Skip to content

Impella 1000434Recalls & reported problems

Abiomed, Inc. · Kit, Low Profile, 14Fr, Short, 1UP · Class 2

All models of Impella Heart Pump (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813502012996), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Flush” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K222113); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 22, 2026Open, Classified

    Potential for thrombus formation during prolonged use of the introducer.

    Z-2599-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Flush1
  • Difficult to Remove1

Patient problems

  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Difficult to Flush; Difficult to Remove

    IMPELLA CP 10 WAS INSERTED THROUGH THE 14FR LOW PROFILE SHEATH VIA THE FEMORAL ARTERY IN A 65 YEAR OLD FEMALE WHO PRESENTED IN AMI/CGS WITHOUT INCIDENT. THE PATIENT¿S COMORBIDITIES WERE UNKNOWN. THE PATIENT¿S CLINICAL STATE PRIOR TO INITIATION OF SUPPORT WAS IDENTIFIED AS SCAI SHOCK STAGE C. POST PROCEDURE IN THE ICU THE SHEATH SIDEARM WAS HOOKED UP TO A PRESSURE BAG WITH NS AND HEPARIN. THE FOLLOWING DAY AT 9:30AM IT WAS NOTED THAT THERE WAS MISCOMMUNICATION AND THE SIDEARM FLUSH WAS DISCONTINUED. IT WAS RECONNECTED AT 10:30AM THAT SAME DAY. WHILE ON SUPPORT THE DEVICE PERFORMED AS EXPECTED P-9 3.5L/MIN WITH NO NOTED ALARMS OR CONCERNS. UPON PLANNED REMOVAL, AFTER THE SHEATH WAS REMOVED, DURING STANDARD ASPIRATION OF THE SHEATH IT WAS NOTED TO BE "STUCK". THE SHEATH WAS PULLED BACK SLIGHTLY WITH SOME SUCCESS, HOWEVER, THERE WAS STILL NOTED DIFFICULTY DURING ASPIRATION. THE SYRINGE WAS THEN DISCONNECTED AND THE STOPCOCK WAS LEFT OPEN, WHICH SHOWED ADEQUATE FLOW. AFTER UTILIZING NEW SYRINGE AND ASPIRATION, CLOT WAS NOTED. THEY THEN UTILIZED A SLIP TIP SYRINGE TO ASPIRATE THROUGH THE VALVE AND CLOT WAS NOTED TO HAVE BEEN ASPIRATED FROM THERE AS WELL. SECOND ASPIRATION WAS FOUND TO BE

    Patient problems: Thrombosis/Thrombus

    MAUDE report 24734922 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K222113. One submission often covers many models, so check the product and lot details in each record.

  • Z-2128-2023 · May 26, 2023 · Open, Classified
    Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates