FAST-FIX 72201491Recalls & reported problems

Smith & Nephew, Inc. · ULTRA FAST-FIX ASSEMBLY - CURVED · Class 2 · Implant

All models of FAST-FIX (18)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew Ultra FAST-FIX suture assembly is used in knee procedures such as meniscus repair. Most of the 864 reports describe the device not deploying, pulling out or breaking during surgery, with surgeons sometimes finishing with a back-up product. The reports include 269 injuries and no deaths. Reports from the past year number 40, down from 81 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
862
Deaths reported
0
Injuries reported
269
Recalls naming it
0

Summary

The FDA's MAUDE database contains 862 adverse event reports that list this exact device (DI 03596010597205), received between Mar 30, 2017 and Jun 2, 2026 — classified as 269 injuries, 593 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (194 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 8 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 11 reportsApr 2024: 15 reportsMay 2024: 30 reportsJun 2024: 8 reportsJul 2024: 16 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 5 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 15 reportsMar 2025: 15 reportsApr 2025: 9 reportsMay 2025: 3 reportsJun 2025: 12 reportsJul 2025: 8 reportsAug 2025: 5 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 8 reportsDec 2025: 9 reportsJan 2026: 8 reportsFeb 2026: 1 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 5 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire194
  • Firing Problem156
  • Device Dislodged or Dislocated118
  • Break83
  • Premature Activation56
  • Failure to Advance46
  • Activation, Positioning or Separation Problem30
  • Detachment of Device or Device Component30

Patient problems

  • No Clinical Signs, Symptoms or Conditions517
  • No Known Impact Or Consequence To Patient261
  • No Information49
  • Device Embedded In Tissue or Plaque33
  • No Consequences Or Impact To Patient10
  • Foreign Body In Patient7
  • Injury4
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 862 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, THE T2 IMPLANT OF THE ULTRA FAST FIX COULDN'T BE SECURED AND WAS EASILY PULLED OUT. THE PROCEDURE WAS RESUMED, USING AN S+N EQUIVALENT BACK UP PRODUCT, HOWEVER IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25364573 ↗
  • MalfunctionMay 27, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR PROCEDURE, THE FAST FIX'S SECOND IMPLANT CAME OUT WHILE PULLING THE THREAD. THE PROCEDURE WAS RESUMED AFTER A NON-SIGNIFICANT DELAY, USING A SIMILAR S+N BACK UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25310927 ↗
  • MalfunctionMay 26, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, THE T2 OF AN ULTRA FAST-FIX WAS NOT SECURED AND COULD BE PULLED OUT. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25295834 ↗
  • MalfunctionMay 25, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR PROCEDURE, THE T2 OF THE ULTRA FASTFIX CAME OUT WHILE PULLING THE THREAD. THE PROCEDURE WAS RESUMED, WITH A NON SIGNIFICANT DELAY, USING A SIMILAR S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25271817 ↗
  • MalfunctionMay 25, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING AN ACL SURGERY, THE SECOND IMPLANT OF ONE (1) ULTRA FAST-FIX CAME OUT WHILE PULLING THE THREAD. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25271803 ↗
  • MalfunctionMay 21, 2026Firing Problem

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, THE T2 OF AN ULTRA FAST-FIX FELL OFF FROM THE INSERTER AFTER T1 WAS DEPLOYED. T1 WAS REMOVED WITH TWEEZERS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. ADDITIONAL ANESTHESIA WAS NOT REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT'S STATUS IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25237934 ↗
  • MalfunctionApr 1, 2026Failure to Fire

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, AFTER THE T1 IMPLANTS OF TWO (2) ULTRA FAST-FIX CURVED WERE DEPLOYED, THE T2 IMPLANTS WERE STUCK IN THE HEAD AND COULD NOT BE DEPLOYED. THE PROCEDURE WAS PERFORMED, USING AN EQUIVALENT S+N BACKUP; HOWEVER, IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753615 ↗
  • MalfunctionMar 31, 2026Firing Problem

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, BOTH T IMPLANTS OF ONE (1) ULTRA FAST-FIX CURVED FELL OFF AT THE SAME TIME AS THE FIRST DEPLOYMENT. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP, WITH A NON-SIGNIFICANT DELAY. NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24750561 ↗
  • MalfunctionMar 25, 2026Firing Problem

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, BOTH TS OF THE ULTRA-FAST-FIX FELL OFF AT THE SAME TIME AS THE FIRST DEPLOYMENT OCCURRED. THE TS WERE NOT DEPLOYED THROUGH THE MENISCUS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. THE PATIENT STATUS IS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691598 ↗
  • MalfunctionMar 24, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, AFTER THE SECOND T OF THE ULTRA-FAST-FIX WAS DEPLOYED, T2 FELL AUTOMATICALLY FROM THE INSERTER. T1 WAS REMOVED FROM THE PATIENT WITH SUCTION. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. THE PATIENT WAS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675818 ↗
  • MalfunctionMar 23, 2026Firing Problem

