Skip to content

Paramount Mini GPS PMB8-5-18-80Recalls & reported problems

Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684006597), received between Dec 1, 2017 and Dec 1, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 1, 2017Material Deformation; Material Deformation

    THE PHYSICIAN WAS ATTEMPTING USE A PARAMOUNT MINI GPS BALLOON-EXPANDABLE STENT TO TREAT A 15MM PLAQUE LESION WITH LITTLE CALCIFICATION IN THE RIGHT PROXIMAL RENAL ARTERY EXHIBITING 90-95% STENOSIS. THE ARTERY WAS 5 MM IN DIAMETER, AND WAS EXTREMELY TORTUOUS. THE ARTERY WAS REPORTED TO HAVE A VERY SIGNIFICANT ANGLE OF ORIGIN OFF THE AORTA. THE PROCEDURE USED A 6 FR NON-MEDTRONIC SHEATH AND A 0.018" NON-MEDTRONIC GUIDEWIRE. THE VESSEL WAS PRE-DILATED USING A NON-MEDTRONIC 3X20 DEVICE AT 8ATMS FOR 45 SEC WHICH WAS THEN INCREASED TO 12 ATMS AND BALLOON RUPTURED BUT WAS REMOVED INTACT. THE PHYSICIAN THEN ATTEMPTED TO USE A 5X18 PARAMOUNT MINI BALLOON-EXPANDABLE STENT WHICH FAILED TO CROSS THE LESION. THERE WAS DIFFICULTY REMOVING THE PARAMOUNT MINI STENT. THE VESSEL WAS THEN PRE-DILATED FOR A SECOND TIME WITH A MDT 4X40 PLUS AT 8 ATMS FOR 60 SECONDS. THE PHYSICIAN EXCHANGED THE WIRES TO NON-MEDTRONIC 0.014". THE PHYSICIAN THEN ATTEMPTED TO USE ANOTHER PARAMOUNT MINI BALLOON-EXPANDABLE 5X14 STENT, WHICH ALSO FAILED TO CROSS THE LESION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A 5X40 NON-MEDTRONIC PTA BALLOON INFLATED TO 8 ATMS FOR 60 SECONDS. NO PATIENT INJURY WAS REP ORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7077308 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →