SPEEDSTITCH OM-8850Recalls & reported problems
Smith & Nephew, Inc. · SPEEDSTITCH NEEDLE CASSETTE INSERT · Class 2
All models of SPEEDSTITCH (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Speedstitch needle cassette insert is a suturing tool used during orthopedic procedures such as hip arthroscopy. The 42 reports mostly describe the needle breaking, bending, or separating during surgery, in some cases leaving the needle tip or fragments inside the patient, with no deaths and 30 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 30
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 00817470005929), received between Apr 9, 2018 and Dec 5, 2023 — classified as 30 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Material Separation” (27 reports).
1 other recall cover devices cleared under the same FDA submission (K042031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation27
- Break14
- Material Twisted/Bent6
- Detachment Of Device Component1
- Failure to Cycle1
- Material Deformation1
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient28
- Foreign Body In Patient17
- No Clinical Signs, Symptoms or Conditions7
- Device Embedded In Tissue or Plaque4
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2023Break; Break
IT WAS REPORTED THAT, DURING A HIP SURGERY, THE SPEEDSTITCH NEEDLE'S UNLOADER WAS ALREADY BROKEN IN TWO PIECES WHEN THE PACKET WAS OPENED AND COULD NOT BE USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18267649 ↗MalfunctionMay 22, 2023Failure to Cycle; Material Deformation; Failure to Cycle; Material Deformation
IT WAS REPORTED THAT TWO (2) SPEEDSTITCH NEEDLE WERE NOT GRASPING THE SUTURE WHEN PASSING IT THROUGH THE HIP CAPSULE. EVENTUALLY, THE NEEDLE TIP BENT AND IT WAS NOT POSSIBLE TO PASS OR CATCH THE SUTURE. SURGERY WAS COMPLETED AFTER A 45-MINUTE DELAY, USING A SLINGSHOT SUTURE PASSER DEVICE FROM STRYKER (COMPETITOR DEVICE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16976397 ↗MalfunctionMay 22, 2023Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE, A SPEEDSTICH NEEDLE BROKE OFF IN THE CAPSULAR TISSUE, WHEN CLOSING THE CAPSULE WITH THE DEVICE, THE NEEDLE TIP WAS REMOVED FROM TISSUE, WHICH CAUSED A DELAY, HOWEVER UNKNOWN OF HOW MUCH, AS WELL IT IS UNKNOWN IF THERE WAS A BACK-UP DEVICE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16975954 ↗InjuryDec 9, 2021Break; Break
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY WITH LABRAL REPAIR, THE SPEEDSTITCH NEEDLE BROKE OFF IN THE PATIENT AND NOT ALL THE PIECES WERE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A 30 MIN DELAY USING A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 12968557 ↗InjuryNov 23, 2021Break; Break
IT WAS REPORTED THAT DURING A HIP PROCEDURE, INTERNAL TO PATIENT, THE TIP OF THE SPEEDSTITCH NEEDLE BROKE OFF AND WAS NOT ABLE TO BE "FISHED OUT" OF THE HIP CAPSULE OF THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 12861836 ↗InjuryMay 30, 2021Break; Break
IT WAS REPORTED THAT DURING THE SURGERY THE SPEEDSTITCH NEEDLE TIP BROKE OFF INSIDE THE PATIENT'S ANATOMY, THE BROKEN PIECES WERE FOUND WITH A SCOPE AND THEN REMOVED. DELAY GREATER THAN 30 MINUTES WAS REPORTED. SMITH AND NEPHEW BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. NO OTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11905797 ↗MalfunctionOct 21, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE SPEEDSTITCH NEEDLE FAILED TO PASS SUTURE TAPE. THE MALFUNCTION HAPPENED DURING AN UNKNOWN PROCEDURE OUTSIDE THE PATIENT; BUT IT WAS COMPLETED USING A BACK UP DEVICE WITH NO DELAY OR PATIENT INJURY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10714899 ↗MalfunctionJun 24, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY SURGERY THE NEEDLE TIP WAS FAULTY. NO PATIENT INJURIES REPORTED. BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. UNKNOWN IF THERE WAS A DELAY OF THE SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10189416 ↗MalfunctionJun 24, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING SURGERY THE SPEEDSTITCH NEEDLE WAS DAMAGED AT THE END AND WOULD NOT PASS THE SUTURE. THE PROCEDURE WAS COMPLETED WITH A DELAY OF UNDER 30 MIN AND A BACKUP DEVICE WAS AVAILABLE. NO INJURY TO THE PATIENT WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10189336 ↗MalfunctionJun 24, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING SURGERY THE SPEEDSTITCH NEEDLE WAS DAMAGED AT THE END AND WOULD NOT PASS THE SUTURE. THE PROCEDURE WAS COMPLETED WITH A DELAY OF UNDER 30 MIN AND A BACKUP DEVICE WAS AVAILABLE. NO INJURY TO THE PATIENT WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10189282 ↗MalfunctionJun 23, 2020Material Separation; Material Separation
IT WAS REPORTED THAT, DURING A HIP ARTHROSCOPY, THE SPEEDSTITCH NEEDLE WAS BENDING OVER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10187146 ↗InjuryJun 8, 2020Break; Break
IT WAS REPORTED THAT WHEN THE SURGEON WAS PERFORMING A HIP SCOPE, A SPEEDSTITCH NEEDLE BROKE INSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A COMPETITOR DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10131073 ↗InjuryJun 8, 2020Break; Break
