Senhance X0007035Recalls & reported problems
Asensus Surgical, Inc. · Monopolar L-Hook Electrode · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815440020224), received between Dec 11, 2024 and Dec 11, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Sparking” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2024Sparking; Sparking
NO INJURY OCCURRED TO A PATIENT OR USER. THE EVENT OCCURRED AT (B)(6) IN THE UNITED STATES ON 29-NOV-2024, AND ASENSUS SURGICAL WAS ALSO MADE AWARE ON 29-NOV-2024. THERE WAS A <15 MINUTE DELAY TO THE SURGICAL CASE WITH NO IMPACT TO THE PATIENT, AND THE SURGEON CONVERTED TO TRADITIONAL LAPAROSCOPY TO COMPLETE THE CASE. THE EVENT INFORMATION STATES THAT DURING A SECOND PROCEDURE USING THE SYSTEM ON THIS DATE, THE SURGEON OBSERVED THAT A MONOPOLAR L-HOOK ELECTRODE WAS CONDUCTING ENERGY FROM A SMALL SPOT ON THE SHAFT OF THE INSTRUMENT JUST ABOVE THE TIP. THE INSTRUMENT APPEARED TO BE "ARCHING" ENERGY. THE SURGEON NOTICED A BRIGHT, FLASH-LIKE SPOT ABOUT AN INCH ABOVE THE TIP OF THE INSTRUMENT AS THE INSTRUMENT TOUCHED AN AREA OF FAT NEAR THE GALLBLADDER WHICH PROMPTED THE TEAM TO FURTHER INSPECT THE INSTRUMENT. THE INSTRUMENT WAS INSPECTED AND THE SHAFT COATING WAS NOTICED TO BE PHYSICALLY DAMAGED. NO HARM OR INJURY TO THE PATIENT OCCURRED AS A RESULT OF THIS AND THE INSTRUMENT WAS DEEMED UNUSABLE, AND THE REMAINDER OF THE CASE WAS COMPLETED WITH TRADITIONAL LAPAROSCOPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20904574 ↗
Other System,Surgical,Computer Controlled Instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.