Skip to content

Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80CRecalls & reported problems

Zoll Medical Corporation · Powerheart G5 AED, Fully Automatic, Dual Language (English/Spanish), with Intellisense Adult Pads with CPR feedback device and Battery · Class 3

All models of Powerheart G5 Kit, Auto, US Eng/ESP LA (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Zoll Powerheart G5 is an automated external defibrillator with a battery and pads. Many of its 188 reports describe the unit being unable to obtain readings, failing its self-test, or not powering up. There are no injuries and 2 death reports. There are 2 open recalls, both about units failing self-test after prolonged exposure to extreme temperature or humidity.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
202
Deaths reported
2
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 202 adverse event reports that list this exact device (DI 00812394021208), received between Dec 29, 2023 and Sep 29, 2026 — classified as 2 deaths, 200 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (70 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 12 reportsOct 2024: 9 reportsNov 2024: 7 reportsDec 2024: 4 reportsJan 2025: 8 reportsFeb 2025: 9 reportsMar 2025: 6 reportsApr 2025: 4 reportsMay 2025: 15 reportsJun 2025: 4 reportsJul 2025: 5 reportsAug 2025: 2 reportsSep 2025: 5 reportsOct 2025: 6 reportsNov 2025: 10 reportsDec 2025: 8 reportsJan 2026: 9 reportsFeb 2026: 10 reportsMar 2026: 6 reportsApr 2026: 7 reportsMay 2026: 10 reportsJun 2026: 8 reportsJul 2026: 6 reportsAug 2026: 6 reportsSep 2026: 16 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2025Class IIOpen, Classified

    The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

    Z-1319-2025 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

  2. Feb 13, 2025Class IIOpen, Classified

    The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

    Z-1318-2025 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unable to Obtain Readings70
  • Failure of Device to Self-Test45
  • Failure to Power Up24
  • Use of Device Problem21
  • Defibrillation/Stimulation Problem17
  • Failure to Charge15
  • Incorrect Interpretation of Signal10
  • Output Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions190
  • Insufficient Information7
  • Intraoperative Cardiac Valve Injury2
  • Burn(s)1
  • Cardiac Arrest1
  • Ventricular Fibrillation1

Recent adverse event reports

The 25 most recent of 202 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934155 ↗
  • MalfunctionSep 28, 2026Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED ERROR CODE \"E05\" OBSERVED IN THE ERROR LOG.\\COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914895 ↗
  • MalfunctionSep 25, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN \"ECG FAULT -501\" ERROR MESSAGE.\\COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899953 ↗
  • MalfunctionSep 24, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE DISPLAYED AN \"ECG FAULT -501\" ERROR MESSAGE.\\COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882723 ↗
  • MalfunctionSep 18, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50817078 ↗
  • MalfunctionSep 15, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "SELF-TEST FAILURE FOR P/S V OUT OF RANGE" ERROR MESSAGE.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772256 ↗
  • MalfunctionSep 15, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN \"ECG FAULT -501\" ERROR MESSAGE.\\COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772131 ↗
  • MalfunctionSep 15, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772091 ↗
  • MalfunctionSep 14, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755871 ↗
  • MalfunctionSep 14, 2026Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50753168 ↗
  • MalfunctionSep 14, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50753140 ↗
  • MalfunctionSep 11, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A \"BRIDGE TEST FAILED\" MESSAGE.\\COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50730750 ↗
  • MalfunctionSep 8, 2026Unable to Obtain Readings; Incorrect Interpretation of Signal

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A FEMALE PATIENT (AGE UNKNOWN), THE DEVICE PROMPTED "NO SHOCK ADVISED" FOR A RHYTHM CLINICIANS BELIEVED TO BE SHOCKABLE AND DID NOT DETECT THE ATTACHED ELECTRODE PADS.COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669412 ↗
  • MalfunctionSep 4, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50616440 ↗
  • MalfunctionSep 2, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567517 ↗
  • MalfunctionSep 1, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50543976 ↗
  • MalfunctionAug 28, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50497854 ↗
  • MalfunctionAug 25, 2026Failure to Charge; Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50432446 ↗
  • MalfunctionAug 21, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379344 ↗
  • MalfunctionAug 14, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50255026 ↗
  • MalfunctionAug 14, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50253274 ↗
  • MalfunctionAug 13, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50233632 ↗
  • MalfunctionJul 29, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50071032 ↗
  • MalfunctionJul 17, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "SELF-TEST FAILURE FOR P/S V OUT OF RANGE" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25855046 ↗
  • MalfunctionJul 13, 2026Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE ISSUED A "SHOCK ADVISED" PROMPT FOR A HEART RHYTHM THEY BELIEVED WAS NON-SHOCKABLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 25796692 ↗

FDA clearance and classification

The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.

Common questions about the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C

Has the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1319-2025 (Class II), began on Feb 13, 2025: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C?

The FDA's MAUDE database lists 202 adverse event reports for this device, received between Dec 29, 2023 and Sep 29, 2026. In those reports, the most common specific device problems are “Unable to Obtain Readings” (70), “Failure of Device to Self-Test” (45) and “Failure to Power Up” (24); the most common patient problems are “Intraoperative Cardiac Valve Injury” (2), “Burn(s)” (1) and “Cardiac Arrest” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C?

2 reports are classified as deaths, 0 as injuries and 200 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Powerheart G5 Kit, Auto, US Eng/ESP LA G5A-80C, and is it still sold?

The FDA's device database lists Zoll Medical Corporation as the company responsible for it (brand Powerheart G5 Kit, Auto, US Eng/ESP LA, model G5A-80C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Automated External Defibrillators (Non-Wearable) devices

All Automated External Defibrillators (Non-Wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page