Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80P0Recalls & reported problems
Cardiac Science Corporation · Powerheart G5 AED, Fully Automatic, Dual Language (English/Spanish),no battery or pads · Class 3
All models of Powerheart G5 Kit, Auto, US Eng/ESP LA (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Cardiac Science Powerheart G5 is a fully automatic external defibrillator (AED) that delivers a shock without manual shock delivery. Most of the 33 reports describe the unit failing its routine self-test or being unable to read the patient's heart rhythm, showing messages such as "Unit Failed" or "Check Pads," or reading high impedance through the pads; many were found during routine testing with no patient involved. One report describes a death during an attempted resuscitation, with no injuries and 32 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00812394020577), received between Oct 26, 2020 and Apr 10, 2024 — classified as 1 death, 32 malfunctions.
The most frequently reported specific device problem is “Unable to Obtain Readings” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unable to Obtain Readings15
- Failure of Device to Self-Test13
- High impedance5
- Use of Device Problem2
- Appropriate Device Problem Term/Code Not Available1
- Battery Problem1
- Failure to Power Up1
- Output below Specifications1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- No Patient Involvement3
- Insufficient Information2
- Necrosis1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2024Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19080145 ↗DeathMar 21, 2024Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A FEMALE PATIENT (AGE UNKNOWN), THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER SET OF DEFIB PADS TO CONTINUE TREATING THE PATIENT, HOWEVER, THE SAME ERROR WAS OBSERVED. THE COMPLAINANT INDICATED THAT THE EMS TEAM ARRIVED SHORTLY AFTER AND THEIR EQUIPMENT WAS USED TO TREAT THE PATIENT WITH NO SUCCESS. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.
Patient problems: Necrosis
MAUDE report 18952308 ↗MalfunctionDec 8, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18290677 ↗MalfunctionNov 28, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18216652 ↗MalfunctionSep 1, 2023High impedance; Unable to Obtain Readings; High impedance; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DID NOT DETECT THE ATTACHED ELECTRODE PADS AND FAILED FOR HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17669355 ↗MalfunctionJul 21, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING,THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17374040 ↗MalfunctionJun 26, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17199075 ↗MalfunctionJun 22, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17181788 ↗MalfunctionJun 15, 2023Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17139478 ↗MalfunctionApr 17, 2023High impedance; High impedance
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED FOR HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16751303 ↗MalfunctionApr 10, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16707433 ↗MalfunctionJan 19, 2023Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16207732 ↗MalfunctionJan 9, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16122894 ↗MalfunctionDec 19, 2022Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A RED "X" INDICATOR. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16006270 ↗MalfunctionAug 29, 2022Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15311189 ↗MalfunctionJul 8, 2022Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING,THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14957391 ↗MalfunctionJun 30, 2022High impedance; Unable to Obtain Readings; Appropriate Device Problem Term/Code Not Available; High impedance; Unable to Obtain Readings; Appropriate Device Problem Term/Code Not Available
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DID NOT DETECT THE ATTACHED ELECTRODE PADS AND FAILED FOR HIGH IMPEDANCE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 14874264 ↗MalfunctionMay 31, 2022Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14550475 ↗MalfunctionApr 22, 2022Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "SELF TEST FAILURE FOR P/S V OUT OF RANGE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14188172 ↗MalfunctionMar 20, 2022Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13829725 ↗MalfunctionMar 18, 2022Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13806634 ↗MalfunctionDec 22, 2021Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13059078 ↗MalfunctionOct 27, 2021Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12703229 ↗MalfunctionSep 13, 2021Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12464602 ↗MalfunctionJul 21, 2021Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12205613 ↗
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