND REPLACED WITH ACDF IN (B)(6) 2023.

    Patient problems: Neck Pain

    MAUDE report 18567962 ↗
  • InjuryDec 2, 2023Collapse; Collapse

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C IMPLANTED AT C6/7 LEVEL IN 2016. PATIENT EXPERIENCED PAIN, NUMBNESS, DIZZINESS AND TENSION. THE DEVICE WAS EXPLANTED IN (B)(6) 2023 DUE TO CLINICAL SYMPTOMS.THE DEVICE WAS NOT INTACT AT THE TIME OF REMOVAL.

    Patient problems: Dizziness; Numbness; Pain

    MAUDE report 18253313 ↗
  • MalfunctionNov 1, 2023Collapse; Collapse

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C ARTIFICIAL CERVICAL DISC IMPLANTED AT C5-6 IN 2018. A REVISION SURGERY TO REMOVE M6-C AND CONVERT TO ACDF IN (B)(6) 2023 DUE TO OSTEOLYSIS AND COLLAPSE OF THE M6-C.

    Patient problems: Failure of Implant

    MAUDE report 18052595 ↗
  • MalfunctionOct 11, 2023Material Separation; Material Separation

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C IMPLANTED IN 2013 AT C4/5. IT WAS NOT REPORTED IF PATIENT EXPERIENCED ANY SYMPTOMS HOWEVER IT WAS REPORTED THAT THE M6-C HAD COLLAPSED AND OSTEOLYSIS WAS PRESENT. RADIOGRAPHIC IMAGES SHOW THAT THE PATIENT HAD A 2 LEVEL ACDF AT LEVELS C5/6 AND C6/7. IT IS UNKNOWN WHETHER THAT FUSION WAS PERFORMED BEFORE OR AFTER THE IMPLANTATION OF THE M6-C DISC AT C4/5. THE M6-C DEVICE WAS REMOVED DUE TO COLLAPSED DISC.

    Patient problems: Insufficient Information

    MAUDE report 17917067 ↗
  • InjurySep 26, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION PROVIDED STATES THAT AN M6-C WAS IMPLANTED AT C4/5 ON (B)(6) 2021. THE PATIENT REPORTED INCREASED NECK PAIN. THE M6-C DEVICE WAS REMOVED ON (B)(6) 2023 DUE TO SUSPECTED LOOSENING OF THE DEVICE.

    Patient problems: Neck Pain

    MAUDE report 17821427 ↗
  • InjuryJul 31, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C IMPLANTED IN 2019 AT C5/6 AND C7/T1. PATIENT HAD PREVIOUS FUSION AT C6/7 IN 2016. THE PATIENT EXPERIENCED NON-SPECIFIC NECK PAIN. THE MRI SHOWED SIGNS OF INFECTION. THE M6-C WAS REMOVED DUE TO SUSPECTED INFECTION.

    Patient problems: Neck Pain

    MAUDE report 17427944 ↗
  • InjuryJun 12, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION PROVIDED STATES THAT PATIENT HAD M6-C IMPLANTED AT C5/6 ON (B)(6) 2022. PT EXPERIENTS CERVICAL NECK PAIN AND DISCOMFORT AT IMPLANT LEVEL. THE M6-C WAS REMOVED ON (B)(6) 2023.

    Patient problems: Implant Pain

    MAUDE report 17114116 ↗
  • InjuryMay 2, 2023Collapse; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Collapse; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    INFORMATION PROVIDED STATES THAT A REVISION SURGERY WAS PERFORMED TO REMOVE TWO COLLAPSED M6-C DISCS THAT WERE IMPLANTED 12/13 YEARS AGO. NO ADDITIONAL INFORMATION HAS BEEN FORTHCOMING.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Erosion; Insufficient Information; Neck Pain; Neck Stiffness; Osteolysis

    MAUDE report 16850881 ↗
  • InjuryFeb 23, 2023Patient-Device Incompatibility; Patient-Device Incompatibility

    INFORMATION PROVIDED STATES THE PATIENT EXPERIENCED PAIN AND LYSIS ZONE COULD BE SEEN IN THE BONE 13 YEARS POST IMPLANT OF AN M6-C PROSTHESIS. THE PATIENT HAD BEEN DIAGNOSED WITH MULTIPLE SCLEROSIS AT THE TIME OF INITIAL IMPLANT IN 2010. THE M6-C WAS EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16432435 ↗

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →