Intellisense Adult Defibrillation Pads, Powerheart® G5 XELAED001BRecalls & reported problems

Zoll Medical Corporation · Intellisense Adult defibrillation Pads, non-polarized, for Powerheart® G5 · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

These are Zoll Intellisense adult defibrillation pads used with the Powerheart G5 defibrillator. The 52 reports mostly describe the defibrillator failing to detect the attached pads or showing a "check pads" message, plus high impedance readings. No deaths and no injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
52
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 52 adverse event reports that list this exact device (DI 00812394020959), received between Nov 21, 2020 and Aug 14, 2024 — classified as 52 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (45 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unable to Obtain Readings45
  • High impedance12
  • Failure of Device to Self-Test3
  • Use of Device Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions51
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 52 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCATED DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19980440 ↗
  • MalfunctionJul 1, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING TRAINING BY FACILITY STAFF, THE ASSOCIATED DEFIBRILLATOR FAILED TO DETECT THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643823 ↗
  • MalfunctionJul 1, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING TRAINING BY FACILITY STAFF, THE DEVICE DID NOT DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643811 ↗
  • MalfunctionMar 6, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18847389 ↗
  • MalfunctionMar 5, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE DID NOT DETECT THSEE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18840100 ↗
  • MalfunctionMar 1, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18815200 ↗
  • MalfunctionJan 25, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR DID NOT DETECT THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18578915 ↗
  • MalfunctionNov 20, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18169708 ↗
  • MalfunctionNov 14, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18130861 ↗
  • MalfunctionNov 9, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18106074 ↗
  • MalfunctionNov 9, 2023High impedance; Unable to Obtain Readings; High impedance; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED FOR HIGH IMPEDANCE USING THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18099960 ↗
  • MalfunctionNov 9, 2023High impedance; Unable to Obtain Readings; High impedance; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED FOR HIGH IMPEDANCE USING THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18099790 ↗
  • MalfunctionNov 3, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18067421 ↗
  • MalfunctionNov 2, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18056524 ↗
  • MalfunctionOct 31, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18041554 ↗
  • MalfunctionOct 24, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17997108 ↗
  • MalfunctionOct 24, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17997106 ↗
  • MalfunctionOct 16, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "CHECK PADS" MESSAGE USING THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17938349 ↗
  • MalfunctionOct 13, 2023High impedance; High impedance

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED IMPEDANCE TESTING USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17930625 ↗
  • MalfunctionSep 29, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING INCOMING TESTING, THE ASSOCIATED DEFIBRILLATOR DOES NOT DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17842880 ↗
  • MalfunctionSep 21, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17787764 ↗
  • MalfunctionJun 27, 2023High impedance; High impedance

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE FAILED FOR HIGH IMPEDANCE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17209173 ↗
  • MalfunctionMay 17, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "CHECK PADS" MESSAGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16948060 ↗
  • MalfunctionMay 4, 2023High impedance; Unable to Obtain Readings; High impedance; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR DID NOT DETECT THE ATTACHED ELECTRODE PADS AND FAILED FOR HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16873099 ↗
  • MalfunctionMay 4, 2023High impedance; Unable to Obtain Readings; High impedance; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR DID NOT DETECT THE ATTACHED ELECTRODE PADS AND FAILED FOR HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16873081 ↗

Other Automated External Defibrillators (Non-Wearable) devices