Skip to content

SPEEDSTITCH OM-7000Recalls & reported problems

Smith & Nephew, Inc. · SPEEDSTITCH SUTURING DEVICE HANDLE · Class 2

All models of SPEEDSTITCH (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00817470001402), received between Mar 11, 2019 and Oct 23, 2025 — classified as 1 injury, 15 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K042031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Insufficient Device Problem Information2
  • Material Twisted/Bent2
  • Separation Problem2
  • Detachment of Device or Device Component1
  • Difficult to Open or Close1
  • Mechanical Problem1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2
  • No Information1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT, DURING A SURGERY, IT WAS NOTICED THAT A SPEEDSTITCH DEVICE WAS BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23367408 ↗
  • MalfunctionOct 13, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, THE HANDLE OF TWO (2) SPEEDSTITCH DEVICES KEPT BREAKING OFF WHEN IT WAS BEING INSERTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. RESULTS OF INVESTIGATION FOUND THE DEVICE HAVE A BROKEN NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17930705 ↗
  • MalfunctionMay 19, 2023Break; Break

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY PROCEDURE, THE SPEEDSTITCH DEVICE BROKE. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING AN UNKNOWN ALTERNATIVE DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16963559 ↗
  • MalfunctionApr 11, 2023Break; Break

    IT WAS REPORTED THAT THE SPEEDSTITCH DISPOSABLE NEEDLE BROKE INSIDE THE INSTRUMENT AND THEY WERE UNABLE TO REMOVE IT. NO CASE REPORTED; THEREFORE, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16715711 ↗
  • MalfunctionJan 3, 2022Separation Problem; Separation Problem

    IT WAS REPORTED THAT THE SHAFT OF THE SPEEDSTITCH CAME OFF. IT IS UNKNOWN WHEN THE EVENT OCCURRED. A BACK-UP DEVICE WAS AVAILABLE AND NO DELAYS OCCURRED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13142701 ↗
  • MalfunctionJan 3, 2022Separation Problem; Separation Problem

    IT WAS REPORTED THAT THE SHAFT OF THE SPEEDSTITCH CAME OFF. IT IS UNKNOWN WHEN THE EVENT OCCURRED. A BACK-UP DEVICE WAS AVAILABLE AND NO DELAYS OCCURRED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13142686 ↗
  • MalfunctionOct 29, 2021Break; Break

    IT WAS REPORTED THAT DURING A HIP SURGERY, EXTERNAL TO THE PATIENT, THE SHAFT OF THE SPEEDSTITCH PULLED OFF. THE PROCEDURE WAS COMPLETED WITH 3 MINUTES DELAY USING A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12720825 ↗
  • MalfunctionMay 26, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE SPEEDSTITCH DEVICE WAS NOT TAKING THE NEEDLE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11890503 ↗
  • MalfunctionMay 6, 2021Break; Break

    IT WAS REPORTED THAT THE SPEEDSTITCH WAS BROKEN, NO FURTHER INFORMATION. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT. IT IS UNKNOWN IF THERE WAS A DELAY OR IF A BACKUP DEVICE WAS AVAILABLE AND HOW THE ISSUE WAS RESOLVED. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782639 ↗
  • MalfunctionJan 7, 2021Break; Break

    IT WAS REPORTED THAT DURING A HIP SURGERY THE SPEEDSTITCH BARROW SEPARATED FROM THE HANDLE. NO PATIENT INJURIES OR DELAY REPORTED. SMITH AND NEPHEW BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11131581 ↗
  • MalfunctionDec 8, 2020Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING SURGERY, THE SPEEDSTITCH NEEDLE COULD NOT SEAT ALL THE WAY DOWN AND WOULD NOT RETRACT BACK INSIDE THE INSERTER TUBE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SMITH & NEPHEW DEVICE WAS AVAILABLE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10973065 ↗
  • InjuryOct 22, 2020Break; Break

    IT WAS REPORTED THAT DURING SURGERY, WHEN THE SURGEON WAS USING THE SPEEDSTITCH GUN FOR CAPSULE CLOSURE IN THE HIP AND UPON REMOVING THE DEVICE FROM THE PATIENT THE SHAFT OF THE SPEEDSTITCH DEVICE WOULD PULL OFF AND REMAIN IN THE CANNULA. THE SURGEON WOULD JUST GRAB THE PIECE STICKING OUT OF CANNULA AND REMOVE BY HAND. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR OTHER COMPLICATIONS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10717541 ↗
  • MalfunctionJun 24, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SPEEDSTITCH DEVICE WAS MISSING SCREWS AND NOT WORKING PROPERLY. THIS WAS NOTICED BEFORE THE PROCEDURE; THEREFORE, NO PATIENT WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10189769 ↗
  • MalfunctionJun 24, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SPEEDSTITCH DEVICE WAS MISSING SCREWS AND NOT WORKING PROPERLY. THIS WAS NOTICED BEFORE THE PROCEDURE; THEREFORE, NO PATIENT WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10189726 ↗
  • MalfunctionJul 22, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE WAS NOT WORKING DURING A PROCEDURE. THERE WAS NO DELAY OR PATIENT INJURIES BUT, IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED SINCE NO BACKUP DEVICE WAS AVAILABLE TO COMPLETE IT. ATTEMPTS WERE MADE TO RETRIEVE FURTHER INFORMATION BUT NO RESPONSE HAS BEEN RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8814499 ↗
  • MalfunctionMar 11, 2019Difficult to Open or Close

    IT WAS REPORTED THAT THERE IS AN ISSUE WITH THE DEVICE, THE LEVER IS NOT CLOSING AND OPENING. NO PATIENT INJURIES OR DELAY WERE REPORTED. PROCEDURE WAS FINISHED WITH A COMPETITOR DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8410136 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042031. One submission often covers many models, so check the product and lot details in each record.

  • Z-0611-2012 · Aug 8, 2011 · Terminated
    OPUS SpeedStitch Product Number OM-8086 is a suture cartridge For use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →