Arthrex® AR-4501Recalls & reported problems
Arthrex, Inc. · MENISCAL CINCH™ II · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex Meniscal Cinch II is a knee meniscus repair device that uses a suture. Many of its 396 adverse event reports describe parts breaking, the device coming loose, or the push button sticking and failing to deploy. The reports list 67 injuries and no deaths, and no recalls name this device. Reports fell from 89 in the prior 12 months to 12 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 400
- Deaths reported
- 0
- Injuries reported
- 67
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 400 adverse event reports that list this exact device (DI 00888867299139), received between Jan 30, 2019 and Sep 17, 2026 — classified as 67 injuries, 333 malfunctions.
The most frequently reported specific device problem is “Break” (146 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break146
- Device Dislodged or Dislocated79
- Failure to Advance49
- Physical Resistance/Sticking40
- Material Split, Cut or Torn17
- Material Twisted/Bent16
- Detachment of Device or Device Component13
- Mechanical Jam12
Patient problems
- No Clinical Signs, Symptoms or Conditions253
- Insufficient Information85
- No Known Impact Or Consequence To Patient26
- Failure of Implant15
- Foreign Body In Patient11
- Device Embedded In Tissue or Plaque6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- No Information3
Recent adverse event reports
The 25 most recent of 400 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Device Dislodged or Dislocated
On 3rd September 2026, it was reported by an Arthrex's employee via an Email that for 3 units of AR-4501, Meniscal Cinch¿ II:The first unit of AR-4501's anchor dropped out when the suture was tightened. A 2nd AR-4501 was changed to but both anchors dropped out too. A 3rd AR-4501 was used but the second anchor dropped out again.This was detected during a knee meniscus repair surgery on (b)(6) 2026. The procedure was completed successfully by using a 4th AR-4501. There were no patient harm and no secondary procedure needed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803256 ↗MalfunctionSep 17, 2026Device Dislodged or Dislocated
On 3rd September 2026, it was reported by an Arthrex's employee via an Email that for 3 units of AR-4501, Meniscal Cinch¿ II:The first unit of AR-4501's anchor dropped out when the suture was tightened. A 2nd AR-4501 was changed to but both anchors dropped out too. A 3rd AR-4501 was used but the second anchor dropped out again.This was detected during a knee meniscus repair surgery on (b)(6) 2026. The procedure was completed successfully by using a 4th AR-4501. There were no patient harm and no secondary procedure needed
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803254 ↗MalfunctionSep 17, 2026Device Dislodged or Dislocated
On 3rd September 2026, it was reported by an Arthrex's employee via an Email that for 3 units of AR-4501, Meniscal Cinch¿ II:The first unit of AR-4501's anchor dropped out when the suture was tightened. A 2nd AR-4501 was changed to but both anchors dropped out too. A 3rd AR-4501 was used but the second anchor dropped out again.This was detected during a knee meniscus repair surgery on (b)(6) 2026. The procedure was completed successfully by using a 4th AR-4501. There were no patient harm and no secondary procedure needed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803249 ↗MalfunctionSep 14, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50756681 ↗MalfunctionAug 17, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50277180 ↗MalfunctionAug 14, 2026Device Dislodged or Dislocated
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50252702 ↗MalfunctionAug 13, 2026Device Dislodged or Dislocated
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50241045 ↗MalfunctionAug 10, 2026Physical Resistance/Sticking
It was reported that during the anterior cruciate ligament surgery, the trigger mechanism did not click into place on the plastic and did not lock. Per complaint reporter there was no harm for patient, operator or third party. The surgery was finished successfully with a new device with the same part number. It was not necessary to switch the surgical technique or do a second surgery.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50200868 ↗MalfunctionAug 4, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50129082 ↗MalfunctionApr 15, 2026Physical Resistance/Sticking
