- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813502010237), received between Nov 10, 2017 and Nov 30, 2018 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K122084); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack1
- Deformation Due to Compressive Stress1
- Kinked1
- Leak/Splash1
- Split1
Patient problems
- No Consequences Or Impact To Patient2
- Blood Loss1
- Death1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 30, 2018Leak/Splash; Leak/Splash
PATIENT WAS ADMITTED FOR A HIGH RISK ANGIOPLASTY OF THE CORONARY ARTERIES. ACCESS AT THE LEFT FEMORAL ARTERY FOR PLACEMENT OF THE IMPELLA CP, HUB LEAK OBSERVED WITH BLOOD LOST. 14FR INTRODUCER REMOVED AND REPLACED BY 16FR COOK. NO BLOOD PRODUCTS GIVEN TO CONTROL BLOOD LOSS. PATIENT IS STABLE AND NO EXTENDED HOSPITALIZATION REPORTED. NO OTHER ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Blood Loss; No Consequences Or Impact To Patient
MAUDE report 8121295 ↗MalfunctionJun 18, 2018Crack; Crack
IT WAS REPORTED THAT DURING A HRPCI (HIGH RISK PERCUTANEOUS CORONARY INTERVENTION) PROCEDURE, ON (B)(6) 2018, AN INTRODUCER CRACKED AFTER INSERTION. THE ISSUE WAS IDENTIFIED DURING THE PROCEDURE AS FOLLOWS: UPON REMOVAL OF THE DILATOR, BLEEDING WAS NOTED FROM AROUND THE SHEATH. AFTER REMOVING THE SHEATH, A CRACK IN THE 14F 13CM INTRODUCER WAS FOUND. AMOUNT OF BLOOD LOSS IS UNKNOWN AT THIS TIME. PATIENT IS REPORTED TO BE STABLE, NO EXTENDED HOSPITALIZATION WAS REQUIRED. THE DEVICES ARE NOT AVAILABLE TO RETURN AS THEY WERE DISCARDED. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN RECEIVED AS OF DATE OF THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7611196 ↗DeathNov 10, 2017Kinked; Split; Deformation Due to Compressive Stress; Kinked; Split; Deformation Due to Compressive Stress
ADDITIONAL INFORMATION WAS RECEIVED THAT INDICATED THE FOLLOWING: THE IMPELLA WAS ATTEMPTED TO BE PLACED ON (B)(6), AT 10:15 AM. THE ATTEMPT WAS ABORTED, AND PATIENT EXPIRED ON (B)(6) AT 10:51 AM. SALVAGE SITUATION AND ACCESS ATTEMPTS WERE CEASED. PHYSICIAN'S PLAN WAS TO PLACE IMPELLA CP FOR SUPPORT. IN THE ICU THE PATIENT CODED DURING PREP TO CCL (CARDIAC CARE LAB). CPR STARTED AND CONTINUED THRU OUT ATTEMPTED IMPELLA INSERTION. IN CCL SHEATH WAS BEING PLACED INTO RIGHT FEMORAL. IMPELLA WAS OPENED AND PREPPED READY FOR IMPLANT. VISUALIZATION WAS POOR DURING PLACEMENT OF PIGTAIL AND WOULD NOT ADVANCE TO THE LEFT VENTRICLE. WHEN FLUORO VISUALIZATION WAS MADE IT SHOWED THE WIRE LOOPED WITHIN ARTERY. DECISION WAS MADE TO TRY THE LEFT FEMORAL ARTERY FOR ACCESS, BUT COULD NOT GAIN ACCESS TO THE ARTERY. WITH NO ROSC (RETURN OF SPONTANEOUS CIRCULATION) PROCEDURE WAS ABORTED. THE INITIAL INSERTION WAS THE SAME SIDE AS THE A-LINE THAT WAS PLACE POST IABP. PHYSICIAN SAID INSERTION WAS POTENTIALLY LOWER THAN PREFERRED. ANGULATION OF STICK IS UNKNOWN. PATIENT WAS OBESE WITH LARGE PANIS (AN ABNORMAL LAYER OF FIBROVASCULAR TISSUE OR GRANULATION TISSUE). THE LFA WAS NOT ACCESSED DUE TO ACTIVE CPR
Patient problems: Death
MAUDE report 7022665 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122084. One submission often covers many models, so check the product and lot details in each record.
- Z-2679-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
- Z-2680-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
- Z-2678-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
- Z-1467-2021 · Feb 22, 2021 · TerminatedAngioDynamics Safe Sheath Ultralite - Vascular Introducer/Catheter Introducer Catalogue #: VI10 - Product Usage: For the introduction of various types of pacing leads and catheter
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page