This Lifeline View DDU-A2300 is a Defibtech automated external defibrillator, a portable device used to deliver a shock to someone in cardiac arrest. The 34 reports describe units that would not power on, showed a blank or flashing red status indicator with a service code, or had batteries that drained too early. None of these reports describe a patient event, and no deaths or injuries are mentioned. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 39
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 00815098020201), received between Feb 3, 2026 and Sep 30, 2026 — classified as 39 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Power Up19
- Premature Discharge of Battery11
- Battery Problem3
- Device Sensing Problem1
- Failure to Deliver Shock/Stimulation1
- Inappropriate/Inadequate Shock/Stimulation1
- Material Deformation1
- No Audible Prompt/Feedback1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
Recent adverse event reports
The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Battery Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953766 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
A customer reported that their AED's ASI is flashing red, and the device is reporting a service code. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935598 ↗MalfunctionSep 23, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873422 ↗MalfunctionSep 14, 2026Failure to Power Up
A customer reported that their AED's ASI is blank, and the AED will not power on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50755240 ↗MalfunctionSep 2, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50568788 ↗MalfunctionSep 2, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50568786 ↗MalfunctionAug 26, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50452381 ↗MalfunctionAug 13, 2026No Audible Prompt/Feedback
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50236250 ↗MalfunctionAug 13, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50235414 ↗MalfunctionAug 11, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50203269 ↗MalfunctionAug 3, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50116580 ↗MalfunctionJun 29, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED WILL NOT POWER ON, BUT IT POWERED ON WITH A SPARE BATTERY PACK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25651164 ↗MalfunctionJun 19, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25562682 ↗MalfunctionJun 17, 2026Premature Discharge of Battery
A CUSTOMER REPORTED THEIR AED IS FLASHING RED AND PROMPTING A "REPLACE BATTERY PACK NOW" WARNING. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25531907 ↗MalfunctionMay 21, 2026Battery Problem
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS FLASHING RED, AND THE DEVICE IS REPORTING A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25246172 ↗MalfunctionMay 19, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER REPORTED THE AED MAINTENANCE FREQUENCY AS EVERY SIX MONTHS. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25217735 ↗MalfunctionMay 15, 2026Failure to Power Up
A CUSTOMER REPORTED THAT UPON INSERTING A NEW BATTERY PACK, THEIR AED POWERED ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25187011 ↗MalfunctionMay 12, 2026Failure to Power Up
AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED WOULD NOT POWER ON; HOWEVER, WHEN TESTED WITH A SPARE BATTERY PACK, THE AED POWERED ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25151387 ↗MalfunctionMay 8, 2026Material Deformation
A CUSTOMER REPORTED THEY COULD NOT ATTACH THE PATIENT ELECTRODE PADS TO THEIR AED THE CUSTOMER PROVIDED A PHOTO OF THE AED WITH A DAMAGED CONNECTOR IN THE PAD RECEPTACLE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125745 ↗MalfunctionMay 5, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25082640 ↗MalfunctionApr 28, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019391 ↗MalfunctionApr 27, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S BATTERY WAS DEPLETED. WHEN TESTED WITH A SPARE BATTERY PACK, THE AED POWERED ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24998136 ↗MalfunctionApr 10, 2026Premature Discharge of Battery
A CUSTOMER REPORTED THAT THEIR AED BATTERY WAS DEPLETED. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24839764 ↗MalfunctionApr 10, 2026Failure to Power Up
A CUSTOMER REPORTED THAT THEIR AED'S ASI IS FLASHING RED, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24839763 ↗MalfunctionMar 25, 2026Power Problem
AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED UNIT WOULD AUTO POWER OFF. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690183 ↗
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page