Skip to content

Boston Scientific TCN Electrodes recallZ-2403-2026 · Boston Scientific Neuromodulation Corporation · initiated Apr 29, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reusable electrodes may not meet expected performance levels.

FDA root cause category: Under Investigation by firm

What the company is doing

On April 29, 2026, Boston Scientific issued a Urgent Medical device Removal Notification via Federal Express/email/hand delivery. Boston Scientific asked consignees to take the following actions: 1. DO NOT use affected product. Remove affected products from your facility s inventory, and segregate the units in a secure place. 2. For products already in use, immediately discard according to your facility s procedure. 3. Unused products should be returned to Boston Scientific in accordance with the enclosed instructions. 4. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 5. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 6. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions.

Recall details

Recall number
Z-2403-2026
Status
Open, Classified
Initiated by the company
Apr 29, 2026
Posted by the FDA
Jun 11, 2026
Recalling firm
Boston Scientific Neuromodulation Corporation
Quantity
49,251 units
Product code
GXI
FDA submissions
K050084, K183177
Distribution
Worldwide distribution - US Nationwide and the countries of PE, MX, PR, CO, MY, CA.

Product description

Boston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: TCN-5-3m, TCN Electrode, NITINOL, 3M CABLE, 5 cm x 0.45 mm (0.0175 in) REF: TCN-10-3M, TCN Electrode, NITINOL, 3M CABLE, 10 cm x 0.45 mm (0.0175 in); REF: TCN-15-3M, TCN Electrode, NITINOL, 3M CABLE, 15 cm x 0.45 mm (0.0175 in) REF: TCN-20-3M, TCN Electrode, NITINOL, 3M CABLE, 20 cm x 0.45 mm (0.0175 in)

Code information (models, lots, serial numbers)

All Lots: / UDI: TCN-10: 00813250013054; TCN-10-3M: 00813250017502; TCN-15: 00813250013061; TCN-15-3M: 00813250017519; TCN-20: 00813250013078; TCN-20-3M: 00813250017526; TCN-5: 00813250013047; TCN-5-3M:00813250017533

Devices named in this recall

8 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98939).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page