Lifeline View DDU-E2300Recalls & reported problems
Defibtech, L.L.C. · Class 3
All models of Lifeline View (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Lifeline View DDU-E2300 defibrillator from Defibtech is a portable device used to shock the heart during sudden cardiac arrest. The 75 reports mostly describe the battery draining on its own, or the unit failing to power up or show a display; reports rose from 1 in the prior 12 months to 74 in the last 12 months. There are no deaths and 1 injury. One open recall, from March 2025, concerns missing Italian text on the labeling and instructions for use.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 78
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 78 adverse event reports that list this exact device (DI 00815098020232), received between Mar 26, 2025 and Sep 30, 2026 — classified as 1 injury, 77 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (50 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Mar 18, 2025Class IIOpen, Classified
It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).
Z-1692-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Discharge of Battery50
- Failure to Power Up19
- Battery Problem3
- Device Sensing Problem3
- No Audible Prompt/Feedback2
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions77
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 78 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953770 ↗MalfunctionSep 26, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50908147 ↗MalfunctionSep 26, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50908143 ↗MalfunctionSep 26, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906478 ↗MalfunctionSep 23, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873417 ↗MalfunctionSep 22, 2026Failure to Power Up
An international distributor reported that their customer's AED battery was depleted. When tested with a spare battery pack, the AED powered on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853442 ↗MalfunctionSep 22, 2026Premature Discharge of Battery
An international distributor reported their customer's AED battery was depleted. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853441 ↗MalfunctionSep 1, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50545160 ↗MalfunctionSep 1, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50545157 ↗MalfunctionSep 1, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50545153 ↗MalfunctionAug 31, 2026Premature Discharge of Battery
An international distributor reported that their customer's AED screen is blank. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50524487 ↗MalfunctionAug 24, 2026No Audible Prompt/Feedback
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419052 ↗MalfunctionAug 24, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419051 ↗MalfunctionAug 12, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50223295 ↗InjuryAug 10, 2026Premature Discharge of Battery
An international distributor reported their customer's AED ASI is flashing red, and the device is reporting a service code. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50195596 ↗MalfunctionAug 6, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50160571 ↗MalfunctionAug 5, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147539 ↗MalfunctionAug 5, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147527 ↗MalfunctionAug 5, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147523 ↗MalfunctionAug 5, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50143776 ↗MalfunctionAug 3, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50122806 ↗MalfunctionJul 28, 2026
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50057441 ↗MalfunctionJul 22, 2026Device Sensing Problem
Patient problems: Insufficient Information
MAUDE report 25906964 ↗MalfunctionJul 21, 2026Premature Discharge of Battery
AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED BATTERY PACK WAS DEPLETED, WHEN TESTED WITH A SPARE BATTERY PACK, THE AED DID POWER ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25899272 ↗MalfunctionJul 21, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25899134 ↗
FDA clearance and classification
Cleared through 510(k) K081259 on Jun 15, 2009 (decision: Substantially Equivalent). Submitted by Defibtech, LLC as “Ddu-2300 Semiautomatic External Defibrillator And Accessories”. The FDA review took 409 days. FDA record ↗
The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗
FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.
Common questions about the Lifeline View DDU-E2300
Has the Lifeline View DDU-E2300 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1692-2025 (Class II), began on Mar 18, 2025: It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Lifeline View DDU-E2300?
The FDA's MAUDE database lists 78 adverse event reports for this device, received between Mar 26, 2025 and Sep 30, 2026. In those reports, the most common specific device problems are “Premature Discharge of Battery” (50), “Failure to Power Up” (19) and “Battery Problem” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Lifeline View DDU-E2300?
None of the reports are classified as deaths, 1 as an injury and 77 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Lifeline View DDU-E2300 FDA approved?
It was cleared through 510(k) K081259 on Jun 15, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Lifeline View DDU-E2300, and is it still sold?
The FDA's device database lists Defibtech, L.L.C. as the company responsible for it (brand Lifeline View, model DDU-E2300). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,163 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,012 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,384 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,719 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,582 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page