Skip to content

Impella 0052-0011Recalls & reported problems

Abiomed, Inc. · Axillary Insertion Kit, PKGD · Class 2

All models of Impella Heart Pump (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813502010015), received between Mar 8, 2026 and May 1, 2026 — classified as 1 death, 2 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Patient Device Interaction Problem2
  • Difficult to Advance1
  • Fluid/Blood Leak1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2
  • Vascular Dissection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2026Fluid/Blood Leak; Patient Device Interaction Problem

    AN IMPELLA 5.5 DEVICE WAS INSERTED VIA THE LEFT AXILLARY ARTERY IN A 63-YEAR-OLD MALE PATIENT WITH CARDIOMYOPATHY AND A HISTORY OF KNOWN CORONARY ARTERY DISEASE. THE PATIENT¿S CLINICAL STATE PRIOR TO INITIATION OF SUPPORT WAS IDENTIFIED AS SCAI SHOCK STAGE D. LEAKING WAS OBSERVED FROM BETWEEN THE 10MM GELWEAVE GRAFT AND 23F SHEATH DESPITE BEING LOCKED WITH GRAFT LOCKS AND TWO ADDITIONAL HEAVY TIES. THE PATIENT SURVIVED TO EXPLANT. BLEEDING MAY BE ASSOCIATED WITH ACCESS-SITE COMPLICATIONS AND GRAFT INTERFACE ISSUES.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25057328 ↗
  • InjuryApr 1, 2026Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    AN IMPELLA 5.5 DEVICE WAS INSERTED VIA THE RIGHT AXILLARY ARTERY IN A 34-YEAR-OLD FEMALE PATIENT PRESENTING WITH ACUTE MYOCARDIAL INFARCTION COMPLICATED BY CARDIOGENIC SHOCK (AMI/CGS), SCAI SHOCK STAGE E. THE PATIENT¿S COMORBIDITIES WERE UNKNOWN. A GRAFT WAS SEWN ONTO THE RIGHT AXILLARY ARTERY, AND THE SHEATH AND GRAFT WERE SECURED IN PLACE. DURING THE PROCEDURE, THE SURGEON ATTEMPTED TO ADVANCE A 0.035-INCH J-WIRE BUT WAS UNABLE TO PASS IT THROUGH THE ANASTOMOSIS. THE ANASTOMOSIS WAS TAKEN DOWN, AT WHICH TIME DISSECTION OF THE AXILLARY ARTERY WAS IDENTIFIED. A PATCH WAS PLACED TO REPAIR THE ARTERY, AND A 10 MM GRAFT WAS SUBSEQUENTLY ANASTOMOSED TO THE PATCH. THE PATIENT REMAINED ON IMPELLA SUPPORT FOLLOWING THE REPAIR. THE REPORTED VASCULAR DISSECTION IS CONSISTENT WITH ACCESS-RELATED ARTERIAL INJURY DURING SURGICAL MANIPULATION AND GRAFT ANASTOMOSIS IN THE SETTING OF URGENT MECHANICAL CIRCULATORY SUPPORT PLACEMENT

    Patient problems: Vascular Dissection

    MAUDE report 24754145 ↗
  • DeathMar 8, 2026Adverse Event Without Identified Device or Use Problem

    AN IMPELLA 5.5 DEVICE WAS INSERTED VIA THE RIGHT AXILLARY ARTERY IN A 64 YEAR-OLD MALE PATIENT PRESENTING WITH ACUTE MYOCARDIAL INFARCTION COMPLICATED BY CARDIOGENIC SHOCK (AMI/CGS), SCAI SHOCK STAGE D, WITH A HISTORY OF KNOWN CORONARY ARTERY DISEASE (CAD) AND RENAL INSUFFICIENCY. THE PATIENT UNDERWENT IMPELLA 5.5 PLACEMENT IN THE OPERATING ROOM, AND A JP DRAIN WAS PLACED, WHICH DEMONSTRATED ONGOING DRAINAGE. THE PATIENT RECEIVED 1 UNIT OF PACKED RED BLOOD CELLS (PRBCS). THE ANTICOAGULATION LEVEL WAS FOUND TO BE HIGH, PROMPTING THE CARE TEAM TO STOP HEPARIN, AFTER WHICH THE BLEEDING STOPPED. HEPARIN WAS LATER RESTARTED AT A LOWER DOSE. THE PATIENT REMAINED ON SUPPORT. THE REPORTED MAJOR BLEED REQUIRING BLOOD TRANSFUSION AND HEMOSTASIS IS CONSISTENT WITH ACCESS-SITE COMPLICATIONS AND THE ANTICOAGULATION/PURGE REQUIREMENTS ASSOCIATED WITH IMPELLA SUPPORT. THE BLEEDING WAS RESOLVED WITH STANDARD INTERVENTION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24542076 ↗

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page