Skip to content

Impella 1000535Recalls & reported problems

Abiomed, Inc. · ABMD Intro Kit, 14Fr x Long, Pkgd US · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K222113); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2686-2026 · Root cause: Under Investigation by firm · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K222113. One submission often covers many models, so check the product and lot details in each record.

  • Z-2128-2023 · May 26, 2023 · Open, Classified
    Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page