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, BOTH T'S OF THE ULTRA FAST-FIX FELL OFF AT THE SAME TIME AS THE FIRST DEPLOYMENT OCCURRED. THE T;S WERE NOT DEPLOYED THROUGH THE MENISCUS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP. THE PATIENT STATUS IS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663112 ↗
  • MalfunctionMar 23, 2026Firing Problem

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, BOTH T'S OF THE ULTRA FAST-FIX FELL OFF AT THE SAME TIME AS THE FIRST DEPLOYMENT OCCURRED. THE T'S WERE NOT DEPLOYED THROUGH THE MENISCUS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP. THE PATIENT STATUS IS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663010 ↗
  • MalfunctionFeb 23, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, THE T2 IMPLANT OF ONE (1) ULTRA FAST-FIX CURVED DID NOT HUNG UP TO THE JOINT CAVITY. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP, WITH A NON-SIGNIFICANT DELAY. NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24414536 ↗
  • MalfunctionJan 23, 2026Material Twisted/Bent; Firing Problem

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, BOTH T'S OF ONE (1) ULTRA FAST FIX CURVED FALL OFF AT THE SAME TIME AT THE FIRST DEPLOYMENT. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP; HOWEVER, IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24168344 ↗
  • MalfunctionJan 23, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, THE T2 ANCHOR OF ONE (1) ULTRA FAST FIX CURVED FELL OFF FROM THE MENISCUS. T1 IMPLANT WAS SUCCESSFULLY REMOVED WITH SUCTION. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP, WITH A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24168270 ↗
  • MalfunctionJan 23, 2026Firing Problem

    IT WAS REPORTED THAT, DURING A MENISCUS REPAIR, BOTH T'S OF ONE (1) ULTRA FAST FIX CURVED FELL OFF AT THE SAME TIME DURING THE FIRST DEPLOYMENT. T1 WAS DEPLOYED THROUGH THE MENISCUS AND WAS REMOVED. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP, WITH A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24167746 ↗
  • MalfunctionJan 19, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, TWO (2) ULTRA FAST-FIX T2 IMPLANTS FAILED TO DEPLOY. THE T1 IMPLANT WAS REMOVED FROM INSIDE THE PATIENT USING A GRASPER. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24111923 ↗
  • MalfunctionJan 15, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, THE T2 OF ONE (1) ULTRA FAST-FIX FAILED TO DEPLOY. T1 WAS SUCCESSFULLY REMOVED USING GRASPERS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. THE PATIENT'S STATUS IS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24090600 ↗
  • MalfunctionJan 15, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, THE ULTRA FAST-FIX T2 IMPLANT DISLODGED. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING AN EQUIVALENT S&N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24079629 ↗
  • MalfunctionJan 12, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, THE ULTRA FAST-FIX T2 IMPLANT DISLODGED. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING AN EQUIVALENT S&N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24048376 ↗
  • MalfunctionJan 5, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, T2 OF ONE FAST FIX FELL OFF FROM THE MENISCUS. T1 WAS REMOVED WITH SUCTION. THE PROCEDURE WAS RESUMED, AFTER A NON SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23967518 ↗
  • MalfunctionDec 17, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY THE ULTRA FAST-FIX T2 IMPLANT FELL OUT FROM THE MENISUCS WHEN PUSHING THE KNOT AFTER SUTURING. IN CONSEQUENCE, BOTH IMPLANTS WERE REMOVED FROM INSIDE THE PATIENT USING SUCTION. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP AFTER A NON-SIGNIFICANT SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23829935 ↗
  • MalfunctionDec 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, THE ULTRA FAST-FIX T IMPLANT FELL OUT WHEN PUSHING THE KNOT AFTER SUTURING. BOTH IMPLANTS WERE REMOVED FROM THE INSIDE THE PATIENT USING SUCTION. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23820974 ↗
  • MalfunctionDec 15, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, WHEN THE T2 OF AN ULTRA FAST-FIX WAS DEPLOYED, IT FELL OFF BY ITSELF. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802530 ↗
  • MalfunctionDec 15, 2025Firing Problem; Firing Problem

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, THE ULTRA FAST-FIX T IMPLANTS WERE DEPLOYED SIMULTANEOUSLY WHEN DEPLOYING THE T1 IMPLANT. BOTH IMPLANTS WERE REMOVED. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802501 ↗

FDA clearance and classification

The FDA regulates product code GAT (suture, nonabsorbable, synthetic, polyethylene) under 21 CFR 878.5000 as a Class II device. 21 CFR 878.5000 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the FAST-FIX 72201491

Has the FAST-FIX 72201491 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010597205), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the FAST-FIX 72201491?

The FDA's MAUDE database lists 862 adverse event reports for this device, received between Mar 30, 2017 and Jun 2, 2026. In those reports, the most common specific device problems are “Failure to Fire” (194), “Firing Problem” (156) and “Device Dislodged or Dislocated” (118); the most common patient problems are “Device Embedded In Tissue or Plaque” (33), “Foreign Body In Patient” (7) and “Injury” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the FAST-FIX 72201491?

None of the reports are classified as deaths, 269 as injuries and 593 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the FAST-FIX 72201491 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the FAST-FIX 72201491, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand FAST-FIX, model 72201491). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →