Paramount Mini GPS PMB8-5-14-80Recalls & reported problems
Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684006566), received between Mar 5, 2019 and Mar 5, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
Patient problems
- Patient Problem/Medical Problem1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 5, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE PARAMOUNT MINI STENT TO TREAT A RIGHT DISTAL LESION IN THE ILIAC ARTERY COMMON OF A PATIENT IN A PROCEDURE. NO DAMAGE NOTED TO PACKAGING NOTED, NO ISSUES NOTED WHEN REMOVING THE DEVICES FROM THE HOOP/TRAY, DEVICES WAS PREPPED AS PER IFU. IT WAS REPORTED THAT ONE OF THE STENT¿S DISLODGEMENT OCCURRED DURING DELIVERY TO/AT LESION, IT JUST CAME LOOSE. THE DISLODGED SENT WAS REMOVED USING A SNARE. THE SECOND STENT DISLODGED BUT WAS AND FULLY EXPANDED IN THE EXTERNAL ILIAC ARTERY. THE DEVICES DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT, RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICES, EXCESSIVE FORCE WAS NOT USED. NO FURTHER PATIENT INJURY REPORTED FOR THIS EVENT.
Patient problems: Patient Problem/Medical Problem
MAUDE report 8389447 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page