GP General Purpose Drainage Catheter GPL2-1230HRecalls & reported problems
Uresil, LLC · HYDROPHILIC COATED GP LOCKING CATHETER WITH DEPTH MARKS, 12F · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817640020066), received between May 15, 2025 and May 15, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
Patient problems
- Bacteremia1
- High White Blood Cell Count1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2025Crack; Crack
NARRATIVE FROM HEALTHCARE FACILITY: PATIENT PRESENTING WITH PERSISTENT RLQ PERIAPPENDICEAL ABSCESS, ASSOCIATED RECURRENT INFECTION, AND BROKEN/MALFUNCTIONING PERCUTANEOUS DRAIN. THE CURRENT MANAGEMENT FRAMEWORK IS TO HAVE THIS CAVITY FULLY DRAINED AND COLLAPSED AND THEN TO UNDERGO INTERVAL APPENDECTOMY. ON INITIAL ADMISSION EXAM, HR IN 120S, NO MEASURED FEVER IN ED, VERY TENDER IN RLQ. DRAIN NOT HOLDING SUCTION (LUER LOCK CRACKED). WBC 20. SEPSIS FLUIDS INITIATED. CT SCAN DEMONSTRATES PERSISTENT (BUT DECREASED) WALLED-OFF ABSCESS. PREVIOUS 8CM, NOW 5CM. THE HUB [LUER] OF THE DRAIN WAS FRACTURED. THEREFORE, THE JP [JACKSON-PRATT] BULB COULD NOT BE ATTACHED TO EFFECTIVELY DRAIN THE ABSCESS. THE DRAIN MALFUNCTION DIRECTLY LEAD TO ABSCESS PROGRESSION, ELEVATED WHITE BLOOD CELL COUNT AND FEVER. THE CLINICAL PICTURE WAS CONSISTENT WITH BACTEREMIA. THE PATIENT BECAME SO ILL, THAT SHE COULD NOT WAIT FOR OUTPATIENT TREATMENT. SHE CAME TO THE ER AND WAS DIRECTLY ADMITTED TO THE HOSPITAL. PATIENT HAS SINCE BEEN DISCHARGED FROM THE HOSPITAL.
Patient problems: Bacteremia; High White Blood Cell Count
MAUDE report 22030987 ↗
Other Catheter, Nephrostomy, General & Plastic Surgery devices
- NA M0067001210Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY2 deaths48 reports
- NA M0064201110Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY6 reports
- Flexima™ M001271800Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY6 reports
- Skater™ All-Purpose and Nephrostomy Set 756608025Argon Medical Devices, Inc. · Catheter, Nephrostomy, General & Plastic Surgery5 reports
- Flexima™ M001271790Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY4 reports
- Re-Entry Malecot Nephrostomy Catheter Set M0064101060Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY4 reports
- SKATER™ All Purpose and Nephrostomy Drainage Set 756608035Argon Medical Devices, Inc. · Catheter, Nephrostomy, General & Plastic Surgery4 reports
- Flexima™ M001271860Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page