IT WAS REPORTED THAT WHEN THE SURGEON WAS PERFORMING A HIP SCOPE, A SPEEDSTITCH NEEDLE BROKE INSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACKUP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10130935 ↗InjuryJun 8, 2020Break; Break
IT WAS REPORTED THAT WHEN THE SURGEON WAS PERFORMING A HIP SCOPE, A SPEEDSTITCH NEEDLE BROKE INSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACKUP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10130880 ↗InjuryMay 30, 2020Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE, WHEN DEPLOYING THE DEVICE TO PASS THE TAPE OR SUTURE THE NEEDLE WOULD JAM OR BREAK THE TIP OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103835 ↗InjuryMay 30, 2020Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE, WHEN DEPLOYING THE DEVICE TO PASS THE TAPE OR SUTURE THE NEEDLE WOULD JAM OR BREAK THE TIP OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103830 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE, WHEN DEPLOYING THE DEVICE TO PASS THE TAPE OR SUTURE THE NEEDLE WOULD JAM OR BREAK THE TIP OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103818 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE, WHEN DEPLOYING THE DEVICE TO PASS THE TAPE OR SUTURE THE NEEDLE WOULD JAM OR BREAK THE TIP OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103816 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE IN THE HIP, THE NEEDLE WOULD FAIL TO PASS THE STITCH OR MINI TAPE AND WOULD SOMETIMES BREAK OFF OR BEND. THE BROKEN PIECE COULD NOT BE RETRIEVED SINCE IT WAS BURIED IN CAPSULE TISSUE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103814 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE IN THE HIP, THE NEEDLE WOULD FAIL TO PASS THE STITCH OR MINI TAPE AND WOULD SOMETIMES BREAK OFF OR BEND. THE BROKEN PIECE COULD NOT BE RETRIEVED SINCE IT WAS BURIED IN CAPSULE TISSUE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103813 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHILE USING THE SPEEDSTITCH TO CLOSE THE CAPSULE IN THE HIP, THE NEEDLE WOULD FAIL TO PASS THE STITCH OR MINI TAPE AND WOULD SOMETIMES BREAK OFF OR BEND. THE BROKEN PIECE COULD NOT BE RETRIEVED SINCE IT WAS BURIED IN CAPSULE TISSUE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 10103808 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP CAPSULE CLOSURE WHEN USING A SPEEDSTITCHES, WHEN DEPLOYING THE NEEDLES TO PASS THE SUTURES OR TAPE, THE NEEDLES TIP WOULD BEND AND IN SOME CASES IT BROKE OFF AT THE TIP. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACKUP DEVICE, NO SIGNIFICANT DELAY REPORTED. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BE BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10103765 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP CAPSULE CLOSURE WHEN USING A SPEEDSTITCHES, WHEN DEPLOYING THE NEEDLES TO PASS THE SUTURES OR TAPE, THE NEEDLES TIP WOULD BEND AND IN SOME CASES IT BROKE OFF AT THE TIP. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACKUP DEVICE, NO SIGNIFICANT DELAY REPORTED. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BE BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10103763 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP CAPSULE CLOSURE WHEN USING A SPEEDSTITCHES, WHEN DEPLOYING THE NEEDLES TO PASS THE SUTURES OR TAPE, THE NEEDLES TIP WOULD BEND AND IN SOME CASES IT BROKE OFF AT THE TIP. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACKUP DEVICE, NO SIGNIFICANT DELAY REPORTED. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BE BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10103762 ↗InjuryMay 30, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP CAPSULE CLOSURE WHEN USING A SPEEDSTITCHES, WHEN DEPLOYING THE NEEDLES TO PASS THE SUTURES OR TAPE, THE NEEDLES TIP WOULD BEND AND IN SOME CASES IT BROKE OFF AT THE TIP. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACKUP DEVICE, NO SIGNIFICANT DELAY REPORTED. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BE BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10103760 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042031. One submission often covers many models, so check the product and lot details in each record.
- Z-0611-2012 · Aug 8, 2011 · TerminatedOPUS SpeedStitch Product Number OM-8086 is a suture cartridge For use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices
- FAST-FIX 72202468Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 death6,626 reports
- FAST-FIX 72201491Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE863 reports
- FAST-FIX 72202469Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE561 reports
- FAST-FIX 72205676Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE439 reports
- Arthrex® AR-4501Arthrex, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE403 reports
- FAST-FIX 72205324Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE383 reports
- LUPINE ORTHOCORD 210712Medos International Sàrl · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE357 reports
- FAST-FIX 72201494Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE249 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page