Visi-Pro PXB35-05-27-080Recalls & reported problems
Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant
All models of Visi-Pro (66)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684036228), received between Jun 2, 2020 and Jun 2, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K061566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2020Burst Container or Vessel; Positioning Problem; Burst Container or Vessel; Positioning Problem
PHYSICIAN WAS ATTEMPTING TO USE A VISI PRO BALLOON EXPANDABLE STENT DURING PROCEDURE TO TREAT A SEVERELY CALCIFIED LESION IN THE PROXIMAL COMMON ILIAC ARTERY WITH 75% STENOSIS. THE VESSEL DIAMETER AND LESION LENGTH ARE 6MM AND 20MM RESPECTIVELY. THE VESSEL WAS MODERATELY TORTUOUS. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUES NOTED WHEN REMOVING DEVICE FROM HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. IT WAS REPORTED THAT THE BALLOON RUPTURED CIRCUMFERENTIALLY DURING STENT DEPLOYMENT. INFLATION DEVICE WAS USED TO INFLATE BALLOON. THE BALLOON RUPTURED AT 8ATM. THE LESION WAS PRE DILATED USING A 8X60X80 EVERCROSS PTA BALLOON. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THERE WAS NO RESISTANCE WHEN ADVANCING THE DEVICE. THE BALLOON DID NOT FRAGMENT. A NON-MEDTRONIC DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE VISIPRO STENT WAS IMPLANTED AT THE TARGET LESION AND POST DILATED WITH A NON-MEDTRONIC COVERED STENT. THERE WAS NO VESSEL DAMAGE RELATED TO THE BALLOON RUPTURE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10110978 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061566. One submission often covers many models, so check the product and lot details in each record.
- Z-1547-2008 · Jul 1, 2008 · Terminatedev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PX
- Z-1548-2008 · Jul 1, 2008 · Terminatedev3" sterile EO, single Use, Visi-Pro Balloon-Expandable Biliary Stent System. Model Numbers: PXB35-06-27-135, PXB35-06-37-080, PXB35-07-37-080, PXB35-08-17-080 , PXB35-08-27
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.