Impella 1000434Recalls & reported problems
Abiomed, Inc. · Kit, Low Profile, 14Fr, Short, 1UP · Class 2
All models of Impella Heart Pump (34)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813502012996), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Difficult to Flush” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K222113); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 22, 2026Open, Classified
Potential for thrombus formation during prolonged use of the introducer.
Z-2599-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Flush1
- Difficult to Remove1
Patient problems
- Thrombosis/Thrombus1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Difficult to Flush; Difficult to Remove
IMPELLA CP 10 WAS INSERTED THROUGH THE 14FR LOW PROFILE SHEATH VIA THE FEMORAL ARTERY IN A 65 YEAR OLD FEMALE WHO PRESENTED IN AMI/CGS WITHOUT INCIDENT. THE PATIENT¿S COMORBIDITIES WERE UNKNOWN. THE PATIENT¿S CLINICAL STATE PRIOR TO INITIATION OF SUPPORT WAS IDENTIFIED AS SCAI SHOCK STAGE C. POST PROCEDURE IN THE ICU THE SHEATH SIDEARM WAS HOOKED UP TO A PRESSURE BAG WITH NS AND HEPARIN. THE FOLLOWING DAY AT 9:30AM IT WAS NOTED THAT THERE WAS MISCOMMUNICATION AND THE SIDEARM FLUSH WAS DISCONTINUED. IT WAS RECONNECTED AT 10:30AM THAT SAME DAY. WHILE ON SUPPORT THE DEVICE PERFORMED AS EXPECTED P-9 3.5L/MIN WITH NO NOTED ALARMS OR CONCERNS. UPON PLANNED REMOVAL, AFTER THE SHEATH WAS REMOVED, DURING STANDARD ASPIRATION OF THE SHEATH IT WAS NOTED TO BE "STUCK". THE SHEATH WAS PULLED BACK SLIGHTLY WITH SOME SUCCESS, HOWEVER, THERE WAS STILL NOTED DIFFICULTY DURING ASPIRATION. THE SYRINGE WAS THEN DISCONNECTED AND THE STOPCOCK WAS LEFT OPEN, WHICH SHOWED ADEQUATE FLOW. AFTER UTILIZING NEW SYRINGE AND ASPIRATION, CLOT WAS NOTED. THEY THEN UTILIZED A SLIP TIP SYRINGE TO ASPIRATE THROUGH THE VALVE AND CLOT WAS NOTED TO HAVE BEEN ASPIRATED FROM THERE AS WELL. SECOND ASPIRATION WAS FOUND TO BE
Patient problems: Thrombosis/Thrombus
MAUDE report 24734922 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K222113. One submission often covers many models, so check the product and lot details in each record.
- Z-2128-2023 · May 26, 2023 · Open, ClassifiedAbiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Other Introducer, Catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page