ON 27TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR 2 UNITS AR-4501, MENISCAL CINCH¿ II, THE FIRST BUTTON COULD NOT BE PUSHED, THEN A NEW AR-4501 WAS USED, BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING USE IN A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING THE THIRD ONE. NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED. KNOT PUSHER AND CUTTER PART NUMBER IS AR-11790. 1ST APRIL 2026 - THE COMPLAINT INITIATOR REPLIED THAT THE DEFECT WAS DISCOVERED WHILE THE DEVICE WAS IN THE PATIENT. THE IMPLANT DID NOT DEPLOY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24887168 ↗MalfunctionMar 30, 2026Physical Resistance/Sticking
ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE FIRST BUTTON WAS STUCK AND IMPLANT FAILED TO DEPLOY. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24713617 ↗MalfunctionJan 23, 2026Physical Resistance/Sticking
ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE PUSH BUTTON OF THE FIRST IMPLANT WAS STUCK. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24159844 ↗MalfunctionDec 23, 2025Positioning Failure; Physical Resistance/Sticking; Positioning Failure; Physical Resistance/Sticking
ON 3RD DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT FIXED SUCCESSFULLY BUT WHEN SETTING THE SECOND IMPLANT, THE BUTTON COULD NOT BE RETRIEVED. THE SUTURE WAS STUCK AS WELL. THIS WAS DETECTED DURING AN OPEN REDUCTION AND INTERNAL FIXATION OF TIBIAL FRACTURES WITH PLATES SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881366 ↗MalfunctionDec 17, 2025Break; Break
ON 27TH NOV 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT WAS SET PROPERLY HOWEVER THE SECOND IMPLANT WAS BROKEN. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23827716 ↗MalfunctionNov 24, 2025Positioning Failure; Positioning Failure
ON 4TH NOV 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT WAS SET PROPERLY BUT THE SECOND IMPLANT COULD NOT BE SET. THIS WAS DETECTED DURING AN ARTHROSCOPIC PATELLA RESHAPING, LEFT KNEE MENISCUS SUTURING SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630289 ↗MalfunctionNov 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 3RD NOV 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT WAS SET PROPERLY BUT WHEN TRIGGERED THE SECOND IMPLANT, THE BUTTON WAS STUCK. THIS WAS DETECTED DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION AND MENISCUS INJURY PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23618252 ↗MalfunctionOct 27, 2025Break; Break
ON 15TH OCT 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 3 UNITS OF AR-4501, MENISCAL CINCH¿ II, BOTH IMPLANTS WERE SET PROPERLY BUT THE SUTURE BROKE WHEN USING THE KNOT PUSHER. ANOTHER TWO NEW AR-4501 WERE OPENED BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER BRAND PRODUCT. THERE WERE NO PATIENT HARM OR SECONDARY PROCEDURE NEEDED. PLEASE NOTE: THIS CASE WAS CREATED BECAUSE THE ORIGINAL CASE (B)(4) ENCOUNTERED SYSTEM ISSUES DURING THE UPDATE PROCESS. AS A RESULT, IT VOIDED THE ORIGINAL CASE, AND A NEW CASE (B)(4) WAS CREATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23392412 ↗MalfunctionOct 27, 2025Break; Break
ON 15TH OCT 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 3 UNITS OF AR-4501, MENISCAL CINCH¿ II, BOTH IMPLANTS WERE SET PROPERLY BUT THE SUTURE BROKE WHEN USING THE KNOT PUSHER. ANOTHER TWO NEW AR-4501 WERE OPENED BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER BRAND PRODUCT. THERE WERE NO PATIENT HARM OR SECONDARY PROCEDURE NEEDED. PLEASE NOTE: THIS CASE WAS CREATED BECAUSE THE ORIGINAL CASE (B)(4) ENCOUNTERED SYSTEM ISSUES DURING THE UPDATE PROCESS. AS A RESULT, IT VOIDED THE ORIGINAL CASE, AND A NEW CASE (B)(4) WAS CREATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23392314 ↗MalfunctionOct 27, 2025Break; Break
ON 15TH OCT 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 3 UNITS OF AR-4501, MENISCAL CINCH¿ II, BOTH IMPLANTS WERE SET PROPERLY BUT THE SUTURE BROKE WHEN USING THE KNOT PUSHER. ANOTHER TWO NEW AR-4501 WERE OPENED BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER BRAND PRODUCT. THERE WERE NO PATIENT HARM OR SECONDARY PROCEDURE NEEDED. PLEASE NOTE: THIS CASE WAS CREATED BECAUSE THE ORIGINAL CASE (B)(4) ENCOUNTERED SYSTEM ISSUES DURING THE UPDATE PROCESS. AS A RESULT, IT VOIDED THE ORIGINAL CASE, AND A NEW CASE (B)(4) WAS CREATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23392176 ↗MalfunctionOct 23, 2025Failure to Advance; Failure to Fire; Failure to Advance; Failure to Fire
IT WAS REPORTED THAT DURING A MENISCAL REPAIR SURGERY THE TRIGGER GOT LOCKED AND IT WAS NOT POSSIBLE TO FIRE INTO THE MENISCUS. SOME PARTS BROKE OFF AND COULD BE RETRIEVED OUT OF THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23369477 ↗MalfunctionAug 29, 2025Failure to Fire; Failure to Fire
ON 08/12/2025, A DISTRIBUTOR REPORTED VIA EMAIL THAT AN AR-4501 MENISCAL CINCH II EXPERIENCED AN ISSUE DURING ITS SECOND SUTURE DEPLOYMENT, IN WHICH THE IMPLANT WAS NOT RELEASED. NO ADDITIONAL DETAILS WERE PROVIDED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 08/18/2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22921754 ↗MalfunctionAug 20, 2025Break; Failure to Advance; Device Dislodged or Dislocated; Break; Failure to Advance; Device Dislodged or Dislocated
ON 4TH AUGUST 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR 1 UNIT OF AR-4501, MENISCAL CINCH¿ II, BOTH IMPLANTS WERE DEPLOY AND SET PROPERLY, WHEN RETRIEVING THE SUTURE, IT BROKE, AND BOTH IMPLANTS WAS PULLED OUT. THIS WAS DETECTED DURING A MENISCUS REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WAS NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22854564 ↗MalfunctionAug 13, 2025Break; Break
ON 07/29/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-4501 MENISCAL CINCH¿ II BROKE. THIS OCCURRED DURING A CASE. THE SURGEON INSERTED THE DEVICE INTO THE POSTERIOR HORN, PERFORMED THE FIRST SHOT WITHOUT ANY PROBLEMS, BUT WHEN PERFORMING THE SECOND SHOT AND PROCEEDING TO REMOVE THE CANNULA, IT WAS BROKEN.
Patient problems: Insufficient Information
MAUDE report 22797538 ↗MalfunctionJul 28, 2025Break; Failure to Advance; Break; Failure to Advance
ON 7TH JULY 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT WAS SET SUCCESSFULLY BUT THE SUTURE BROKE WHILE SETTING THE SECOND IMPLANT. THIS WAS DETECTED DURING A KNEE MENISCUS REPAIR PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WAS NO PATIENT HARM, AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22637948 ↗MalfunctionJul 28, 2025Failure to Advance; Failure to Advance
ON 7TH JULY 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-4501, MENISCAL CINCH¿ II, THE FIRST IMPLANT FAILED TO SET. THIS WAS DETECTED DURING A KNEE MENISCUS REPAIR PROCEDURE ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-4501. THERE WAS NO PATIENT HARM, AND NO SECONDARY PROCEDURE NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22637935 ↗
FDA clearance and classification
The FDA regulates product code GAT (suture, nonabsorbable, synthetic, polyethylene) under 21 CFR 878.5000 as a Class II device. 21 CFR 878.5000 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Arthrex® AR-4501
Has the Arthrex® AR-4501 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888867299139), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Arthrex® AR-4501?
The FDA's MAUDE database lists 400 adverse event reports for this device, received between Jan 30, 2019 and Sep 17, 2026. In those reports, the most common specific device problems are “Break” (146), “Device Dislodged or Dislocated” (79) and “Failure to Advance” (49); the most common patient problems are “Failure of Implant” (15), “Foreign Body In Patient” (11) and “Device Embedded In Tissue or Plaque” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Arthrex® AR-4501?
None of the reports are classified as deaths, 67 as injuries and 333 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Arthrex® AR-4501 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the Arthrex® AR-4501, and is it still sold?
The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-4501). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices
- FAST-FIX 72202468Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 death6,566 reports
- FAST-FIX 72201491Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE862 reports
- FAST-FIX 72202469Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE561 reports
- FAST-FIX 72205676Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE428 reports
- FAST-FIX 72205324Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE381 reports
- LUPINE ORTHOCORD 210712Medos International Sàrl · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE352 reports
- FAST-FIX 72201494Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE249 reports
- ETHIBOND EXCEL X557HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE187 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page